The second gamma-H2AX assay inter-comparison exercise carried out in the framework of the European biodosimetry network (RENEB)

  • Jayne Moquet
  • Stephen Barnard
  • Albena Staynova
  • Carita Lindholm
  • Octávia Monteiro Gil
  • Vanda Martins
  • Ute Rößler
  • Anne Vral
  • Charlot Vandevoorde
  • Maria Wojewódzka
  • Kai Rothkamm

Abstract

PURPOSE: Within the EU RENEB project, seven laboratories have taken part in training and harmonisation activities to strengthen triage gamma-H2AX-based radiation exposure assessment. This has culminated in a second triage biodosimetry exercise.

MATERIALS AND METHODS: Whole blood and separated lymphocyte samples were homogenously irradiated with (60)Co gamma rays at 0.5, 2.5 (blind samples), 0 and 2 Gy (reference samples). Following post-exposure incubations of 4 and 24 h, 16 samples were shipped on ice packs to each partner. The samples were stained and scored for gamma-H2AX foci, using manual and/or automated fluorescence microscope scoring strategies. Dose estimates were obtained and used to assign triage categories to the samples.

RESULTS: Average dose estimates across all the laboratories correlated well with true doses. The most accurate assignment of triage category was achieved by manual scoring of the 4-h blood and lymphocyte samples. Only three samples out of a total of 46 were miscategorized in a way that could have adversely effected the clinical management of a radiation casualty.

CONCLUSIONS: This inter-comparison exercise has demonstrated that following a recent acute radiation exposure, the gamma-H2AX assay could be a useful triage tool that can be successfully applied across a network of laboratories.