Robotic versus conventional ablation for common-type atrial flutter: a prospective randomized trial to evaluate the effectiveness of remote catheter navigation

Standard

Robotic versus conventional ablation for common-type atrial flutter: a prospective randomized trial to evaluate the effectiveness of remote catheter navigation. / Steven, Daniel; Rostock, Thomas; Servatius, Helge; Hoffmann, Boris; Drewitz, Imke; Müllerleile, Kai; Meinertz, Thomas; Willems, Stephan.

In: HEART RHYTHM, Vol. 5, No. 11, 11.2008, p. 1556-1560.

Research output: SCORING: Contribution to journalSCORING: Journal articleResearchpeer-review

Harvard

APA

Vancouver

Bibtex

@article{a1f675d664504b818cb9b7892daa63f8,
title = "Robotic versus conventional ablation for common-type atrial flutter: a prospective randomized trial to evaluate the effectiveness of remote catheter navigation",
abstract = "BACKGROUND: Conventional catheter ablation for common-type atrial flutter (AFL) is a widely established therapy but has not been compared with the use of a robotic navigation system (RNS) thus far.OBJECTIVES: The purpose of this study was to investigate the feasibility of a new, nonmagnetic RNS with regard to safety, efficacy, and X-ray exposure to investigator and patient compared with the conventional ablation approach in patients with AFL.METHODS: Fifty patients (65.7 +/- 9.3 years, 40 male) undergoing de novo catheter ablation for AFL were randomly assigned to conventional or RNS-guided cavotricuspid isthmus (CTI) ablation.RESULTS: Complete bidirectional isthmus block was achieved for all patients without occurrence of procedure-related complications. The fluoroscopy time and the investigator X-ray exposure (8.2 +/- 4.6 vs. 5.8 +/- 3.6, P = .038; and 8.2 +/- 4.6 vs. 1.9 +/- 1.1 minutes, P<.001) as well as the mean radiofrequency (RF) duration and the energy delivered were significantly higher in the conventional than in the RNS group (321.7 +/- 214.6 vs. 496.4 +/- 213.9 seconds, P = .006; 8279 +/- 5767 vs. 16,308 +/- 6870 J, P<.001, respectively). The overall procedure time in the RNS group was significantly longer than in the conventional group (79.2 +/- 30.6 vs. 58.4 +/- 17.7 minutes; P = .04) but significantly decreased comparing the first 10 with the last 10 patients in the RNS group (105.3 +/- 34.8 vs. 60.6 +/- 6.3 minutes; P = .003). Starting ablation during AFL, bidirectional block instantly after termination was observed in 90% of the RNS and 50% of the conventionally treated patients (P = .03).CONCLUSION: The present study demonstrates the safety and feasibility of RNS for performing CTI ablation in patients with common-type AFL for use in the clinical routine. As a result of the remote navigation, X-ray exposure and RF duration to achieve bidirectional block were significantly decreased and occurred more often immediately after AFL termination. These findings are consistent with increased catheter stability and RF application efficacy using RNS compared with conventional catheter manipulation.",
keywords = "Aged, Atrial Flutter/surgery, Catheter Ablation/instrumentation, Equipment Design, Female, Heart Conduction System/surgery, Humans, Male, Middle Aged, Prospective Studies, Robotics, Treatment Outcome",
author = "Daniel Steven and Thomas Rostock and Helge Servatius and Boris Hoffmann and Imke Drewitz and Kai M{\"u}llerleile and Thomas Meinertz and Stephan Willems",
year = "2008",
month = nov,
doi = "10.1016/j.hrthm.2008.08.028",
language = "English",
volume = "5",
pages = "1556--1560",
journal = "HEART RHYTHM",
issn = "1547-5271",
publisher = "Elsevier",
number = "11",

}

RIS

TY - JOUR

T1 - Robotic versus conventional ablation for common-type atrial flutter: a prospective randomized trial to evaluate the effectiveness of remote catheter navigation

AU - Steven, Daniel

AU - Rostock, Thomas

AU - Servatius, Helge

AU - Hoffmann, Boris

AU - Drewitz, Imke

AU - Müllerleile, Kai

AU - Meinertz, Thomas

AU - Willems, Stephan

PY - 2008/11

Y1 - 2008/11

N2 - BACKGROUND: Conventional catheter ablation for common-type atrial flutter (AFL) is a widely established therapy but has not been compared with the use of a robotic navigation system (RNS) thus far.OBJECTIVES: The purpose of this study was to investigate the feasibility of a new, nonmagnetic RNS with regard to safety, efficacy, and X-ray exposure to investigator and patient compared with the conventional ablation approach in patients with AFL.METHODS: Fifty patients (65.7 +/- 9.3 years, 40 male) undergoing de novo catheter ablation for AFL were randomly assigned to conventional or RNS-guided cavotricuspid isthmus (CTI) ablation.RESULTS: Complete bidirectional isthmus block was achieved for all patients without occurrence of procedure-related complications. The fluoroscopy time and the investigator X-ray exposure (8.2 +/- 4.6 vs. 5.8 +/- 3.6, P = .038; and 8.2 +/- 4.6 vs. 1.9 +/- 1.1 minutes, P<.001) as well as the mean radiofrequency (RF) duration and the energy delivered were significantly higher in the conventional than in the RNS group (321.7 +/- 214.6 vs. 496.4 +/- 213.9 seconds, P = .006; 8279 +/- 5767 vs. 16,308 +/- 6870 J, P<.001, respectively). The overall procedure time in the RNS group was significantly longer than in the conventional group (79.2 +/- 30.6 vs. 58.4 +/- 17.7 minutes; P = .04) but significantly decreased comparing the first 10 with the last 10 patients in the RNS group (105.3 +/- 34.8 vs. 60.6 +/- 6.3 minutes; P = .003). Starting ablation during AFL, bidirectional block instantly after termination was observed in 90% of the RNS and 50% of the conventionally treated patients (P = .03).CONCLUSION: The present study demonstrates the safety and feasibility of RNS for performing CTI ablation in patients with common-type AFL for use in the clinical routine. As a result of the remote navigation, X-ray exposure and RF duration to achieve bidirectional block were significantly decreased and occurred more often immediately after AFL termination. These findings are consistent with increased catheter stability and RF application efficacy using RNS compared with conventional catheter manipulation.

AB - BACKGROUND: Conventional catheter ablation for common-type atrial flutter (AFL) is a widely established therapy but has not been compared with the use of a robotic navigation system (RNS) thus far.OBJECTIVES: The purpose of this study was to investigate the feasibility of a new, nonmagnetic RNS with regard to safety, efficacy, and X-ray exposure to investigator and patient compared with the conventional ablation approach in patients with AFL.METHODS: Fifty patients (65.7 +/- 9.3 years, 40 male) undergoing de novo catheter ablation for AFL were randomly assigned to conventional or RNS-guided cavotricuspid isthmus (CTI) ablation.RESULTS: Complete bidirectional isthmus block was achieved for all patients without occurrence of procedure-related complications. The fluoroscopy time and the investigator X-ray exposure (8.2 +/- 4.6 vs. 5.8 +/- 3.6, P = .038; and 8.2 +/- 4.6 vs. 1.9 +/- 1.1 minutes, P<.001) as well as the mean radiofrequency (RF) duration and the energy delivered were significantly higher in the conventional than in the RNS group (321.7 +/- 214.6 vs. 496.4 +/- 213.9 seconds, P = .006; 8279 +/- 5767 vs. 16,308 +/- 6870 J, P<.001, respectively). The overall procedure time in the RNS group was significantly longer than in the conventional group (79.2 +/- 30.6 vs. 58.4 +/- 17.7 minutes; P = .04) but significantly decreased comparing the first 10 with the last 10 patients in the RNS group (105.3 +/- 34.8 vs. 60.6 +/- 6.3 minutes; P = .003). Starting ablation during AFL, bidirectional block instantly after termination was observed in 90% of the RNS and 50% of the conventionally treated patients (P = .03).CONCLUSION: The present study demonstrates the safety and feasibility of RNS for performing CTI ablation in patients with common-type AFL for use in the clinical routine. As a result of the remote navigation, X-ray exposure and RF duration to achieve bidirectional block were significantly decreased and occurred more often immediately after AFL termination. These findings are consistent with increased catheter stability and RF application efficacy using RNS compared with conventional catheter manipulation.

KW - Aged

KW - Atrial Flutter/surgery

KW - Catheter Ablation/instrumentation

KW - Equipment Design

KW - Female

KW - Heart Conduction System/surgery

KW - Humans

KW - Male

KW - Middle Aged

KW - Prospective Studies

KW - Robotics

KW - Treatment Outcome

U2 - 10.1016/j.hrthm.2008.08.028

DO - 10.1016/j.hrthm.2008.08.028

M3 - SCORING: Journal article

C2 - 18984532

VL - 5

SP - 1556

EP - 1560

JO - HEART RHYTHM

JF - HEART RHYTHM

SN - 1547-5271

IS - 11

ER -