Initial Clinical Experience Using the Low-Profile Altura Endograft System With Double D-Shaped Proximal Stents for Endovascular Aneurysm Repair

Standard

Initial Clinical Experience Using the Low-Profile Altura Endograft System With Double D-Shaped Proximal Stents for Endovascular Aneurysm Repair. / Krievins, Dainis; Krämer, Albrecht; Savlovskis, Janis; Oszkinis, Georgij; Debus, E Sebastian; Oberhuber, Alexander; Zarins, Christopher K.

In: J ENDOVASC THER, Vol. 25, No. 3, 06.2018, p. 379-386.

Research output: SCORING: Contribution to journalSCORING: Journal articleResearchpeer-review

Harvard

APA

Vancouver

Bibtex

@article{bd4ef3aea3d64341b65a61e1ef9d2c9b,
title = "Initial Clinical Experience Using the Low-Profile Altura Endograft System With Double D-Shaped Proximal Stents for Endovascular Aneurysm Repair",
abstract = "PURPOSE: To report the initial clinical results of endovascular aneurysm repair (EVAR) using the low-profile (14-F) Altura Endograft System, which features a double {"}D-shaped{"} stent design with suprarenal fixation and modular iliac components that are deployed from distal to proximal.METHODS: From 2011 to 2015, 90 patients (mean age 72.8±8.3 years; 79 men) with abdominal aortic aneurysm (AAA; mean diameter 53.8±5.7 mm) were treated at 10 clinical sites in 2 prospective, controlled clinical studies using the Altura endograft. Outcomes evaluated included mortality, major adverse events (MAEs: all-cause death, stroke, paraplegia, myocardial infarction, respiratory failure, bowel ischemia, and blood loss ≥1000 mL), and clinical success (freedom from procedure-related death, type I/III endoleak, migration, thrombosis, and reintervention).RESULTS: Endografts were successfully implanted in 89 (99%) patients; the single failure was due to delivery system malfunction before insertion in the early-generation device. One (1%) patient died and 4 patients underwent reinterventions (1 type I endoleak, 2 iliac limb stenoses, and 1 endograft occlusion) within the first 30 days. During a median follow-up of 12.5 months (range 11.5-50.9), there were no aneurysm ruptures, surgical conversions, or AAA-related deaths. The cumulative MAE rates were 3% (3/89) at 6 months and 7% (6/89) at 1 year. Two patients underwent coil embolization of type II endoleaks at 6.5 months and 2.2 years, respectively. Clinical success was 94% (84/89) at 30 days, 98% (85/87) at 6 months, and 99% (82/83) at 1 year.CONCLUSION: Early results suggest that properly selected AAA patients can be safely treated using the Altura Endograft System with favorable midterm outcome. Thus, further clinical investigation is warranted to evaluate the role of this device in the treatment of AAA.",
keywords = "Aged, Aged, 80 and over, Aortic Aneurysm, Abdominal/diagnostic imaging, Blood Vessel Prosthesis, Blood Vessel Prosthesis Implantation/adverse effects, Endovascular Procedures/adverse effects, Female, Humans, Male, Middle Aged, Patient Selection, Postoperative Complications/etiology, Prosthesis Design, Prosthesis Failure, Risk Assessment, Risk Factors, Stents, Time Factors, Treatment Outcome",
author = "Dainis Krievins and Albrecht Kr{\"a}mer and Janis Savlovskis and Georgij Oszkinis and Debus, {E Sebastian} and Alexander Oberhuber and Zarins, {Christopher K}",
year = "2018",
month = jun,
doi = "10.1177/1526602818771973",
language = "English",
volume = "25",
pages = "379--386",
journal = "J ENDOVASC THER",
issn = "1526-6028",
publisher = "International Society of Endovascular Specialists",
number = "3",

}

RIS

TY - JOUR

T1 - Initial Clinical Experience Using the Low-Profile Altura Endograft System With Double D-Shaped Proximal Stents for Endovascular Aneurysm Repair

AU - Krievins, Dainis

AU - Krämer, Albrecht

AU - Savlovskis, Janis

AU - Oszkinis, Georgij

AU - Debus, E Sebastian

AU - Oberhuber, Alexander

AU - Zarins, Christopher K

PY - 2018/6

Y1 - 2018/6

N2 - PURPOSE: To report the initial clinical results of endovascular aneurysm repair (EVAR) using the low-profile (14-F) Altura Endograft System, which features a double "D-shaped" stent design with suprarenal fixation and modular iliac components that are deployed from distal to proximal.METHODS: From 2011 to 2015, 90 patients (mean age 72.8±8.3 years; 79 men) with abdominal aortic aneurysm (AAA; mean diameter 53.8±5.7 mm) were treated at 10 clinical sites in 2 prospective, controlled clinical studies using the Altura endograft. Outcomes evaluated included mortality, major adverse events (MAEs: all-cause death, stroke, paraplegia, myocardial infarction, respiratory failure, bowel ischemia, and blood loss ≥1000 mL), and clinical success (freedom from procedure-related death, type I/III endoleak, migration, thrombosis, and reintervention).RESULTS: Endografts were successfully implanted in 89 (99%) patients; the single failure was due to delivery system malfunction before insertion in the early-generation device. One (1%) patient died and 4 patients underwent reinterventions (1 type I endoleak, 2 iliac limb stenoses, and 1 endograft occlusion) within the first 30 days. During a median follow-up of 12.5 months (range 11.5-50.9), there were no aneurysm ruptures, surgical conversions, or AAA-related deaths. The cumulative MAE rates were 3% (3/89) at 6 months and 7% (6/89) at 1 year. Two patients underwent coil embolization of type II endoleaks at 6.5 months and 2.2 years, respectively. Clinical success was 94% (84/89) at 30 days, 98% (85/87) at 6 months, and 99% (82/83) at 1 year.CONCLUSION: Early results suggest that properly selected AAA patients can be safely treated using the Altura Endograft System with favorable midterm outcome. Thus, further clinical investigation is warranted to evaluate the role of this device in the treatment of AAA.

AB - PURPOSE: To report the initial clinical results of endovascular aneurysm repair (EVAR) using the low-profile (14-F) Altura Endograft System, which features a double "D-shaped" stent design with suprarenal fixation and modular iliac components that are deployed from distal to proximal.METHODS: From 2011 to 2015, 90 patients (mean age 72.8±8.3 years; 79 men) with abdominal aortic aneurysm (AAA; mean diameter 53.8±5.7 mm) were treated at 10 clinical sites in 2 prospective, controlled clinical studies using the Altura endograft. Outcomes evaluated included mortality, major adverse events (MAEs: all-cause death, stroke, paraplegia, myocardial infarction, respiratory failure, bowel ischemia, and blood loss ≥1000 mL), and clinical success (freedom from procedure-related death, type I/III endoleak, migration, thrombosis, and reintervention).RESULTS: Endografts were successfully implanted in 89 (99%) patients; the single failure was due to delivery system malfunction before insertion in the early-generation device. One (1%) patient died and 4 patients underwent reinterventions (1 type I endoleak, 2 iliac limb stenoses, and 1 endograft occlusion) within the first 30 days. During a median follow-up of 12.5 months (range 11.5-50.9), there were no aneurysm ruptures, surgical conversions, or AAA-related deaths. The cumulative MAE rates were 3% (3/89) at 6 months and 7% (6/89) at 1 year. Two patients underwent coil embolization of type II endoleaks at 6.5 months and 2.2 years, respectively. Clinical success was 94% (84/89) at 30 days, 98% (85/87) at 6 months, and 99% (82/83) at 1 year.CONCLUSION: Early results suggest that properly selected AAA patients can be safely treated using the Altura Endograft System with favorable midterm outcome. Thus, further clinical investigation is warranted to evaluate the role of this device in the treatment of AAA.

KW - Aged

KW - Aged, 80 and over

KW - Aortic Aneurysm, Abdominal/diagnostic imaging

KW - Blood Vessel Prosthesis

KW - Blood Vessel Prosthesis Implantation/adverse effects

KW - Endovascular Procedures/adverse effects

KW - Female

KW - Humans

KW - Male

KW - Middle Aged

KW - Patient Selection

KW - Postoperative Complications/etiology

KW - Prosthesis Design

KW - Prosthesis Failure

KW - Risk Assessment

KW - Risk Factors

KW - Stents

KW - Time Factors

KW - Treatment Outcome

U2 - 10.1177/1526602818771973

DO - 10.1177/1526602818771973

M3 - SCORING: Journal article

C2 - 29673298

VL - 25

SP - 379

EP - 386

JO - J ENDOVASC THER

JF - J ENDOVASC THER

SN - 1526-6028

IS - 3

ER -