Initial Clinical Experience Using the Low-Profile Altura Endograft System With Double D-Shaped Proximal Stents for Endovascular Aneurysm Repair
Standard
Initial Clinical Experience Using the Low-Profile Altura Endograft System With Double D-Shaped Proximal Stents for Endovascular Aneurysm Repair. / Krievins, Dainis; Krämer, Albrecht; Savlovskis, Janis; Oszkinis, Georgij; Debus, E Sebastian; Oberhuber, Alexander; Zarins, Christopher K.
in: J ENDOVASC THER, Jahrgang 25, Nr. 3, 06.2018, S. 379-386.Publikationen: SCORING: Beitrag in Fachzeitschrift/Zeitung › SCORING: Zeitschriftenaufsatz › Forschung › Begutachtung
Harvard
APA
Vancouver
Bibtex
}
RIS
TY - JOUR
T1 - Initial Clinical Experience Using the Low-Profile Altura Endograft System With Double D-Shaped Proximal Stents for Endovascular Aneurysm Repair
AU - Krievins, Dainis
AU - Krämer, Albrecht
AU - Savlovskis, Janis
AU - Oszkinis, Georgij
AU - Debus, E Sebastian
AU - Oberhuber, Alexander
AU - Zarins, Christopher K
PY - 2018/6
Y1 - 2018/6
N2 - PURPOSE: To report the initial clinical results of endovascular aneurysm repair (EVAR) using the low-profile (14-F) Altura Endograft System, which features a double "D-shaped" stent design with suprarenal fixation and modular iliac components that are deployed from distal to proximal.METHODS: From 2011 to 2015, 90 patients (mean age 72.8±8.3 years; 79 men) with abdominal aortic aneurysm (AAA; mean diameter 53.8±5.7 mm) were treated at 10 clinical sites in 2 prospective, controlled clinical studies using the Altura endograft. Outcomes evaluated included mortality, major adverse events (MAEs: all-cause death, stroke, paraplegia, myocardial infarction, respiratory failure, bowel ischemia, and blood loss ≥1000 mL), and clinical success (freedom from procedure-related death, type I/III endoleak, migration, thrombosis, and reintervention).RESULTS: Endografts were successfully implanted in 89 (99%) patients; the single failure was due to delivery system malfunction before insertion in the early-generation device. One (1%) patient died and 4 patients underwent reinterventions (1 type I endoleak, 2 iliac limb stenoses, and 1 endograft occlusion) within the first 30 days. During a median follow-up of 12.5 months (range 11.5-50.9), there were no aneurysm ruptures, surgical conversions, or AAA-related deaths. The cumulative MAE rates were 3% (3/89) at 6 months and 7% (6/89) at 1 year. Two patients underwent coil embolization of type II endoleaks at 6.5 months and 2.2 years, respectively. Clinical success was 94% (84/89) at 30 days, 98% (85/87) at 6 months, and 99% (82/83) at 1 year.CONCLUSION: Early results suggest that properly selected AAA patients can be safely treated using the Altura Endograft System with favorable midterm outcome. Thus, further clinical investigation is warranted to evaluate the role of this device in the treatment of AAA.
AB - PURPOSE: To report the initial clinical results of endovascular aneurysm repair (EVAR) using the low-profile (14-F) Altura Endograft System, which features a double "D-shaped" stent design with suprarenal fixation and modular iliac components that are deployed from distal to proximal.METHODS: From 2011 to 2015, 90 patients (mean age 72.8±8.3 years; 79 men) with abdominal aortic aneurysm (AAA; mean diameter 53.8±5.7 mm) were treated at 10 clinical sites in 2 prospective, controlled clinical studies using the Altura endograft. Outcomes evaluated included mortality, major adverse events (MAEs: all-cause death, stroke, paraplegia, myocardial infarction, respiratory failure, bowel ischemia, and blood loss ≥1000 mL), and clinical success (freedom from procedure-related death, type I/III endoleak, migration, thrombosis, and reintervention).RESULTS: Endografts were successfully implanted in 89 (99%) patients; the single failure was due to delivery system malfunction before insertion in the early-generation device. One (1%) patient died and 4 patients underwent reinterventions (1 type I endoleak, 2 iliac limb stenoses, and 1 endograft occlusion) within the first 30 days. During a median follow-up of 12.5 months (range 11.5-50.9), there were no aneurysm ruptures, surgical conversions, or AAA-related deaths. The cumulative MAE rates were 3% (3/89) at 6 months and 7% (6/89) at 1 year. Two patients underwent coil embolization of type II endoleaks at 6.5 months and 2.2 years, respectively. Clinical success was 94% (84/89) at 30 days, 98% (85/87) at 6 months, and 99% (82/83) at 1 year.CONCLUSION: Early results suggest that properly selected AAA patients can be safely treated using the Altura Endograft System with favorable midterm outcome. Thus, further clinical investigation is warranted to evaluate the role of this device in the treatment of AAA.
KW - Aged
KW - Aged, 80 and over
KW - Aortic Aneurysm, Abdominal/diagnostic imaging
KW - Blood Vessel Prosthesis
KW - Blood Vessel Prosthesis Implantation/adverse effects
KW - Endovascular Procedures/adverse effects
KW - Female
KW - Humans
KW - Male
KW - Middle Aged
KW - Patient Selection
KW - Postoperative Complications/etiology
KW - Prosthesis Design
KW - Prosthesis Failure
KW - Risk Assessment
KW - Risk Factors
KW - Stents
KW - Time Factors
KW - Treatment Outcome
U2 - 10.1177/1526602818771973
DO - 10.1177/1526602818771973
M3 - SCORING: Journal article
C2 - 29673298
VL - 25
SP - 379
EP - 386
JO - J ENDOVASC THER
JF - J ENDOVASC THER
SN - 1526-6028
IS - 3
ER -