Two-Hour Algorithm for Rapid Triage of Suspected Acute Myocardial Infarction Using a High-Sensitivity Cardiac Troponin I Assay

  • T Nestelberger
  • J Boeddinghaus
  • J Greenslade
  • WA Parsonage
  • M Than
  • D Wussler
  • P Lopez-Ayala
  • T Zimmermann
  • M Meier
  • V Troester
  • P Badertscher
  • Luca Koechlin
  • Karin Wildi
  • Mahnoor Anwar
  • Michael Freese
  • Dagmar Keller
  • Tobias Reichlin
  • Raphael Twerenbold
  • Louise Cullen
  • Christian Mueller
  • APACE Investigators

Abstract

BACKGROUND
We aimed to derive and externally validate a 0/2-h algorithm using the high-sensitivity cardiac troponin I (hs-cTnI)-Access assay.

METHODS
We enrolled patients presenting to the emergency department with symptoms suggestive of acute myocardial infarction (AMI) in 2 prospective diagnostic studies using central adjudication. Two independent cardiologists adjudicated the final diagnosis, including all available medical information including cardiac imaging. hs-cTnI-Access concentrations were measured at presentation and after 2 h in a blinded fashion.

RESULTS
AMI was the adjudicated final diagnosis in 164 of 1131 (14.5%) patients in the derivation cohort. Rule-out by the hs-cTnI-Access 0/2-h algorithm was defined as 0-h hs-cTnI-Access concentration <4 ng/L in patients with an onset of chest pain >3 h (direct rule-out) or a 0-h hs-cTnI-Access concentration <5 ng/L and an absolute change within 2 h <5 ng/L in all other patients. Derived thresholds for rule-in were a 0-h hs-cTnI-Access concentration ≥50 ng/L (direct rule-in) or an absolute change within 2 h ≥20 ng/L. In the derivation cohort, these cutoffs ruled out 55% of patients with a negative predictive value (NPV) of 99.8% (95% CI, 99.3–100) and sensitivity of 99.4% (95% CI, 96.5–99.9), and ruled in 30% of patients with a positive predictive value (PPV) of 73% (95% CI, 66.1–79). In the validation cohort, AMI was the adjudicated final diagnosis in 88 of 1280 (6.9%) patients. These cutoffs ruled out 77.9% of patients with an NPV of 99.8% (95% CI, 99.3–100) and sensitivity of 97.7% (95% CI, 92.0–99.7), and ruled in 5.8% of patients with a PPV of 77% (95% CI, 65.8–86) in the validation cohort.

CONCLUSIONS
Safety and efficacy of the l hs-cTnI-Access 0/2-h algorithm for triage toward rule-out or rule-in of AMI are very high.

Bibliografische Daten

OriginalspracheEnglisch
ISSN0009-9147
DOIs
StatusVeröffentlicht - 09.2019
Extern publiziertJa