Second-generation cryoballoon-based pulmonary vein isolation: Lessons from a five-year follow-up
Standard
Second-generation cryoballoon-based pulmonary vein isolation: Lessons from a five-year follow-up. / Heeger, Christian-Hendrik; Subin, Behnam; Wissner, Erik; Fink, Thomas; Mathew, Shibu; Maurer, Tilman; Lemes, Christine; Rillig, Andreas; Wohlmuth, Peter; Reissmann, Bruno; Tilz, Roland Richard; Ouyang, Feifan; Kuck, Karl-Heinz; Metzner, Andreas.
in: INT J CARDIOL, Jahrgang 312, 01.08.2020, S. 73-80.Publikationen: SCORING: Beitrag in Fachzeitschrift/Zeitung › SCORING: Zeitschriftenaufsatz › Forschung › Begutachtung
Harvard
APA
Vancouver
Bibtex
}
RIS
TY - JOUR
T1 - Second-generation cryoballoon-based pulmonary vein isolation: Lessons from a five-year follow-up
AU - Heeger, Christian-Hendrik
AU - Subin, Behnam
AU - Wissner, Erik
AU - Fink, Thomas
AU - Mathew, Shibu
AU - Maurer, Tilman
AU - Lemes, Christine
AU - Rillig, Andreas
AU - Wohlmuth, Peter
AU - Reissmann, Bruno
AU - Tilz, Roland Richard
AU - Ouyang, Feifan
AU - Kuck, Karl-Heinz
AU - Metzner, Andreas
N1 - Copyright © 2020 Elsevier B.V. All rights reserved.
PY - 2020/8/1
Y1 - 2020/8/1
N2 - BACKGROUND: Pulmonary vein isolation (PVI) using the second-generation cryoballoon (CB2) for atrial fibrillation (AF) treatment has demonstrated encouraging procedural success rates and mid-term results. However, long-term follow-up data on outcome is sparse. The current study is reporting on five-year clinical outcomes following CB2-based PVI in patients with paroxysmal AF (PAF) and persistent AF (PersAF).METHODS: A total of 139 patients underwent index CB2-based PVI patients (PAF: n = 105, 76% and PersAF: n = 34, 34%) in two electrophysiology centers. Freeze-cycle duration was 240 s. After successful PVI a bonus freeze-cycle of the same duration was applied in the first 71 patients while the bonus-freeze was omitted in 68 following patients. Three patients (2.2%) were lost to follow-up.RESULTS: After a median follow-up duration of 60 months (interquartile range: 46, 72 months) 74/136 (54.4%) patients remained in stable sinus rhythm (PAF: 62/104, 59.6%; PersAF: 15/32, 46.9%). Significant differences were observed concerning 5-year clinical outcome between PAF and PersAF patients (p = 0.0315). After a mean of 1.32 ± 0.6 procedures (2nd and 3rd procedure by radiofrequency ablation) and a median follow-up duration of 60 (37, 68) months 90/136 (66.2%) patients remained in stable sinus rhythm (PAF: 72/104 (69.2%), PersAF: 21/32 (65.6%), p = 0.0276). For the comparison of bonus-freeze vs no-bonus-freeze protocols no differences were observed (53.5% vs 57.4%, p = 0.650).CONCLUSIONS: The five-year single-procedure success rate for CB2-based PVI was 54.4% and increased to 66.2% following repeat RF-based procedures. No differences were detected comparing bonus-freeze and no-bonus-freeze protocols.
AB - BACKGROUND: Pulmonary vein isolation (PVI) using the second-generation cryoballoon (CB2) for atrial fibrillation (AF) treatment has demonstrated encouraging procedural success rates and mid-term results. However, long-term follow-up data on outcome is sparse. The current study is reporting on five-year clinical outcomes following CB2-based PVI in patients with paroxysmal AF (PAF) and persistent AF (PersAF).METHODS: A total of 139 patients underwent index CB2-based PVI patients (PAF: n = 105, 76% and PersAF: n = 34, 34%) in two electrophysiology centers. Freeze-cycle duration was 240 s. After successful PVI a bonus freeze-cycle of the same duration was applied in the first 71 patients while the bonus-freeze was omitted in 68 following patients. Three patients (2.2%) were lost to follow-up.RESULTS: After a median follow-up duration of 60 months (interquartile range: 46, 72 months) 74/136 (54.4%) patients remained in stable sinus rhythm (PAF: 62/104, 59.6%; PersAF: 15/32, 46.9%). Significant differences were observed concerning 5-year clinical outcome between PAF and PersAF patients (p = 0.0315). After a mean of 1.32 ± 0.6 procedures (2nd and 3rd procedure by radiofrequency ablation) and a median follow-up duration of 60 (37, 68) months 90/136 (66.2%) patients remained in stable sinus rhythm (PAF: 72/104 (69.2%), PersAF: 21/32 (65.6%), p = 0.0276). For the comparison of bonus-freeze vs no-bonus-freeze protocols no differences were observed (53.5% vs 57.4%, p = 0.650).CONCLUSIONS: The five-year single-procedure success rate for CB2-based PVI was 54.4% and increased to 66.2% following repeat RF-based procedures. No differences were detected comparing bonus-freeze and no-bonus-freeze protocols.
KW - Atrial Fibrillation/diagnostic imaging
KW - Catheter Ablation
KW - Cryosurgery
KW - Follow-Up Studies
KW - Humans
KW - Pulmonary Veins/diagnostic imaging
KW - Recurrence
KW - Time Factors
KW - Treatment Outcome
U2 - 10.1016/j.ijcard.2020.03.062
DO - 10.1016/j.ijcard.2020.03.062
M3 - SCORING: Journal article
C2 - 32241572
VL - 312
SP - 73
EP - 80
JO - INT J CARDIOL
JF - INT J CARDIOL
SN - 0167-5273
ER -