Prospective multi-centre study on a composite ceramic femoral component in total knee arthroplasty: Five-year clinical and radiological outcomes

  • Philipp Bergschmidt
  • Rainer Bader
  • Dirk Ganzer
  • Christian Hauzeur
  • Christoph H Lohmann
  • Alexander Krüger
  • Wolfgang Ruether
  • Domenico Tigani
  • Nicola Rani
  • José Luis Esteve
  • Fernando Lopez Prats
  • Claudio Zorzi
  • Vincenzo Madonna
  • Stefano Rigotti
  • Francesco Benazzo
  • Stefano Marco Paolo Rossi
  • Wolfram Mittelmeier

Abstract

BACKGROUND: Enhanced wear resistance of ceramics in general and improved mechanical characteristics of composite ceramics in terms of strength and resistance meet the demands for application in TKA. The aim of this prospective international multi-centre study was to evaluate the 5-year clinical and radiological outcomes of an unconstrained TKA with a composite ceramic femoral component.

METHODS: A total of 107 patients (109 knees) underwent TKA with the MULTIGEN-PLUS Ceramic Knee at seven centres in three European countries. Clinical and radiological assessments were performed preoperatively and postoperatively at 3, 12, 24 and 60 months, using HSS, WOMAC, SF-36 and standardised radiographs.

RESULTS: Mean HSS and WOMAC increased significantly from 55.1±11.5 (21-83) and 48.1±16.6 (3-90) preoperatively to 85.6±9.6 (49-98) and 73.3±20.4 (17-100) at 60 months. Mean SF-36 showed significant improvements in patients' quality of life (49.1±17.6 (12-96) preoperatively versus 67.7±23.1 (12-100) at 60 months). Non-progressive radiolucent lines (<1 mm) were observed around the femoral component in four cases. Neither implant migration nor loosening were registered. Kaplan-Meier survivorship was 96.0% at 60 months (92.1-100%, CI 95%).

CONCLUSIONS: Five-year implant survival rate of the ceramic knee is comparable to other metallic and ceramic unconstrained TKA systems. Although the assessment of long-term implant survivorship is still pending, the ceramic implants represent a promising solution for patients with allergies against metallic components and furthermore for the general osteoarthritis population due to enhanced wear resistance.

Bibliografische Daten

OriginalspracheEnglisch
ISSN0968-0160
DOIs
StatusVeröffentlicht - 06.2015
PubMed 25805085