Improved procedural efficacy of pulmonary vein isolation using the novel second-generation cryoballoon

  • Alexander Fürnkranz
  • Stefano Bordignon
  • Boris Schmidt
  • Melanie Gunawardene
  • Britta Schulte-Hahn
  • Verena Urban
  • Frank Bode
  • Bernd Nowak
  • Julian K R Chun

Abstract

INTRODUCTION: The cryoballoon technology has the potential to isolate a pulmonary vein (PV) with a single energy application. However, using the first-generation cryoballoon (CB-1G) repeated freezing or additional focal ablation is often necessary. The novel second-generation cryoballoon (CB-2G) features a widened zone of optimal cooling comprising the whole frontal hemisphere. The aim of this study was to investigate the impact of the novel design on procedural efficacy of cryoballoon PV isolation (CB-PVI).

METHODS AND RESULTS: Single transseptal CB-PVI using an endoluminal spiral mapping catheter was performed in 60 consecutive patients (CB-1G, 28 mm, 300 seconds application time: 30 patients; CB-2G, 28 mm, 240 seconds application time: 30 patients). When compared to the CB-1G, using the CB-2G increased single-shot PVI rate from 51% to 84% (P < 0.001) and decreased procedure duration (128 ± 27 vs 98 ± 30 minutes; P < 0.001), and fluoroscopy exposure time (19.5 ± 7.4 vs 13.4 ± 5.3 min; P = 0.001). Effective CB-2G PVI could be performed with increased real-time PVI visualization rate (49% vs 76%; P < 0.001). Time to PVI (T(PVI)) was shorter in the CB-2G group (79 ± 60 vs. 52 ± 36 seconds; P = 0.049). Procedure-related complications occurred in 2 patients in the CB-1G group and 1 patient in the CB-2G group.

CONCLUSIONS: The CB-2G significantly improved procedural efficacy compared to the CB-1G and provided reliable T(PVI) measurement. T(PVI) may be used to adjust application time and number individually in future studies. Final conclusions regarding the safety profile of the CB-2G requires additional research.

Bibliografische Daten

OriginalspracheEnglisch
ISSN1045-3873
DOIs
StatusVeröffentlicht - 01.05.2013
Extern publiziertJa
PubMed 23398599