Comparison of three different serum-free light-chain assays-implications on diagnostic and therapeutic monitoring of multiple myeloma

Beteiligte Einrichtungen

Abstract

The measurement of serum-free light chains (FLC) is standard of care in the diagnosis and management of multiple myeloma (MM). The revised international myeloma working group (IMWG) implemented the involved FLC/noninvolved FLC (iFLC/niFLC) ratio as a biomarker for MM requiring treatment. Recently, a new definition of high-risk smoldering MM (SMM) including iFLC/niFLC ratio was published. These recommendations were solely based on a single assay method (Freelite assay). Today, two additional assays, N Latex FLC and ELISA-based Sebia FLC, are available. Here, we report on a single-center-study comparing results of all three different assays for FLC correlation and its potential implications for diagnostic and clinical use. In total, 187 samples from 47 MM patients were examined, and determination of FLC was performed. Comparison analyses showed similar FLC results for Sebia FLC and N Latex FLC assay with markedly lower absolute values for κ/λ ratio compared with Freelite. Values of λ FLC exhibited high variability. The ratio of iFLC/niFLC showed significant discrepancies among these assays. Our data demonstrate that the three available assays may result in markedly discrepant results, and should not be used interchangeably to monitor patients. Furthermore, modifications of the assay-specific diagnostic (iFLC/niFLC) thresholds for SMM and MM are recommended.

Bibliografische Daten

OriginalspracheEnglisch
ISSN2044-5385
DOIs
StatusVeröffentlicht - 09.01.2020
PubMed 31915365