Analyse der Passgenauigkeit der Prothesenköpfe der Ascension® PyroCarbon-Fingermittelgelenksprothese

Standard

Analyse der Passgenauigkeit der Prothesenköpfe der Ascension® PyroCarbon-Fingermittelgelenksprothese. / Ries, C; Zhang, W; Müller, L P; Hohendorff, B.

in: HANDCHIR MIKROCHIR P, Jahrgang 47, Nr. 5, 10.2015, S. 322-7.

Publikationen: SCORING: Beitrag in Fachzeitschrift/ZeitungSCORING: ZeitschriftenaufsatzForschungBegutachtung

Harvard

APA

Vancouver

Bibtex

@article{bc324639b20d43d785c0caa3fed1ce3b,
title = "Analyse der Passgenauigkeit der Prothesenk{\"o}pfe der Ascension{\textregistered} PyroCarbon-Fingermittelgelenksprothese",
abstract = "BACKGROUND: The Ascension pyroCarbon proximal interphalangeal (PIP) total joint is available in 4 different sizes, and ideally, the prosthesis head will be flush with the bone. Fit of the Ascension pyrocarbon PIP joint prosthesis has not yet been investigated.MATERIALS AND METHODS: The components of the Ascension pyrocarbon PIP total joint were inserted in 287 phalanges of human cadaver specimens. The distances from the edge of the component head to the edge of the bone were electronically measured radially, ulnarly, dorsally and palmarly on radiographs in posterior-anterior and lateral views.RESULTS: Only one finger had a precise fit of the heads of both, the proximal and corresponding distal component in relation to the bone. Only in 17 (5 proximal, 12 distal) prosthesis components the head did the bone fit on all sides. Overall, the proximal component head tends to be too large, while the distal component head tends to be too small.CONCLUSION: With occasional exceptions, the proximal and distal component heads of the Ascension pyrocarbon PIP total joint do not accomodate the dimensions of finger phalanges.",
keywords = "Aged, Aged, 80 and over, Arthroplasty, Replacement/methods, Carbon, Female, Finger Joint/diagnostic imaging, Humans, Male, Middle Aged, Prosthesis Design, Prosthesis Failure, Prosthesis Fitting, Radiography",
author = "C Ries and W Zhang and M{\"u}ller, {L P} and B Hohendorff",
note = "{\textcopyright} Georg Thieme Verlag KG Stuttgart · New York.",
year = "2015",
month = oct,
doi = "10.1055/s-0035-1555867",
language = "Deutsch",
volume = "47",
pages = "322--7",
journal = "HANDCHIR MIKROCHIR P",
issn = "0722-1819",
publisher = "Georg Thieme Verlag KG",
number = "5",

}

RIS

TY - JOUR

T1 - Analyse der Passgenauigkeit der Prothesenköpfe der Ascension® PyroCarbon-Fingermittelgelenksprothese

AU - Ries, C

AU - Zhang, W

AU - Müller, L P

AU - Hohendorff, B

N1 - © Georg Thieme Verlag KG Stuttgart · New York.

PY - 2015/10

Y1 - 2015/10

N2 - BACKGROUND: The Ascension pyroCarbon proximal interphalangeal (PIP) total joint is available in 4 different sizes, and ideally, the prosthesis head will be flush with the bone. Fit of the Ascension pyrocarbon PIP joint prosthesis has not yet been investigated.MATERIALS AND METHODS: The components of the Ascension pyrocarbon PIP total joint were inserted in 287 phalanges of human cadaver specimens. The distances from the edge of the component head to the edge of the bone were electronically measured radially, ulnarly, dorsally and palmarly on radiographs in posterior-anterior and lateral views.RESULTS: Only one finger had a precise fit of the heads of both, the proximal and corresponding distal component in relation to the bone. Only in 17 (5 proximal, 12 distal) prosthesis components the head did the bone fit on all sides. Overall, the proximal component head tends to be too large, while the distal component head tends to be too small.CONCLUSION: With occasional exceptions, the proximal and distal component heads of the Ascension pyrocarbon PIP total joint do not accomodate the dimensions of finger phalanges.

AB - BACKGROUND: The Ascension pyroCarbon proximal interphalangeal (PIP) total joint is available in 4 different sizes, and ideally, the prosthesis head will be flush with the bone. Fit of the Ascension pyrocarbon PIP joint prosthesis has not yet been investigated.MATERIALS AND METHODS: The components of the Ascension pyrocarbon PIP total joint were inserted in 287 phalanges of human cadaver specimens. The distances from the edge of the component head to the edge of the bone were electronically measured radially, ulnarly, dorsally and palmarly on radiographs in posterior-anterior and lateral views.RESULTS: Only one finger had a precise fit of the heads of both, the proximal and corresponding distal component in relation to the bone. Only in 17 (5 proximal, 12 distal) prosthesis components the head did the bone fit on all sides. Overall, the proximal component head tends to be too large, while the distal component head tends to be too small.CONCLUSION: With occasional exceptions, the proximal and distal component heads of the Ascension pyrocarbon PIP total joint do not accomodate the dimensions of finger phalanges.

KW - Aged

KW - Aged, 80 and over

KW - Arthroplasty, Replacement/methods

KW - Carbon

KW - Female

KW - Finger Joint/diagnostic imaging

KW - Humans

KW - Male

KW - Middle Aged

KW - Prosthesis Design

KW - Prosthesis Failure

KW - Prosthesis Fitting

KW - Radiography

U2 - 10.1055/s-0035-1555867

DO - 10.1055/s-0035-1555867

M3 - SCORING: Zeitschriftenaufsatz

C2 - 26221758

VL - 47

SP - 322

EP - 327

JO - HANDCHIR MIKROCHIR P

JF - HANDCHIR MIKROCHIR P

SN - 0722-1819

IS - 5

ER -