A phantom approach to interscanner comparability of computed tomographic brain perfusion parameters

  • Christian Mathys
  • Konrad Rybacki
  • Hans-Jörg Wittsack
  • Rotem Shlomo Lanzman
  • Falk Roland Miese
  • Stephan Macht
  • Sven Eicker
  • Gerd Meyer zu Hörste
  • Gerald Antoch
  • Bernd Turowski

Beteiligte Einrichtungen

Abstract

OBJECTIVE: This study aimed to create a phantom, which allows reproducible computed tomography perfusion experiments, and to identify the influence of contrast bolus configuration, scan parameters, and scanner hardware on the calculation of perfusion parameters.

METHODS: A discoid perfusion phantom with centrifugally directed flow was constructed. Brain parenchyma was simulated by inert polyoxymethylene spheres. Repeated measurements were performed with variations of the above-mentioned factors, and their effects on perfusion results were analyzed.

RESULTS: Calculated flow values measured during experiments were reproducible and showed good correlation with the true flow (R = 0.995, P < 0.01). Tube voltage, injection rate of the contrast agent, the mathematical perfusion algorithm, and the hardware of the scanner hardware had a reproducible influence on calculated perfusion results.

CONCLUSIONS: In the long term, perfusion phantoms might be helpful in identifying hardware-specific and protocol-related factors of different computed tomography scanners to improve comparability of different scanners and scanning protocols.