Transpulmonary thermodilution using femoral indicator injection

Standard

Transpulmonary thermodilution using femoral indicator injection : a prospective trial in patients with a femoral and a jugular central venous catheter. / Saugel, Bernd; Umgelter, Andreas; Schuster, Tibor; Phillip, Veit; Schmid, Roland M; Huber, Wolfgang.

In: CRIT CARE, Vol. 14, No. 3, 01.01.2010, p. R95.

Research output: SCORING: Contribution to journalSCORING: Journal articleResearchpeer-review

Harvard

APA

Vancouver

Bibtex

@article{2000432396cf40ac8a2c6567fa4cddcd,
title = "Transpulmonary thermodilution using femoral indicator injection: a prospective trial in patients with a femoral and a jugular central venous catheter",
abstract = "INTRODUCTION: Advanced hemodynamic monitoring using transpulmonary thermodilution (TPTD) is established for measurement of cardiac index (CI), global end-diastolic volume index (GEDVI) and extra-vascular lung water index (EVLWI). TPTD requires indicator injection via a central venous catheter (usually placed via the jugular or subclavian vein). However, superior vena cava access is often not feasible due to the clinical situation. This study investigates the conformity of TPTD using femoral access.METHODS: This prospective study involved an 18-month trial at a medical intensive care unit at a university hospital. Twenty-four patients with both a superior and an inferior vena cava catheter at the same time were enrolled in the study.RESULTS: TPTD-variables were calculated from TPTD curves after injection of the indicator bolus via jugular access (TPTDjug) and femoral access (TPTDfem). GEDVIfem and GEDVIjug were significantly correlated (rm = 0.88; P < 0.001), but significantly different (1,034 +/- 275 vs. 793 +/- 180 mL/m2; P < 0.001). Bland-Altman analysis demonstrated a bias of +241 mL/m2 (limits of agreement: -9 and +491 mL/m2). GEDVIfem, CIfem and ideal body weight were independently associated with the bias (GEDVIfem-GEDVIjug). A correction formula of GEDVIjug after femoral TPTD, was calculated. EVLWIfem and EVLWIjug were significantly correlated (rm = 0.93; P < 0.001). Bland-Altman analysis revealed a bias of +0.83 mL/kg (limits of agreement: -2.61 and +4.28 mL/kg). Furthermore, CIfem and CIjug were significantly correlated (rm = 0.95; P < 0.001). Bland-Altman analysis demonstrated a bias of +0.29 L/min/m2 (limits of agreement -0.40 and +0.97 L/min/m2; percentage-error 16%).CONCLUSIONS: TPTD after femoral injection of the thermo-bolus provides precise data on GEDVI with a high correlation, but a self-evident significant bias related to the augmented TPTD-volume. After correction of GEDVIfem using a correction formula, GEDVIfem shows high predictive capabilities for GEDVIjug. Regarding CI and EVLWI, accurate TPTD-data is obtained using femoral access.",
keywords = "Aged, Aged, 80 and over, Catheterization, Central Venous, Female, Femoral Vein, Germany, Hemodynamics, Hospitals, University, Humans, Indicators and Reagents, Intensive Care Units, Male, Middle Aged, Monitoring, Physiologic, Prospective Studies, Thermodilution, Vena Cava, Inferior, Vena Cava, Superior",
author = "Bernd Saugel and Andreas Umgelter and Tibor Schuster and Veit Phillip and Schmid, {Roland M} and Wolfgang Huber",
year = "2010",
month = jan,
day = "1",
doi = "10.1186/cc9030",
language = "English",
volume = "14",
pages = "R95",
journal = "CRIT CARE",
issn = "1364-8535",
publisher = "Springer Science + Business Media",
number = "3",

}

RIS

TY - JOUR

T1 - Transpulmonary thermodilution using femoral indicator injection

T2 - a prospective trial in patients with a femoral and a jugular central venous catheter

AU - Saugel, Bernd

AU - Umgelter, Andreas

AU - Schuster, Tibor

AU - Phillip, Veit

AU - Schmid, Roland M

AU - Huber, Wolfgang

PY - 2010/1/1

Y1 - 2010/1/1

N2 - INTRODUCTION: Advanced hemodynamic monitoring using transpulmonary thermodilution (TPTD) is established for measurement of cardiac index (CI), global end-diastolic volume index (GEDVI) and extra-vascular lung water index (EVLWI). TPTD requires indicator injection via a central venous catheter (usually placed via the jugular or subclavian vein). However, superior vena cava access is often not feasible due to the clinical situation. This study investigates the conformity of TPTD using femoral access.METHODS: This prospective study involved an 18-month trial at a medical intensive care unit at a university hospital. Twenty-four patients with both a superior and an inferior vena cava catheter at the same time were enrolled in the study.RESULTS: TPTD-variables were calculated from TPTD curves after injection of the indicator bolus via jugular access (TPTDjug) and femoral access (TPTDfem). GEDVIfem and GEDVIjug were significantly correlated (rm = 0.88; P < 0.001), but significantly different (1,034 +/- 275 vs. 793 +/- 180 mL/m2; P < 0.001). Bland-Altman analysis demonstrated a bias of +241 mL/m2 (limits of agreement: -9 and +491 mL/m2). GEDVIfem, CIfem and ideal body weight were independently associated with the bias (GEDVIfem-GEDVIjug). A correction formula of GEDVIjug after femoral TPTD, was calculated. EVLWIfem and EVLWIjug were significantly correlated (rm = 0.93; P < 0.001). Bland-Altman analysis revealed a bias of +0.83 mL/kg (limits of agreement: -2.61 and +4.28 mL/kg). Furthermore, CIfem and CIjug were significantly correlated (rm = 0.95; P < 0.001). Bland-Altman analysis demonstrated a bias of +0.29 L/min/m2 (limits of agreement -0.40 and +0.97 L/min/m2; percentage-error 16%).CONCLUSIONS: TPTD after femoral injection of the thermo-bolus provides precise data on GEDVI with a high correlation, but a self-evident significant bias related to the augmented TPTD-volume. After correction of GEDVIfem using a correction formula, GEDVIfem shows high predictive capabilities for GEDVIjug. Regarding CI and EVLWI, accurate TPTD-data is obtained using femoral access.

AB - INTRODUCTION: Advanced hemodynamic monitoring using transpulmonary thermodilution (TPTD) is established for measurement of cardiac index (CI), global end-diastolic volume index (GEDVI) and extra-vascular lung water index (EVLWI). TPTD requires indicator injection via a central venous catheter (usually placed via the jugular or subclavian vein). However, superior vena cava access is often not feasible due to the clinical situation. This study investigates the conformity of TPTD using femoral access.METHODS: This prospective study involved an 18-month trial at a medical intensive care unit at a university hospital. Twenty-four patients with both a superior and an inferior vena cava catheter at the same time were enrolled in the study.RESULTS: TPTD-variables were calculated from TPTD curves after injection of the indicator bolus via jugular access (TPTDjug) and femoral access (TPTDfem). GEDVIfem and GEDVIjug were significantly correlated (rm = 0.88; P < 0.001), but significantly different (1,034 +/- 275 vs. 793 +/- 180 mL/m2; P < 0.001). Bland-Altman analysis demonstrated a bias of +241 mL/m2 (limits of agreement: -9 and +491 mL/m2). GEDVIfem, CIfem and ideal body weight were independently associated with the bias (GEDVIfem-GEDVIjug). A correction formula of GEDVIjug after femoral TPTD, was calculated. EVLWIfem and EVLWIjug were significantly correlated (rm = 0.93; P < 0.001). Bland-Altman analysis revealed a bias of +0.83 mL/kg (limits of agreement: -2.61 and +4.28 mL/kg). Furthermore, CIfem and CIjug were significantly correlated (rm = 0.95; P < 0.001). Bland-Altman analysis demonstrated a bias of +0.29 L/min/m2 (limits of agreement -0.40 and +0.97 L/min/m2; percentage-error 16%).CONCLUSIONS: TPTD after femoral injection of the thermo-bolus provides precise data on GEDVI with a high correlation, but a self-evident significant bias related to the augmented TPTD-volume. After correction of GEDVIfem using a correction formula, GEDVIfem shows high predictive capabilities for GEDVIjug. Regarding CI and EVLWI, accurate TPTD-data is obtained using femoral access.

KW - Aged

KW - Aged, 80 and over

KW - Catheterization, Central Venous

KW - Female

KW - Femoral Vein

KW - Germany

KW - Hemodynamics

KW - Hospitals, University

KW - Humans

KW - Indicators and Reagents

KW - Intensive Care Units

KW - Male

KW - Middle Aged

KW - Monitoring, Physiologic

KW - Prospective Studies

KW - Thermodilution