Transcatheter valve-in-valve implantation versus redo surgical aortic valve replacement in patients with failed aortic bioprostheses

Standard

Transcatheter valve-in-valve implantation versus redo surgical aortic valve replacement in patients with failed aortic bioprostheses. / Silaschi, Miriam; Wendler, Olaf; Seiffert, Moritz; Castro, Liesa; Lubos, Edith; Schirmer, Johannes; Blankenberg, Stefan; Reichenspurner, Hermann; Schäfer, Ulrich; Treede, Hendrik; MacCarthy, Philip; Conradi, Lenard.

In: INTERACT CARDIOV TH, Vol. 24, No. 1, 01.2017, p. 63-70.

Research output: SCORING: Contribution to journalSCORING: Journal articleResearchpeer-review

Harvard

APA

Vancouver

Bibtex

@article{323fe058f8364aa18e352fa5ef0d6d0a,
title = "Transcatheter valve-in-valve implantation versus redo surgical aortic valve replacement in patients with failed aortic bioprostheses",
abstract = "OBJECTIVES: Transcatheter aortic valve-in-valve implantation (ViV) is a new treatment for failing bioprostheses (BP) in patients with high surgical risk. However, comparative data, using standard repeat surgical aortic valve replacement (redo-SAVR), are scarce. We compared outcomes after ViV with those after conventional redo-SAVR in two European centres with established interventional programmes.METHODS: In-hospital databases were retrospectively screened for patients ≥60 years, treated for failing aortic BP. Cases of infective endocarditis or combined procedures were excluded. End-points were adjudicated according to the Valve Academic Research Consortium (VARC-2) criteria.RESULTS: From 2002 to 2015, 130 patients were treated (ViV: n = 71, redo-SAVR: n = 59). Age and logistic EuroSCORE I scores were higher with ViV (78.6 ± 7.5 vs 72.9 ± 6.6 years, P < 0.01; 25.1 ± 18.9 vs 16.8 ± 9.3%, P < 0.01). The 30-day mortality rate was not significantly different (4.2 and 5.1%, respectively) (P = 1.0). Device success was achieved in 52.1% (ViV) and 91.5% (P < 0.01). No stroke was observed after ViV but in 3.4% after redo-SAVR (P = 0.2). Intensive care stay was longer after redo-SAVR (3.4 ± 2.9 vs 2.0 ± 1.8 days, P < 0.01). Mean transvalvular gradients were higher post-ViV (19.7 ± 7.7 vs12.2 ± 5.7 mmHg, P < 0.01), whereas the rate of permanent pacemaker implantation was lower (9.9 vs 25.4%, P < 0.01). Survival rates at 90 and180 days were 94.2 and 92.3% vs 92.8 and 92.8% (P = 0.87), respectively.CONCLUSIONS: Despite a higher risk profile in the ViV group, early mortality rates were not different compared with those of surgery. Although ViV resulted in elevated transvalvular gradients and therefore a lower rate of device success, mortality rates were similar to those with redo-SAVR. At present, both techniques serve as complementary approaches, and allow individualized patient care with excellent outcomes.",
keywords = "Aged, Aged, 80 and over, Aortic Valve Stenosis/surgery, Bioprosthesis, Cohort Studies, Female, Heart Valve Prosthesis, Humans, Male, Reoperation, Retrospective Studies, Transcatheter Aortic Valve Replacement/methods, Treatment Outcome",
author = "Miriam Silaschi and Olaf Wendler and Moritz Seiffert and Liesa Castro and Edith Lubos and Johannes Schirmer and Stefan Blankenberg and Hermann Reichenspurner and Ulrich Sch{\"a}fer and Hendrik Treede and Philip MacCarthy and Lenard Conradi",
note = "{\textcopyright} The Author 2016. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.",
year = "2017",
month = jan,
doi = "10.1093/icvts/ivw300",
language = "English",
volume = "24",
pages = "63--70",
journal = "INTERACT CARDIOV TH",
issn = "1569-9293",
publisher = "European Association for Cardio-Thoracic Surgery",
number = "1",

}

RIS

TY - JOUR

T1 - Transcatheter valve-in-valve implantation versus redo surgical aortic valve replacement in patients with failed aortic bioprostheses

AU - Silaschi, Miriam

AU - Wendler, Olaf

AU - Seiffert, Moritz

AU - Castro, Liesa

AU - Lubos, Edith

AU - Schirmer, Johannes

AU - Blankenberg, Stefan

AU - Reichenspurner, Hermann

AU - Schäfer, Ulrich

AU - Treede, Hendrik

AU - MacCarthy, Philip

AU - Conradi, Lenard

N1 - © The Author 2016. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.

PY - 2017/1

Y1 - 2017/1

N2 - OBJECTIVES: Transcatheter aortic valve-in-valve implantation (ViV) is a new treatment for failing bioprostheses (BP) in patients with high surgical risk. However, comparative data, using standard repeat surgical aortic valve replacement (redo-SAVR), are scarce. We compared outcomes after ViV with those after conventional redo-SAVR in two European centres with established interventional programmes.METHODS: In-hospital databases were retrospectively screened for patients ≥60 years, treated for failing aortic BP. Cases of infective endocarditis or combined procedures were excluded. End-points were adjudicated according to the Valve Academic Research Consortium (VARC-2) criteria.RESULTS: From 2002 to 2015, 130 patients were treated (ViV: n = 71, redo-SAVR: n = 59). Age and logistic EuroSCORE I scores were higher with ViV (78.6 ± 7.5 vs 72.9 ± 6.6 years, P < 0.01; 25.1 ± 18.9 vs 16.8 ± 9.3%, P < 0.01). The 30-day mortality rate was not significantly different (4.2 and 5.1%, respectively) (P = 1.0). Device success was achieved in 52.1% (ViV) and 91.5% (P < 0.01). No stroke was observed after ViV but in 3.4% after redo-SAVR (P = 0.2). Intensive care stay was longer after redo-SAVR (3.4 ± 2.9 vs 2.0 ± 1.8 days, P < 0.01). Mean transvalvular gradients were higher post-ViV (19.7 ± 7.7 vs12.2 ± 5.7 mmHg, P < 0.01), whereas the rate of permanent pacemaker implantation was lower (9.9 vs 25.4%, P < 0.01). Survival rates at 90 and180 days were 94.2 and 92.3% vs 92.8 and 92.8% (P = 0.87), respectively.CONCLUSIONS: Despite a higher risk profile in the ViV group, early mortality rates were not different compared with those of surgery. Although ViV resulted in elevated transvalvular gradients and therefore a lower rate of device success, mortality rates were similar to those with redo-SAVR. At present, both techniques serve as complementary approaches, and allow individualized patient care with excellent outcomes.

AB - OBJECTIVES: Transcatheter aortic valve-in-valve implantation (ViV) is a new treatment for failing bioprostheses (BP) in patients with high surgical risk. However, comparative data, using standard repeat surgical aortic valve replacement (redo-SAVR), are scarce. We compared outcomes after ViV with those after conventional redo-SAVR in two European centres with established interventional programmes.METHODS: In-hospital databases were retrospectively screened for patients ≥60 years, treated for failing aortic BP. Cases of infective endocarditis or combined procedures were excluded. End-points were adjudicated according to the Valve Academic Research Consortium (VARC-2) criteria.RESULTS: From 2002 to 2015, 130 patients were treated (ViV: n = 71, redo-SAVR: n = 59). Age and logistic EuroSCORE I scores were higher with ViV (78.6 ± 7.5 vs 72.9 ± 6.6 years, P < 0.01; 25.1 ± 18.9 vs 16.8 ± 9.3%, P < 0.01). The 30-day mortality rate was not significantly different (4.2 and 5.1%, respectively) (P = 1.0). Device success was achieved in 52.1% (ViV) and 91.5% (P < 0.01). No stroke was observed after ViV but in 3.4% after redo-SAVR (P = 0.2). Intensive care stay was longer after redo-SAVR (3.4 ± 2.9 vs 2.0 ± 1.8 days, P < 0.01). Mean transvalvular gradients were higher post-ViV (19.7 ± 7.7 vs12.2 ± 5.7 mmHg, P < 0.01), whereas the rate of permanent pacemaker implantation was lower (9.9 vs 25.4%, P < 0.01). Survival rates at 90 and180 days were 94.2 and 92.3% vs 92.8 and 92.8% (P = 0.87), respectively.CONCLUSIONS: Despite a higher risk profile in the ViV group, early mortality rates were not different compared with those of surgery. Although ViV resulted in elevated transvalvular gradients and therefore a lower rate of device success, mortality rates were similar to those with redo-SAVR. At present, both techniques serve as complementary approaches, and allow individualized patient care with excellent outcomes.

KW - Aged

KW - Aged, 80 and over

KW - Aortic Valve Stenosis/surgery

KW - Bioprosthesis

KW - Cohort Studies

KW - Female

KW - Heart Valve Prosthesis

KW - Humans

KW - Male

KW - Reoperation

KW - Retrospective Studies

KW - Transcatheter Aortic Valve Replacement/methods

KW - Treatment Outcome

U2 - 10.1093/icvts/ivw300

DO - 10.1093/icvts/ivw300

M3 - SCORING: Journal article

C2 - 27624351

VL - 24

SP - 63

EP - 70

JO - INTERACT CARDIOV TH

JF - INTERACT CARDIOV TH

SN - 1569-9293

IS - 1

ER -