The novel Tenzing 7 delivery catheter designed to deliver intermediate catheters to the face of embolus without crossing: clinical performance predicted in anatomically challenging model
Standard
The novel Tenzing 7 delivery catheter designed to deliver intermediate catheters to the face of embolus without crossing: clinical performance predicted in anatomically challenging model. / Frölich, Andreas Maximilian; Kim, Warren; Stribrny, Knut; Jansen, Olav; Möhlenbruch, Markus; Szikora, Istvan; Wodarg, Fritz; Fiehler, Jens; Otto, Kim; Chou, Tony; Buhk, Jan Hendrik; English, Joey.
In: J NEUROINTERV SURG, Vol. 13, No. 8, 08.2021, p. 722-726.Research output: SCORING: Contribution to journal › SCORING: Journal article › Research › peer-review
Harvard
APA
Vancouver
Bibtex
}
RIS
TY - JOUR
T1 - The novel Tenzing 7 delivery catheter designed to deliver intermediate catheters to the face of embolus without crossing: clinical performance predicted in anatomically challenging model
AU - Frölich, Andreas Maximilian
AU - Kim, Warren
AU - Stribrny, Knut
AU - Jansen, Olav
AU - Möhlenbruch, Markus
AU - Szikora, Istvan
AU - Wodarg, Fritz
AU - Fiehler, Jens
AU - Otto, Kim
AU - Chou, Tony
AU - Buhk, Jan Hendrik
AU - English, Joey
N1 - © Author(s) (or their employer(s)) 2021. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ.
PY - 2021/8
Y1 - 2021/8
N2 - BACKGROUND: In large vessel occlusionstroke, navigation of aspiration catheters (AC) can be impeded by vessel tortuosity and the ophthalmic artery origin. A novel tapered delivery catheter was designed to facilitate delivery without disturbing the embolus. We assessed AC deliverability in vitro and validated the observations in a first-in-human experience.METHODS: In a vascular model with three challenging craniocervical scenarios, two commercial AC were advanced from the carotid to the middle cerebral artery by four neurointerventionalists. Catheter deliverability with standard microwire and microcatheter (MC) combinations and the Tenzing 7 (T7) Delivery Catheter (Route 92 Medical, San Mateo, CA) were compared. Operators rated aspects of catheter deliverability on a 5-point scale. Results were compared with device delivery patterns at a neurovascular center before and after clinical introduction of T7.RESULTS: In vitro, success rate and speed were higher with T7 (96%; mean 30±10 s) than with MC (65%; 72±47 s, p<0.001 each), with fewer interactions with the occlusion site (T7: 54% vs MC: 77%, p=0.004). T7 received superior ratings regarding carotid artery deflection (T7: 2, IQR1-3 vs MC: 3, IQR2-3, p<0.001), guide catheter pushback (T7: 2, IQR1-3 vs MC: 3, IQR3-3, p<0.001) and ophthalmic artery passage (T7: 1.5, IQR1-2 vs MC: 4, IQR3-5, p<0.001). Before introduction of T7 at a single center, delivery of AC to a large vessel occlusion without crossing was achieved in 15/123 cases (12%). With T7, this rate was 28/31 patients (90.3%).CONCLUSION: Compared with microcatheter and microwire combinations, T7 improves aspiration catheter delivery in vitro, minimizing the need to cross the occlusion. Initial clinical experience appears to validate the model's observations.
AB - BACKGROUND: In large vessel occlusionstroke, navigation of aspiration catheters (AC) can be impeded by vessel tortuosity and the ophthalmic artery origin. A novel tapered delivery catheter was designed to facilitate delivery without disturbing the embolus. We assessed AC deliverability in vitro and validated the observations in a first-in-human experience.METHODS: In a vascular model with three challenging craniocervical scenarios, two commercial AC were advanced from the carotid to the middle cerebral artery by four neurointerventionalists. Catheter deliverability with standard microwire and microcatheter (MC) combinations and the Tenzing 7 (T7) Delivery Catheter (Route 92 Medical, San Mateo, CA) were compared. Operators rated aspects of catheter deliverability on a 5-point scale. Results were compared with device delivery patterns at a neurovascular center before and after clinical introduction of T7.RESULTS: In vitro, success rate and speed were higher with T7 (96%; mean 30±10 s) than with MC (65%; 72±47 s, p<0.001 each), with fewer interactions with the occlusion site (T7: 54% vs MC: 77%, p=0.004). T7 received superior ratings regarding carotid artery deflection (T7: 2, IQR1-3 vs MC: 3, IQR2-3, p<0.001), guide catheter pushback (T7: 2, IQR1-3 vs MC: 3, IQR3-3, p<0.001) and ophthalmic artery passage (T7: 1.5, IQR1-2 vs MC: 4, IQR3-5, p<0.001). Before introduction of T7 at a single center, delivery of AC to a large vessel occlusion without crossing was achieved in 15/123 cases (12%). With T7, this rate was 28/31 patients (90.3%).CONCLUSION: Compared with microcatheter and microwire combinations, T7 improves aspiration catheter delivery in vitro, minimizing the need to cross the occlusion. Initial clinical experience appears to validate the model's observations.