Textile Aging Characterization on New Generations of Explanted Commercial Endoprostheses: A Preliminary Study

Standard

Textile Aging Characterization on New Generations of Explanted Commercial Endoprostheses: A Preliminary Study. / Bussmann, A; Heim, F; Delay, C; Girsowicz, E; Del Tatto, B; Dion, D; Papillon, J; Lejay, A; Chakfé, J; Thaveau, F; Georg, Y; Chakfé, N; Geprovas Collaborative Retrieval Program.

In: EUR J VASC ENDOVASC, Vol. 54, No. 3, 09.2017, p. 378-386.

Research output: SCORING: Contribution to journalSCORING: Journal articleResearchpeer-review

Harvard

Bussmann, A, Heim, F, Delay, C, Girsowicz, E, Del Tatto, B, Dion, D, Papillon, J, Lejay, A, Chakfé, J, Thaveau, F, Georg, Y, Chakfé, N & Geprovas Collaborative Retrieval Program 2017, 'Textile Aging Characterization on New Generations of Explanted Commercial Endoprostheses: A Preliminary Study', EUR J VASC ENDOVASC, vol. 54, no. 3, pp. 378-386. https://doi.org/10.1016/j.ejvs.2017.06.004

APA

Bussmann, A., Heim, F., Delay, C., Girsowicz, E., Del Tatto, B., Dion, D., Papillon, J., Lejay, A., Chakfé, J., Thaveau, F., Georg, Y., Chakfé, N., & Geprovas Collaborative Retrieval Program (2017). Textile Aging Characterization on New Generations of Explanted Commercial Endoprostheses: A Preliminary Study. EUR J VASC ENDOVASC, 54(3), 378-386. https://doi.org/10.1016/j.ejvs.2017.06.004

Vancouver

Bibtex

@article{ff9bbdba655c4730a6b94d21f6eb6c82,
title = "Textile Aging Characterization on New Generations of Explanted Commercial Endoprostheses: A Preliminary Study",
abstract = "OBJECTIVE/BACKGROUND: The goal of the present study was to assess the aging phenomena on second-generation textile endoprostheses (EPs) through explant analysis and to establish a preliminary classification of observed defects and material damages.METHODS: From January 2011 to June 2016 110 second- and recent-generation EPs were collected as a part of a European collaborative retrieval program. The analysis focused on the first 41 consecutive commercial EPs collected between 2011 and 2014 and made from polyethylene terephthalate. Explants were submitted to a standardized evaluation protocol, which included data recording, eye-naked evaluation, cleaning of organic remnants, and structural analysis under numerical optical microscopy. Observations were reported using a classification based on 15 features evaluating the fabric, the stitches between the fabric and the stents, and the stents. The total surface area of the holes within the fabric was measured.RESULTS: EPs were implanted for thoracic and abdominal procedures in 12 and 29 cases, respectively. The mean ± SD duration of implantation was 34 ± 26 months (range 2 days-8 years). Sixty-four percent of the samples demonstrated at least one defect caused by compression damage potentially related to the insertion of the EP within the delivery system, which promoted holes and tears. Ninety-five percent of all EPs demonstrated at least one type of abrasion on the stitches. The degradation of the stitches and the number of ruptures increased with duration of implantation. Stent degradation was rare and consisted of corrosion and rupture. Cumulated holed surface area increased with time and was measured up to 13.5 mm2.CONCLUSION: Various aging-related phenomena on commercial textile EPs were identified and classified. Main damaging mechanisms were related to compression and abrasion leading to tears and holes in the fabric and rupture of stitches.",
keywords = "Blood Vessel Prosthesis, Blood Vessel Prosthesis Implantation/adverse effects, Device Removal, Endovascular Procedures/adverse effects, Equipment Failure Analysis, Europe, Humans, Preliminary Data, Program Evaluation, Prosthesis Design, Prosthesis Failure, Risk Factors, Stents, Surface Properties, Time Factors, Treatment Outcome",
author = "A Bussmann and F Heim and C Delay and E Girsowicz and {Del Tatto}, B and D Dion and J Papillon and A Lejay and J Chakf{\'e} and F Thaveau and Y Georg and N Chakf{\'e} and {Geprovas Collaborative Retrieval Program} and Sebastian Debus",
note = "Copyright {\textcopyright} 2017 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.",
year = "2017",
month = sep,
doi = "10.1016/j.ejvs.2017.06.004",
language = "English",
volume = "54",
pages = "378--386",
journal = "EUR J VASC ENDOVASC",
issn = "1078-5884",
publisher = "W.B. Saunders Ltd",
number = "3",

}

RIS

TY - JOUR

T1 - Textile Aging Characterization on New Generations of Explanted Commercial Endoprostheses: A Preliminary Study

AU - Bussmann, A

AU - Heim, F

AU - Delay, C

AU - Girsowicz, E

AU - Del Tatto, B

AU - Dion, D

AU - Papillon, J

AU - Lejay, A

AU - Chakfé, J

AU - Thaveau, F

AU - Georg, Y

AU - Chakfé, N

AU - Geprovas Collaborative Retrieval Program

AU - Debus, Sebastian

N1 - Copyright © 2017 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

PY - 2017/9

Y1 - 2017/9

N2 - OBJECTIVE/BACKGROUND: The goal of the present study was to assess the aging phenomena on second-generation textile endoprostheses (EPs) through explant analysis and to establish a preliminary classification of observed defects and material damages.METHODS: From January 2011 to June 2016 110 second- and recent-generation EPs were collected as a part of a European collaborative retrieval program. The analysis focused on the first 41 consecutive commercial EPs collected between 2011 and 2014 and made from polyethylene terephthalate. Explants were submitted to a standardized evaluation protocol, which included data recording, eye-naked evaluation, cleaning of organic remnants, and structural analysis under numerical optical microscopy. Observations were reported using a classification based on 15 features evaluating the fabric, the stitches between the fabric and the stents, and the stents. The total surface area of the holes within the fabric was measured.RESULTS: EPs were implanted for thoracic and abdominal procedures in 12 and 29 cases, respectively. The mean ± SD duration of implantation was 34 ± 26 months (range 2 days-8 years). Sixty-four percent of the samples demonstrated at least one defect caused by compression damage potentially related to the insertion of the EP within the delivery system, which promoted holes and tears. Ninety-five percent of all EPs demonstrated at least one type of abrasion on the stitches. The degradation of the stitches and the number of ruptures increased with duration of implantation. Stent degradation was rare and consisted of corrosion and rupture. Cumulated holed surface area increased with time and was measured up to 13.5 mm2.CONCLUSION: Various aging-related phenomena on commercial textile EPs were identified and classified. Main damaging mechanisms were related to compression and abrasion leading to tears and holes in the fabric and rupture of stitches.

AB - OBJECTIVE/BACKGROUND: The goal of the present study was to assess the aging phenomena on second-generation textile endoprostheses (EPs) through explant analysis and to establish a preliminary classification of observed defects and material damages.METHODS: From January 2011 to June 2016 110 second- and recent-generation EPs were collected as a part of a European collaborative retrieval program. The analysis focused on the first 41 consecutive commercial EPs collected between 2011 and 2014 and made from polyethylene terephthalate. Explants were submitted to a standardized evaluation protocol, which included data recording, eye-naked evaluation, cleaning of organic remnants, and structural analysis under numerical optical microscopy. Observations were reported using a classification based on 15 features evaluating the fabric, the stitches between the fabric and the stents, and the stents. The total surface area of the holes within the fabric was measured.RESULTS: EPs were implanted for thoracic and abdominal procedures in 12 and 29 cases, respectively. The mean ± SD duration of implantation was 34 ± 26 months (range 2 days-8 years). Sixty-four percent of the samples demonstrated at least one defect caused by compression damage potentially related to the insertion of the EP within the delivery system, which promoted holes and tears. Ninety-five percent of all EPs demonstrated at least one type of abrasion on the stitches. The degradation of the stitches and the number of ruptures increased with duration of implantation. Stent degradation was rare and consisted of corrosion and rupture. Cumulated holed surface area increased with time and was measured up to 13.5 mm2.CONCLUSION: Various aging-related phenomena on commercial textile EPs were identified and classified. Main damaging mechanisms were related to compression and abrasion leading to tears and holes in the fabric and rupture of stitches.

KW - Blood Vessel Prosthesis

KW - Blood Vessel Prosthesis Implantation/adverse effects

KW - Device Removal

KW - Endovascular Procedures/adverse effects

KW - Equipment Failure Analysis

KW - Europe

KW - Humans

KW - Preliminary Data

KW - Program Evaluation

KW - Prosthesis Design

KW - Prosthesis Failure

KW - Risk Factors

KW - Stents

KW - Surface Properties

KW - Time Factors

KW - Treatment Outcome

U2 - 10.1016/j.ejvs.2017.06.004

DO - 10.1016/j.ejvs.2017.06.004

M3 - SCORING: Journal article

C2 - 28716448

VL - 54

SP - 378

EP - 386

JO - EUR J VASC ENDOVASC

JF - EUR J VASC ENDOVASC

SN - 1078-5884

IS - 3

ER -