Test method for mechanical properties of implantable catheters according to DIN 10555-3

Standard

Test method for mechanical properties of implantable catheters according to DIN 10555-3. / Busch, J D; Schröder, H; Sellenschloh, K; Adam, G; Ittrich, H; Huber, G.

In: J MECH BEHAV BIOMED, Vol. 82, 06.2018, p. 183-186.

Research output: SCORING: Contribution to journalSCORING: Journal articleResearchpeer-review

Harvard

APA

Vancouver

Bibtex

@article{3d6c8fe0c76348b3b387b6a96c31c5da,
title = "Test method for mechanical properties of implantable catheters according to DIN 10555-3",
abstract = "To enable causal analysis of port catheter failure, this study aimed to develop an experimental setup for uniaxial tensile tests that addresses the specific requirements of highly elastic medical catheters; and to quantify parameters of the catheters' mechanical competence with respect to effects of artificial aging. Segments of 6F-polyurethane catheters were tested in their native status, after chemical and after mechanical aging. Tension experiments were performed with a rate of 220 mm/min until catheter failure. Material behavior was analyzed based on load cell measurements of the universal test system and an additional optical distance registration. The Young's modulus, the ultimate stress and the ultimate strain were determined. Chemical aging significantly decreased Young's modulus (84%; p = 0.001) and ultimate stress (83%; p < 0.001), whereas mechanical aged samples demonstrated similar results for the Young's modulus (p = 0.772) and a non-significant rise of ultimate stress (13%; p = 0.128). Ultimate strain did not differ significantly regardless of the pretreatment. The results proof reliability, reproducibility and sensitivity to quantify artificial aging induced variations and also promise to detect deviations in material features caused by long-term clinical usage of catheters.",
keywords = "Journal Article",
author = "Busch, {J D} and H Schr{\"o}der and K Sellenschloh and G Adam and H Ittrich and G Huber",
note = "Copyright {\textcopyright} 2018 Elsevier Ltd. All rights reserved.",
year = "2018",
month = jun,
doi = "10.1016/j.jmbbm.2018.03.025",
language = "English",
volume = "82",
pages = "183--186",
journal = "J MECH BEHAV BIOMED",
issn = "1751-6161",
publisher = "Elsevier BV",

}

RIS

TY - JOUR

T1 - Test method for mechanical properties of implantable catheters according to DIN 10555-3

AU - Busch, J D

AU - Schröder, H

AU - Sellenschloh, K

AU - Adam, G

AU - Ittrich, H

AU - Huber, G

N1 - Copyright © 2018 Elsevier Ltd. All rights reserved.

PY - 2018/6

Y1 - 2018/6

N2 - To enable causal analysis of port catheter failure, this study aimed to develop an experimental setup for uniaxial tensile tests that addresses the specific requirements of highly elastic medical catheters; and to quantify parameters of the catheters' mechanical competence with respect to effects of artificial aging. Segments of 6F-polyurethane catheters were tested in their native status, after chemical and after mechanical aging. Tension experiments were performed with a rate of 220 mm/min until catheter failure. Material behavior was analyzed based on load cell measurements of the universal test system and an additional optical distance registration. The Young's modulus, the ultimate stress and the ultimate strain were determined. Chemical aging significantly decreased Young's modulus (84%; p = 0.001) and ultimate stress (83%; p < 0.001), whereas mechanical aged samples demonstrated similar results for the Young's modulus (p = 0.772) and a non-significant rise of ultimate stress (13%; p = 0.128). Ultimate strain did not differ significantly regardless of the pretreatment. The results proof reliability, reproducibility and sensitivity to quantify artificial aging induced variations and also promise to detect deviations in material features caused by long-term clinical usage of catheters.

AB - To enable causal analysis of port catheter failure, this study aimed to develop an experimental setup for uniaxial tensile tests that addresses the specific requirements of highly elastic medical catheters; and to quantify parameters of the catheters' mechanical competence with respect to effects of artificial aging. Segments of 6F-polyurethane catheters were tested in their native status, after chemical and after mechanical aging. Tension experiments were performed with a rate of 220 mm/min until catheter failure. Material behavior was analyzed based on load cell measurements of the universal test system and an additional optical distance registration. The Young's modulus, the ultimate stress and the ultimate strain were determined. Chemical aging significantly decreased Young's modulus (84%; p = 0.001) and ultimate stress (83%; p < 0.001), whereas mechanical aged samples demonstrated similar results for the Young's modulus (p = 0.772) and a non-significant rise of ultimate stress (13%; p = 0.128). Ultimate strain did not differ significantly regardless of the pretreatment. The results proof reliability, reproducibility and sensitivity to quantify artificial aging induced variations and also promise to detect deviations in material features caused by long-term clinical usage of catheters.

KW - Journal Article

U2 - 10.1016/j.jmbbm.2018.03.025

DO - 10.1016/j.jmbbm.2018.03.025

M3 - SCORING: Journal article

C2 - 29605811

VL - 82

SP - 183

EP - 186

JO - J MECH BEHAV BIOMED

JF - J MECH BEHAV BIOMED

SN - 1751-6161

ER -