BACKGROUND: Transcatheter aortic valve implantation (TAVI) has become routine for the treatment of high-risk patients with aortic stenosis. We assessed safety and feasibility of a left ventricular apical access and closure device combined with second-generation transapical (TA) TAVI transcatheter heart valves (THV).
METHODS AND RESULTS: Three elderly, comorbid patients (logEuroSCORE I 13.0-31.1%) received transapical aortic valve implantation (TA-AVI) via the Apica ASC device (Apica Cardiovascular Ltd., Galway, Ireland) using second-generation THV (Medtronic Engager [Medtronic 3F Therapeutics, Santa Ana, California, United States], JenaValve [JenaValve Technology GmbH, Munich, Germany], Symetis Acurate [Symetis S.A., Ecublens, Switzerland]). Access was gained using a non-rib-spreading technique and a novel access and closure device. THV deployment was successful with excellent hemodynamic outcome (no PVL, n = 2; trace PVL, n = 1; mean transvalvular gradients, 5-19 mm Hg) and complete apical hemostasis. No periprocedural major adverse events occurred and Valve Academic Research Consortium-2-defined composite end point of device success was met in all cases.
CONCLUSION: Safety and feasibility of TA-AVI using the ASC device with second-generation THV was demonstrated. Combining latest available technology is a major step toward improved functional outcome and decreased surgical trauma in TA-AVI. Potentially, technical enhancements may eventually pave the way toward a fully percutaneous TA-AVI procedure.
Bibliographical data
Original language | English |
---|---|
ISSN | 0171-6425 |
DOIs | |
Publication status | Published - 09.2014 |
Comment Deanary
Georg Thieme Verlag KG Stuttgart · New York.
PubMed | 24911901 |
---|