Single-centre experience with next-generation devices for transapical aortic valve implantation

Standard

Single-centre experience with next-generation devices for transapical aortic valve implantation. / Seiffert, Moritz; Conradi, Lenard; Kloth, Benjamin; Koschyk, Dietmar; Schirmer, Johannes; Schnabel, Renate B; Blankenberg, Stefan; Reichenspurner, Hermann; Diemert, Patrick; Treede, Hendrik.

In: EUR J CARDIO-THORAC, Vol. 47, No. 1, 01.2015, p. 39-45; discussion 45.

Research output: SCORING: Contribution to journalSCORING: Journal articleResearchpeer-review

Harvard

APA

Vancouver

Bibtex

@article{db381ba19064477ebfb201e8a634108c,
title = "Single-centre experience with next-generation devices for transapical aortic valve implantation",
abstract = "OBJECTIVES: Transcatheter aortic valve implantation (TAVI) has become an established procedure in patients with aortic stenosis and high surgical risk. Experience with first-generation transcatheter heart valves (THVs) is broad but limitations, e.g. paravalvular regurgitation, have been demonstrated. Much hope rests on the recently Conformit{\'e} Europ{\'e}enne mark approved next-generation devices to improve results in these patients. However, apart from the initial approval studies, clinical data with these new devices are still scarce. We aimed to assess short-term outcomes of 200 consecutive patients who underwent transapical TAVI with next-generation THV at our institution.METHODS: Transapical TAVI was performed in 200 consecutive patients 80.5±6.7 years old (38.5% female) at high surgical risk (log EuroSCORE 20.2±16.5%). Devices implanted were the Engager (Medtronic, Inc., Minneapolis, MN, USA; n=50), JenaValve (JenaValve Technology, Munich, Germany; n=88) and Symetis Acurate (Symetis SA, Ecublens, Switzerland; n=62) THV that were selected by the heart team on an individual basis. Data at baseline, during the procedure and follow-up were analysed according to standardized Valve Academic Research Consortium end points. Median follow-up was 219 days.RESULTS: Implantation was successful in 96.5% of cases. Valve function improved significantly with an increase in effective orifice area from 0.8±0.4 to 1.8±0.3 cm2 and a reduction in mean transvalvalvular gradients from 34.0±17.0 to 11.2±5.4 mmHg. Paravalvular regurgitation was none or trace in 70.3% of patients, Grade 1 in 26.1%, and Grade 2 in 3.5%. No patients developed aortic regurgitation>Grade 2. Major access site complications occurred in 6.5%, major stroke in 1.5% and stage-3 kidney injury in 2.5% of patients. A permanent pacemaker was implanted in 20.5% of patients overall and in 8.0% for a complete heart block. At 30-day follow-up 72.8% of patients were in New York Heart Association class I or II (10.5% at baseline). Overall survival was 91.5% at 30 days and 73.9% at 1 year.CONCLUSIONS: In a real-world clinical setting, next-generation transapical THV yielded positive haemodynamic results. The incidence of relevant paravalvular regurgitation was scarce in this group and clinical outcomes were encouraging during short-term follow-up. Long-term follow-up is required to investigate the durability of these new devices.",
keywords = "Aged, Aged, 80 and over, Aortic Valve/surgery, Aortic Valve Insufficiency/etiology, Female, Heart Valve Prosthesis, Humans, Male, Transcatheter Aortic Valve Replacement/adverse effects, Treatment Outcome",
author = "Moritz Seiffert and Lenard Conradi and Benjamin Kloth and Dietmar Koschyk and Johannes Schirmer and Schnabel, {Renate B} and Stefan Blankenberg and Hermann Reichenspurner and Patrick Diemert and Hendrik Treede",
note = "{\textcopyright} The Author 2014. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.",
year = "2015",
month = jan,
doi = "10.1093/ejcts/ezu041",
language = "English",
volume = "47",
pages = "39--45; discussion 45",
journal = "EUR J CARDIO-THORAC",
issn = "1010-7940",
publisher = "Elsevier",
number = "1",

}

RIS

TY - JOUR

T1 - Single-centre experience with next-generation devices for transapical aortic valve implantation

AU - Seiffert, Moritz

AU - Conradi, Lenard

AU - Kloth, Benjamin

AU - Koschyk, Dietmar

AU - Schirmer, Johannes

AU - Schnabel, Renate B

AU - Blankenberg, Stefan

AU - Reichenspurner, Hermann

AU - Diemert, Patrick

AU - Treede, Hendrik

N1 - © The Author 2014. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.

PY - 2015/1

Y1 - 2015/1

N2 - OBJECTIVES: Transcatheter aortic valve implantation (TAVI) has become an established procedure in patients with aortic stenosis and high surgical risk. Experience with first-generation transcatheter heart valves (THVs) is broad but limitations, e.g. paravalvular regurgitation, have been demonstrated. Much hope rests on the recently Conformité Européenne mark approved next-generation devices to improve results in these patients. However, apart from the initial approval studies, clinical data with these new devices are still scarce. We aimed to assess short-term outcomes of 200 consecutive patients who underwent transapical TAVI with next-generation THV at our institution.METHODS: Transapical TAVI was performed in 200 consecutive patients 80.5±6.7 years old (38.5% female) at high surgical risk (log EuroSCORE 20.2±16.5%). Devices implanted were the Engager (Medtronic, Inc., Minneapolis, MN, USA; n=50), JenaValve (JenaValve Technology, Munich, Germany; n=88) and Symetis Acurate (Symetis SA, Ecublens, Switzerland; n=62) THV that were selected by the heart team on an individual basis. Data at baseline, during the procedure and follow-up were analysed according to standardized Valve Academic Research Consortium end points. Median follow-up was 219 days.RESULTS: Implantation was successful in 96.5% of cases. Valve function improved significantly with an increase in effective orifice area from 0.8±0.4 to 1.8±0.3 cm2 and a reduction in mean transvalvalvular gradients from 34.0±17.0 to 11.2±5.4 mmHg. Paravalvular regurgitation was none or trace in 70.3% of patients, Grade 1 in 26.1%, and Grade 2 in 3.5%. No patients developed aortic regurgitation>Grade 2. Major access site complications occurred in 6.5%, major stroke in 1.5% and stage-3 kidney injury in 2.5% of patients. A permanent pacemaker was implanted in 20.5% of patients overall and in 8.0% for a complete heart block. At 30-day follow-up 72.8% of patients were in New York Heart Association class I or II (10.5% at baseline). Overall survival was 91.5% at 30 days and 73.9% at 1 year.CONCLUSIONS: In a real-world clinical setting, next-generation transapical THV yielded positive haemodynamic results. The incidence of relevant paravalvular regurgitation was scarce in this group and clinical outcomes were encouraging during short-term follow-up. Long-term follow-up is required to investigate the durability of these new devices.

AB - OBJECTIVES: Transcatheter aortic valve implantation (TAVI) has become an established procedure in patients with aortic stenosis and high surgical risk. Experience with first-generation transcatheter heart valves (THVs) is broad but limitations, e.g. paravalvular regurgitation, have been demonstrated. Much hope rests on the recently Conformité Européenne mark approved next-generation devices to improve results in these patients. However, apart from the initial approval studies, clinical data with these new devices are still scarce. We aimed to assess short-term outcomes of 200 consecutive patients who underwent transapical TAVI with next-generation THV at our institution.METHODS: Transapical TAVI was performed in 200 consecutive patients 80.5±6.7 years old (38.5% female) at high surgical risk (log EuroSCORE 20.2±16.5%). Devices implanted were the Engager (Medtronic, Inc., Minneapolis, MN, USA; n=50), JenaValve (JenaValve Technology, Munich, Germany; n=88) and Symetis Acurate (Symetis SA, Ecublens, Switzerland; n=62) THV that were selected by the heart team on an individual basis. Data at baseline, during the procedure and follow-up were analysed according to standardized Valve Academic Research Consortium end points. Median follow-up was 219 days.RESULTS: Implantation was successful in 96.5% of cases. Valve function improved significantly with an increase in effective orifice area from 0.8±0.4 to 1.8±0.3 cm2 and a reduction in mean transvalvalvular gradients from 34.0±17.0 to 11.2±5.4 mmHg. Paravalvular regurgitation was none or trace in 70.3% of patients, Grade 1 in 26.1%, and Grade 2 in 3.5%. No patients developed aortic regurgitation>Grade 2. Major access site complications occurred in 6.5%, major stroke in 1.5% and stage-3 kidney injury in 2.5% of patients. A permanent pacemaker was implanted in 20.5% of patients overall and in 8.0% for a complete heart block. At 30-day follow-up 72.8% of patients were in New York Heart Association class I or II (10.5% at baseline). Overall survival was 91.5% at 30 days and 73.9% at 1 year.CONCLUSIONS: In a real-world clinical setting, next-generation transapical THV yielded positive haemodynamic results. The incidence of relevant paravalvular regurgitation was scarce in this group and clinical outcomes were encouraging during short-term follow-up. Long-term follow-up is required to investigate the durability of these new devices.

KW - Aged

KW - Aged, 80 and over

KW - Aortic Valve/surgery

KW - Aortic Valve Insufficiency/etiology

KW - Female

KW - Heart Valve Prosthesis

KW - Humans

KW - Male

KW - Transcatheter Aortic Valve Replacement/adverse effects

KW - Treatment Outcome

U2 - 10.1093/ejcts/ezu041

DO - 10.1093/ejcts/ezu041

M3 - SCORING: Journal article

C2 - 24578405

VL - 47

SP - 39-45; discussion 45

JO - EUR J CARDIO-THORAC

JF - EUR J CARDIO-THORAC

SN - 1010-7940

IS - 1

ER -