Severe Cobalt Intoxication Due to Prosthesis Wear in Repeated Total Hip Arthroplasty.

Standard

Severe Cobalt Intoxication Due to Prosthesis Wear in Repeated Total Hip Arthroplasty. / Oldenburg, Marcus; Wegner, Ralf; Baur, Xaver.

In: J ARTHROPLASTY, 2008.

Research output: SCORING: Contribution to journalSCORING: Journal articleResearchpeer-review

Harvard

APA

Vancouver

Bibtex

@article{40c52d2c25f6451d8452c1d77d3c5789,
title = "Severe Cobalt Intoxication Due to Prosthesis Wear in Repeated Total Hip Arthroplasty.",
abstract = "Our study describes a 55-year-old man with a total hip prosthesis (ceramic femoral head and polyethylene [PE] inlay). After a fall, the ceramic head broke into several pieces, and a subsequent revision surgery with metal femoral head and PE inlay was performed. Three months later, the so far healthy patient complained of multiorganic symptoms. The subsequent clinical examinations revealed hypothyroidism, peripheral neuropathy, and cardiomyopathy for the first time. In a second revision surgery, a massive deterioration of the metal femoral head by overlooked particles of the broken ceramic head was found. The cobalt concentration in blood exceeded a never before observed level. After replacement of the scraped femoral head, the cobalt concentration decreased considerably, and most of the symptoms disappeared.",
author = "Marcus Oldenburg and Ralf Wegner and Xaver Baur",
year = "2008",
language = "Deutsch",
journal = "J ARTHROPLASTY",
issn = "0883-5403",
publisher = "Churchill Livingstone",

}

RIS

TY - JOUR

T1 - Severe Cobalt Intoxication Due to Prosthesis Wear in Repeated Total Hip Arthroplasty.

AU - Oldenburg, Marcus

AU - Wegner, Ralf

AU - Baur, Xaver

PY - 2008

Y1 - 2008

N2 - Our study describes a 55-year-old man with a total hip prosthesis (ceramic femoral head and polyethylene [PE] inlay). After a fall, the ceramic head broke into several pieces, and a subsequent revision surgery with metal femoral head and PE inlay was performed. Three months later, the so far healthy patient complained of multiorganic symptoms. The subsequent clinical examinations revealed hypothyroidism, peripheral neuropathy, and cardiomyopathy for the first time. In a second revision surgery, a massive deterioration of the metal femoral head by overlooked particles of the broken ceramic head was found. The cobalt concentration in blood exceeded a never before observed level. After replacement of the scraped femoral head, the cobalt concentration decreased considerably, and most of the symptoms disappeared.

AB - Our study describes a 55-year-old man with a total hip prosthesis (ceramic femoral head and polyethylene [PE] inlay). After a fall, the ceramic head broke into several pieces, and a subsequent revision surgery with metal femoral head and PE inlay was performed. Three months later, the so far healthy patient complained of multiorganic symptoms. The subsequent clinical examinations revealed hypothyroidism, peripheral neuropathy, and cardiomyopathy for the first time. In a second revision surgery, a massive deterioration of the metal femoral head by overlooked particles of the broken ceramic head was found. The cobalt concentration in blood exceeded a never before observed level. After replacement of the scraped femoral head, the cobalt concentration decreased considerably, and most of the symptoms disappeared.

M3 - SCORING: Zeitschriftenaufsatz

JO - J ARTHROPLASTY

JF - J ARTHROPLASTY

SN - 0883-5403

ER -