[Retrospective analysis of transpulmonary and pulmonary arterial measurement of cardiac output in ARDS patients]

Standard

[Retrospective analysis of transpulmonary and pulmonary arterial measurement of cardiac output in ARDS patients]. / Zöllner, Christian; Briegel, J; Kilger, E; Haller, M.

In: ANAESTHESIST, Vol. 47, No. 11, 11, 1998, p. 912-917.

Research output: SCORING: Contribution to journalSCORING: Journal articleResearchpeer-review

Harvard

APA

Vancouver

Bibtex

@article{26d98889154f41329154babe6a0bc310,
title = "[Retrospective analysis of transpulmonary and pulmonary arterial measurement of cardiac output in ARDS patients]",
abstract = "OBJECTIVES: To investigate the agreement (and its potential dependency on extra-vascular lung water) between transpulmonary (TPID) and standard pulmonary artery (PAID) thermodilution cardiac output measurements. METHODS: One hundred and sixty simultaneous cardiac output measurements using transpulmonary and pulmonary artery thermodilution techniques were retrospectively compared in 18 patients with acute respiratory distress syndrome. In addition, extravascular lung water was determined using a double indicator technique (temperature and indocyanine green). RESULTS: Mean (+/- SD) difference ({"}bias{"}) was 0.03 L/min (+/- 1.04 L/min), linear regression analysis resulted in TPID = 0.87 PAID + 1.16 (r = 0.91). Mean extra vascular lung water was 1625 mL (minimum-maximum: 403-3266 mL) and therefore markedly elevated as could have been expected in patients with ARDS. Bias (PAID-TPID) was not dependent on extravascular lung water. CONCLUSIONS: Transpulmonary and pulmonary artery thermodilution methods can be used interchangeably. The results demonstrate for the first time in humans that transpulmonary thermodilution provides valid cardiac output values in patients with markedly increased fluid content of the lungs.",
author = "Christian Z{\"o}llner and J Briegel and E Kilger and M Haller",
year = "1998",
language = "Deutsch",
volume = "47",
pages = "912--917",
journal = "ANAESTHESIST",
issn = "0003-2417",
publisher = "Springer",
number = "11",

}

RIS

TY - JOUR

T1 - [Retrospective analysis of transpulmonary and pulmonary arterial measurement of cardiac output in ARDS patients]

AU - Zöllner, Christian

AU - Briegel, J

AU - Kilger, E

AU - Haller, M

PY - 1998

Y1 - 1998

N2 - OBJECTIVES: To investigate the agreement (and its potential dependency on extra-vascular lung water) between transpulmonary (TPID) and standard pulmonary artery (PAID) thermodilution cardiac output measurements. METHODS: One hundred and sixty simultaneous cardiac output measurements using transpulmonary and pulmonary artery thermodilution techniques were retrospectively compared in 18 patients with acute respiratory distress syndrome. In addition, extravascular lung water was determined using a double indicator technique (temperature and indocyanine green). RESULTS: Mean (+/- SD) difference ("bias") was 0.03 L/min (+/- 1.04 L/min), linear regression analysis resulted in TPID = 0.87 PAID + 1.16 (r = 0.91). Mean extra vascular lung water was 1625 mL (minimum-maximum: 403-3266 mL) and therefore markedly elevated as could have been expected in patients with ARDS. Bias (PAID-TPID) was not dependent on extravascular lung water. CONCLUSIONS: Transpulmonary and pulmonary artery thermodilution methods can be used interchangeably. The results demonstrate for the first time in humans that transpulmonary thermodilution provides valid cardiac output values in patients with markedly increased fluid content of the lungs.

AB - OBJECTIVES: To investigate the agreement (and its potential dependency on extra-vascular lung water) between transpulmonary (TPID) and standard pulmonary artery (PAID) thermodilution cardiac output measurements. METHODS: One hundred and sixty simultaneous cardiac output measurements using transpulmonary and pulmonary artery thermodilution techniques were retrospectively compared in 18 patients with acute respiratory distress syndrome. In addition, extravascular lung water was determined using a double indicator technique (temperature and indocyanine green). RESULTS: Mean (+/- SD) difference ("bias") was 0.03 L/min (+/- 1.04 L/min), linear regression analysis resulted in TPID = 0.87 PAID + 1.16 (r = 0.91). Mean extra vascular lung water was 1625 mL (minimum-maximum: 403-3266 mL) and therefore markedly elevated as could have been expected in patients with ARDS. Bias (PAID-TPID) was not dependent on extravascular lung water. CONCLUSIONS: Transpulmonary and pulmonary artery thermodilution methods can be used interchangeably. The results demonstrate for the first time in humans that transpulmonary thermodilution provides valid cardiac output values in patients with markedly increased fluid content of the lungs.

M3 - SCORING: Zeitschriftenaufsatz

VL - 47

SP - 912

EP - 917

JO - ANAESTHESIST

JF - ANAESTHESIST

SN - 0003-2417

IS - 11

M1 - 11

ER -