Reliable and Easy-To-Use Liquid Chromatography-Tandem Mass Spectrometry Method for Simultaneous Analysis of Fluconazole, Isavuconazole, Itraconazole, Hydroxy-Itraconazole, Posaconazole, and Voriconazole in Human Plasma and Serum

Standard

Reliable and Easy-To-Use Liquid Chromatography-Tandem Mass Spectrometry Method for Simultaneous Analysis of Fluconazole, Isavuconazole, Itraconazole, Hydroxy-Itraconazole, Posaconazole, and Voriconazole in Human Plasma and Serum. / Müller, Carsten; Gehlen, David; Blaich, Cornelia; Prozeller, Domenik; Liss, Blasius; Streichert, Thomas; Wiesen, Martin H J.

In: THER DRUG MONIT, Vol. 39, No. 5, 10.2017, p. 505-513.

Research output: SCORING: Contribution to journalSCORING: Journal articleResearchpeer-review

Harvard

APA

Vancouver

Bibtex

@article{9a107fbb250e4cdcb38630a510941869,
title = "Reliable and Easy-To-Use Liquid Chromatography-Tandem Mass Spectrometry Method for Simultaneous Analysis of Fluconazole, Isavuconazole, Itraconazole, Hydroxy-Itraconazole, Posaconazole, and Voriconazole in Human Plasma and Serum",
abstract = "BACKGROUND: A fast and easy-to-use liquid chromatography-tandem mass spectrometry method for the determination and quantification of 6 triazoles [fluconazole (FLZ), isavuconazole (ISZ), itraconazole (ITZ), hydroxy-itraconazole (OH-ITZ), posaconazole (PSZ), and voriconazole (VRZ)] in human plasma and serum was developed and validated for therapeutic drug monitoring.METHODS: Sample preparation was based on protein precipitation with acetonitrile and subsequent centrifugation. Isotope-labeled analogues for each analyte were used as internal standards. Chromatographic separation was achieved using a 50 × 2.1 mm, 1.9 μm polar Hypersil Gold C18 column and mobile phase consisting of 0.1% formic acid/acetonitrile (45%/55%, vol/vol) at a flow rate of 340 μL/min. The triazoles were simultaneously detected using a triple-stage quadrupole mass spectrometer operated in selected reaction monitoring mode with positive heated electrospray ionization within a single runtime of t = 3.00 minutes.RESULTS: Linearity of all azole concentration ranges was verified by the Mandel test and demonstrated for all azoles. All calibration curves were linear and fitted using least squares regression with a weighting factor of the reciprocal concentration. Limits of detection (μg/L/L) were FLZ, 9.3; ISZ, 0.3; ITZ, 0.6; OH-ITZ, 8.6; PSZ, 3.4; and VRZ, 2.1. The lower limits of quantitation (μg/L/liter) were FLZ, 28.3; ISZ, 1.0; ITZ, 1.7; OH-ITZ, 26.2; PSZ, 10.3; and VRZ, 6.3. Intraday and interday precisions ranged from 0.6% to 6.6% for all azoles. Intraday and interday accuracies (%bias) of all analytes were within 10.5%. In addition, we report on a 29-year-old white woman (94 kg body weight) with a history of acute myeloid leukemia who underwent stem cell transplantation. Because of diagnosis of aspergillus pneumonia, antifungal pharmacotherapy was initiated with different application modes and dosages of ISZ, and plasma concentrations were monitored over a time period of 6 months.CONCLUSIONS: A precise and highly sensitive liquid chromatography-tandem mass spectrometry method was developed that enables quantification of triazoles in plasma and serum matrix across therapeutically relevant concentration ranges. It was successfully implemented in our therapeutic drug monitoring routine service and is suitable for routine monitoring of antifungal therapy and in severely ill patients.",
keywords = "Journal Article",
author = "Carsten M{\"u}ller and David Gehlen and Cornelia Blaich and Domenik Prozeller and Blasius Liss and Thomas Streichert and Wiesen, {Martin H J}",
year = "2017",
month = oct,
doi = "10.1097/FTD.0000000000000438",
language = "English",
volume = "39",
pages = "505--513",
journal = "THER DRUG MONIT",
issn = "0163-4356",
publisher = "Lippincott Williams and Wilkins",
number = "5",

}

RIS

TY - JOUR

T1 - Reliable and Easy-To-Use Liquid Chromatography-Tandem Mass Spectrometry Method for Simultaneous Analysis of Fluconazole, Isavuconazole, Itraconazole, Hydroxy-Itraconazole, Posaconazole, and Voriconazole in Human Plasma and Serum

AU - Müller, Carsten

AU - Gehlen, David

AU - Blaich, Cornelia

AU - Prozeller, Domenik

AU - Liss, Blasius

AU - Streichert, Thomas

AU - Wiesen, Martin H J

PY - 2017/10

Y1 - 2017/10

N2 - BACKGROUND: A fast and easy-to-use liquid chromatography-tandem mass spectrometry method for the determination and quantification of 6 triazoles [fluconazole (FLZ), isavuconazole (ISZ), itraconazole (ITZ), hydroxy-itraconazole (OH-ITZ), posaconazole (PSZ), and voriconazole (VRZ)] in human plasma and serum was developed and validated for therapeutic drug monitoring.METHODS: Sample preparation was based on protein precipitation with acetonitrile and subsequent centrifugation. Isotope-labeled analogues for each analyte were used as internal standards. Chromatographic separation was achieved using a 50 × 2.1 mm, 1.9 μm polar Hypersil Gold C18 column and mobile phase consisting of 0.1% formic acid/acetonitrile (45%/55%, vol/vol) at a flow rate of 340 μL/min. The triazoles were simultaneously detected using a triple-stage quadrupole mass spectrometer operated in selected reaction monitoring mode with positive heated electrospray ionization within a single runtime of t = 3.00 minutes.RESULTS: Linearity of all azole concentration ranges was verified by the Mandel test and demonstrated for all azoles. All calibration curves were linear and fitted using least squares regression with a weighting factor of the reciprocal concentration. Limits of detection (μg/L/L) were FLZ, 9.3; ISZ, 0.3; ITZ, 0.6; OH-ITZ, 8.6; PSZ, 3.4; and VRZ, 2.1. The lower limits of quantitation (μg/L/liter) were FLZ, 28.3; ISZ, 1.0; ITZ, 1.7; OH-ITZ, 26.2; PSZ, 10.3; and VRZ, 6.3. Intraday and interday precisions ranged from 0.6% to 6.6% for all azoles. Intraday and interday accuracies (%bias) of all analytes were within 10.5%. In addition, we report on a 29-year-old white woman (94 kg body weight) with a history of acute myeloid leukemia who underwent stem cell transplantation. Because of diagnosis of aspergillus pneumonia, antifungal pharmacotherapy was initiated with different application modes and dosages of ISZ, and plasma concentrations were monitored over a time period of 6 months.CONCLUSIONS: A precise and highly sensitive liquid chromatography-tandem mass spectrometry method was developed that enables quantification of triazoles in plasma and serum matrix across therapeutically relevant concentration ranges. It was successfully implemented in our therapeutic drug monitoring routine service and is suitable for routine monitoring of antifungal therapy and in severely ill patients.

AB - BACKGROUND: A fast and easy-to-use liquid chromatography-tandem mass spectrometry method for the determination and quantification of 6 triazoles [fluconazole (FLZ), isavuconazole (ISZ), itraconazole (ITZ), hydroxy-itraconazole (OH-ITZ), posaconazole (PSZ), and voriconazole (VRZ)] in human plasma and serum was developed and validated for therapeutic drug monitoring.METHODS: Sample preparation was based on protein precipitation with acetonitrile and subsequent centrifugation. Isotope-labeled analogues for each analyte were used as internal standards. Chromatographic separation was achieved using a 50 × 2.1 mm, 1.9 μm polar Hypersil Gold C18 column and mobile phase consisting of 0.1% formic acid/acetonitrile (45%/55%, vol/vol) at a flow rate of 340 μL/min. The triazoles were simultaneously detected using a triple-stage quadrupole mass spectrometer operated in selected reaction monitoring mode with positive heated electrospray ionization within a single runtime of t = 3.00 minutes.RESULTS: Linearity of all azole concentration ranges was verified by the Mandel test and demonstrated for all azoles. All calibration curves were linear and fitted using least squares regression with a weighting factor of the reciprocal concentration. Limits of detection (μg/L/L) were FLZ, 9.3; ISZ, 0.3; ITZ, 0.6; OH-ITZ, 8.6; PSZ, 3.4; and VRZ, 2.1. The lower limits of quantitation (μg/L/liter) were FLZ, 28.3; ISZ, 1.0; ITZ, 1.7; OH-ITZ, 26.2; PSZ, 10.3; and VRZ, 6.3. Intraday and interday precisions ranged from 0.6% to 6.6% for all azoles. Intraday and interday accuracies (%bias) of all analytes were within 10.5%. In addition, we report on a 29-year-old white woman (94 kg body weight) with a history of acute myeloid leukemia who underwent stem cell transplantation. Because of diagnosis of aspergillus pneumonia, antifungal pharmacotherapy was initiated with different application modes and dosages of ISZ, and plasma concentrations were monitored over a time period of 6 months.CONCLUSIONS: A precise and highly sensitive liquid chromatography-tandem mass spectrometry method was developed that enables quantification of triazoles in plasma and serum matrix across therapeutically relevant concentration ranges. It was successfully implemented in our therapeutic drug monitoring routine service and is suitable for routine monitoring of antifungal therapy and in severely ill patients.

KW - Journal Article

U2 - 10.1097/FTD.0000000000000438

DO - 10.1097/FTD.0000000000000438

M3 - SCORING: Journal article

C2 - 28742650

VL - 39

SP - 505

EP - 513

JO - THER DRUG MONIT

JF - THER DRUG MONIT

SN - 0163-4356

IS - 5

ER -