Performance evaluation of a novel multi-pinhole collimator for dopamine transporter SPECT

Standard

Performance evaluation of a novel multi-pinhole collimator for dopamine transporter SPECT. / Tecklenburg, K; Forgács, A; Apostolova, I; Lehnert, W; Klutmann, S; Csirik, J; Garutti, E; Buchert, R.

In: PHYS MED BIOL, Vol. 65, No. 16, 21.08.2020, p. 165015.

Research output: SCORING: Contribution to journalSCORING: Journal articleResearchpeer-review

Harvard

APA

Vancouver

Bibtex

@article{6f3e92b7b8e94fd6aca794fd6f97e64e,
title = "Performance evaluation of a novel multi-pinhole collimator for dopamine transporter SPECT",
abstract = "There is a tradeoff between spatial resolution and count sensitivity in SPECT with conventional collimators. Multi-pinhole (MPH) collimator technology has potential for concurrent improvement of resolution and sensitivity in clinical SPECT of 'small' organs. This study evaluated a novel MPH collimator specifically designed for dopamine transporter (DAT) SPECT with a triple-head SPECT camera. Count sensitivity was measured with a 99mTc point source placed on the lattice points of a 1 cm grid covering the whole field-of-view (FOV). Spatial resolution was assessed with a Derenzo type hot rod phantom. An anthropomorphic striatum phantom was scanned with total activity representative of a typical patient scan and different striatum-to-background activity concentration ratios. Recovery of striatum-to-background contrast was assessed by the contrast-recovery-coefficient. Measurements were repeated with double-head SPECT with fan-beam or low-energy-high-resolution-high-sensitivity (LEHRHS) collimators. A patient referred to DAT SPECT because of suspicion of Parkinson's disease was scanned with both LEHRHS and MPH collimators after a single tracer injection. The axial MPH sensitivity profile was approximately symmetrical around its peak, although it was shifted 7 cm towards the patient to simplify positioning. Peak sensitivity of the triple-head MPH system in the center of the FOV was 620 cps MBq-1 compared to 225 cps MBq-1 for the double-head fan-beam system. Sensitivity of the MPH system decreased towards the edges of the FOV. The full width of the sensitivity profile at 200 cps MBq-1 was 21 cm transaxially and 11 cm axially. In MPH SPECT of the Derenzo phantom all rods with ≥ 5 mm diameter were clearly visible. MPH SPECT improved striatal contrast recovery by ≥ 20% compared to fan-beam SPECT. The patient scan demonstrated good image quality of MPH SPECT with almost PET-like delineation of putamen and caudate nucleus. SPECT with dedicated MPH collimators provides considerable improvement of the resolution-sensitivity tradeoff in DAT SPECT compared to SPECT with fan-beam or LEHRHS collimators.",
keywords = "Dopamine Plasma Membrane Transport Proteins/metabolism, Humans, Neostriatum/metabolism, Parkinson Disease/diagnostic imaging, Phantoms, Imaging, Tomography, Emission-Computed, Single-Photon/instrumentation",
author = "K Tecklenburg and A Forg{\'a}cs and I Apostolova and W Lehnert and S Klutmann and J Csirik and E Garutti and R Buchert",
year = "2020",
month = aug,
day = "21",
doi = "10.1088/1361-6560/ab9067",
language = "English",
volume = "65",
pages = "165015",
journal = "PHYS MED BIOL",
issn = "0031-9155",
publisher = "IOP Publishing Ltd.",
number = "16",

}

RIS

TY - JOUR

T1 - Performance evaluation of a novel multi-pinhole collimator for dopamine transporter SPECT

AU - Tecklenburg, K

AU - Forgács, A

AU - Apostolova, I

AU - Lehnert, W

AU - Klutmann, S

AU - Csirik, J

AU - Garutti, E

AU - Buchert, R

PY - 2020/8/21

Y1 - 2020/8/21

N2 - There is a tradeoff between spatial resolution and count sensitivity in SPECT with conventional collimators. Multi-pinhole (MPH) collimator technology has potential for concurrent improvement of resolution and sensitivity in clinical SPECT of 'small' organs. This study evaluated a novel MPH collimator specifically designed for dopamine transporter (DAT) SPECT with a triple-head SPECT camera. Count sensitivity was measured with a 99mTc point source placed on the lattice points of a 1 cm grid covering the whole field-of-view (FOV). Spatial resolution was assessed with a Derenzo type hot rod phantom. An anthropomorphic striatum phantom was scanned with total activity representative of a typical patient scan and different striatum-to-background activity concentration ratios. Recovery of striatum-to-background contrast was assessed by the contrast-recovery-coefficient. Measurements were repeated with double-head SPECT with fan-beam or low-energy-high-resolution-high-sensitivity (LEHRHS) collimators. A patient referred to DAT SPECT because of suspicion of Parkinson's disease was scanned with both LEHRHS and MPH collimators after a single tracer injection. The axial MPH sensitivity profile was approximately symmetrical around its peak, although it was shifted 7 cm towards the patient to simplify positioning. Peak sensitivity of the triple-head MPH system in the center of the FOV was 620 cps MBq-1 compared to 225 cps MBq-1 for the double-head fan-beam system. Sensitivity of the MPH system decreased towards the edges of the FOV. The full width of the sensitivity profile at 200 cps MBq-1 was 21 cm transaxially and 11 cm axially. In MPH SPECT of the Derenzo phantom all rods with ≥ 5 mm diameter were clearly visible. MPH SPECT improved striatal contrast recovery by ≥ 20% compared to fan-beam SPECT. The patient scan demonstrated good image quality of MPH SPECT with almost PET-like delineation of putamen and caudate nucleus. SPECT with dedicated MPH collimators provides considerable improvement of the resolution-sensitivity tradeoff in DAT SPECT compared to SPECT with fan-beam or LEHRHS collimators.

AB - There is a tradeoff between spatial resolution and count sensitivity in SPECT with conventional collimators. Multi-pinhole (MPH) collimator technology has potential for concurrent improvement of resolution and sensitivity in clinical SPECT of 'small' organs. This study evaluated a novel MPH collimator specifically designed for dopamine transporter (DAT) SPECT with a triple-head SPECT camera. Count sensitivity was measured with a 99mTc point source placed on the lattice points of a 1 cm grid covering the whole field-of-view (FOV). Spatial resolution was assessed with a Derenzo type hot rod phantom. An anthropomorphic striatum phantom was scanned with total activity representative of a typical patient scan and different striatum-to-background activity concentration ratios. Recovery of striatum-to-background contrast was assessed by the contrast-recovery-coefficient. Measurements were repeated with double-head SPECT with fan-beam or low-energy-high-resolution-high-sensitivity (LEHRHS) collimators. A patient referred to DAT SPECT because of suspicion of Parkinson's disease was scanned with both LEHRHS and MPH collimators after a single tracer injection. The axial MPH sensitivity profile was approximately symmetrical around its peak, although it was shifted 7 cm towards the patient to simplify positioning. Peak sensitivity of the triple-head MPH system in the center of the FOV was 620 cps MBq-1 compared to 225 cps MBq-1 for the double-head fan-beam system. Sensitivity of the MPH system decreased towards the edges of the FOV. The full width of the sensitivity profile at 200 cps MBq-1 was 21 cm transaxially and 11 cm axially. In MPH SPECT of the Derenzo phantom all rods with ≥ 5 mm diameter were clearly visible. MPH SPECT improved striatal contrast recovery by ≥ 20% compared to fan-beam SPECT. The patient scan demonstrated good image quality of MPH SPECT with almost PET-like delineation of putamen and caudate nucleus. SPECT with dedicated MPH collimators provides considerable improvement of the resolution-sensitivity tradeoff in DAT SPECT compared to SPECT with fan-beam or LEHRHS collimators.

KW - Dopamine Plasma Membrane Transport Proteins/metabolism

KW - Humans

KW - Neostriatum/metabolism

KW - Parkinson Disease/diagnostic imaging

KW - Phantoms, Imaging

KW - Tomography, Emission-Computed, Single-Photon/instrumentation

U2 - 10.1088/1361-6560/ab9067

DO - 10.1088/1361-6560/ab9067

M3 - SCORING: Journal article

C2 - 32369781

VL - 65

SP - 165015

JO - PHYS MED BIOL

JF - PHYS MED BIOL

SN - 0031-9155

IS - 16

ER -