Left Atrial Appendage Closure Guided by Integrated Echocardiography and Fluoroscopy Imaging Reduces Radiation Exposure

  • Christiane Jungen
  • Tobias Zeus
  • Jan Balzer
  • Christian Eickholt
  • Margot Petersen
  • Eva Kehmeier
  • Verena Veulemans
  • Malte Kelm
  • Stephan Willems
  • Christian Meyer

Abstract

AIMS: To investigate whether percutaneous left atrial appendage (LAA) closure guided by automated real-time integration of 2D-/3D-transesophageal echocardiography (TEE) and fluoroscopy imaging results in decreased radiation exposure.

METHODS AND RESULTS: In this open-label single-center study LAA closure (AmplatzerTM Cardiac Plug) was performed in 34 consecutive patients (8 women; 73.1±8.5 years) with (n = 17, EN+) or without (n = 17, EN-) integrated echocardiography/fluoroscopy imaging guidance (EchoNavigator® [EN]; Philips Healthcare). There were no significant differences in baseline characteristics between both groups. Successful LAA closure was documented in all patients. Radiation dose was reduced in the EN+ group about 52% (EN+: 48.5±30.7 vs. EN-: 93.9±64.4 Gy/cm2; p = 0.01). Corresponding to the radiation dose fluoroscopy time was reduced (EN+: 16.7±7 vs. EN-: 24.0±11.4 min; p = 0.035). These advantages were not at the cost of increased procedure time (89.6±28.8 vs. 90.1±30.2 min; p = 0.96) or periprocedural complications. Contrast media amount was comparable between both groups (172.3±92.7 vs. 197.5±127.8 ml; p = 0.53). During short-term follow-up of at least 3 months (mean: 8.1±5.9 months) no device-related events occurred.

CONCLUSIONS: Automated real-time integration of echocardiography and fluoroscopy can be incorporated into procedural work-flow of percutaneous left atrial appendage closure without prolonging procedure time. This approach results in a relevant reduction of radiation exposure.

TRIAL REGISTRATION: ClinicalTrials.gov NCT01262508.