Head-to-head comparison of two continuous glucose monitoring systems on a cardio-surgical ICU

Standard

Head-to-head comparison of two continuous glucose monitoring systems on a cardio-surgical ICU. / Punke, M A; Decker, C; Petzoldt, M; Reuter, D A; Wodack, K H; Reichenspurner, H; Kubik, M; Kluge, S.

In: J CLIN MONIT COMPUT, Vol. 33, No. 5, 10.2019, p. 895-901.

Research output: SCORING: Contribution to journalSCORING: Journal articleResearchpeer-review

Harvard

APA

Vancouver

Bibtex

@article{8f0d5350aa59408b806dcb948ad7a7c5,
title = "Head-to-head comparison of two continuous glucose monitoring systems on a cardio-surgical ICU",
abstract = "In critical illness hypo-and hyperglycemia have a negative influence on patient outcome. Continuous glucose monitoring (CGM) could help in early detection of hypo-and hyperglycemia. A requirement for these new methods is an acceptable accuracy and precision in clinical practice. In this pilot study we prospectively evaluated the accuracy and precision of two CGM sensors (subcutaneous sensor: Sentrino{\textregistered}, Medtronic and intravasal sensor: Glucoclear{\textregistered}, Edwards) in 20 patients on a cardio-surgical ICU in a head to head comparison. CGM data were recorded for up to 48 h and values were compared with blood-gas-analysis (BGA) values, analysed with Bland-Altman-plots and color-coded surveillance error-grids. Shown are means ± standard deviations. In total 270/255 intravasal/subcutaneous pairs with BGA-values were analysed. The average runtime of the sensors was 28.4 ± 6.4 h. Correlation with BGA values yielded a correlation coefficient of 0.76 (subcutaneous sensor) and 0.92 (intravasal sensor). The Bland Altman Plots revealed an accuracy of 2.5 mg/dl, and a precision of + 43.0 mg/dl to - 38.0 mg/dl (subcutaneous sensor) and an accuracy of - 6.0 mg/dl, and a precision of + 12.4 mg/dl to - 24.4 mg/dl (intravasal sensor). No severe hypoglycemic event, defined as BG level below 40 mg/dl, occurred during treatment. Both sensors showed good accuracy in comparison to the BGA values, however they differ regarding precision, which in case of the subcutaneous sensor is considerable high.",
keywords = "Journal Article",
author = "Punke, {M A} and C Decker and M Petzoldt and Reuter, {D A} and Wodack, {K H} and H Reichenspurner and M Kubik and S Kluge",
year = "2019",
month = oct,
doi = "10.1007/s10877-018-0221-5",
language = "English",
volume = "33",
pages = "895--901",
journal = "J CLIN MONIT COMPUT",
issn = "1387-1307",
publisher = "Springer Netherlands",
number = "5",

}

RIS

TY - JOUR

T1 - Head-to-head comparison of two continuous glucose monitoring systems on a cardio-surgical ICU

AU - Punke, M A

AU - Decker, C

AU - Petzoldt, M

AU - Reuter, D A

AU - Wodack, K H

AU - Reichenspurner, H

AU - Kubik, M

AU - Kluge, S

PY - 2019/10

Y1 - 2019/10

N2 - In critical illness hypo-and hyperglycemia have a negative influence on patient outcome. Continuous glucose monitoring (CGM) could help in early detection of hypo-and hyperglycemia. A requirement for these new methods is an acceptable accuracy and precision in clinical practice. In this pilot study we prospectively evaluated the accuracy and precision of two CGM sensors (subcutaneous sensor: Sentrino®, Medtronic and intravasal sensor: Glucoclear®, Edwards) in 20 patients on a cardio-surgical ICU in a head to head comparison. CGM data were recorded for up to 48 h and values were compared with blood-gas-analysis (BGA) values, analysed with Bland-Altman-plots and color-coded surveillance error-grids. Shown are means ± standard deviations. In total 270/255 intravasal/subcutaneous pairs with BGA-values were analysed. The average runtime of the sensors was 28.4 ± 6.4 h. Correlation with BGA values yielded a correlation coefficient of 0.76 (subcutaneous sensor) and 0.92 (intravasal sensor). The Bland Altman Plots revealed an accuracy of 2.5 mg/dl, and a precision of + 43.0 mg/dl to - 38.0 mg/dl (subcutaneous sensor) and an accuracy of - 6.0 mg/dl, and a precision of + 12.4 mg/dl to - 24.4 mg/dl (intravasal sensor). No severe hypoglycemic event, defined as BG level below 40 mg/dl, occurred during treatment. Both sensors showed good accuracy in comparison to the BGA values, however they differ regarding precision, which in case of the subcutaneous sensor is considerable high.

AB - In critical illness hypo-and hyperglycemia have a negative influence on patient outcome. Continuous glucose monitoring (CGM) could help in early detection of hypo-and hyperglycemia. A requirement for these new methods is an acceptable accuracy and precision in clinical practice. In this pilot study we prospectively evaluated the accuracy and precision of two CGM sensors (subcutaneous sensor: Sentrino®, Medtronic and intravasal sensor: Glucoclear®, Edwards) in 20 patients on a cardio-surgical ICU in a head to head comparison. CGM data were recorded for up to 48 h and values were compared with blood-gas-analysis (BGA) values, analysed with Bland-Altman-plots and color-coded surveillance error-grids. Shown are means ± standard deviations. In total 270/255 intravasal/subcutaneous pairs with BGA-values were analysed. The average runtime of the sensors was 28.4 ± 6.4 h. Correlation with BGA values yielded a correlation coefficient of 0.76 (subcutaneous sensor) and 0.92 (intravasal sensor). The Bland Altman Plots revealed an accuracy of 2.5 mg/dl, and a precision of + 43.0 mg/dl to - 38.0 mg/dl (subcutaneous sensor) and an accuracy of - 6.0 mg/dl, and a precision of + 12.4 mg/dl to - 24.4 mg/dl (intravasal sensor). No severe hypoglycemic event, defined as BG level below 40 mg/dl, occurred during treatment. Both sensors showed good accuracy in comparison to the BGA values, however they differ regarding precision, which in case of the subcutaneous sensor is considerable high.

KW - Journal Article

U2 - 10.1007/s10877-018-0221-5

DO - 10.1007/s10877-018-0221-5

M3 - SCORING: Journal article

C2 - 30421152

VL - 33

SP - 895

EP - 901

JO - J CLIN MONIT COMPUT

JF - J CLIN MONIT COMPUT

SN - 1387-1307

IS - 5

ER -