Fetal blood flow velocimetry by phase-contrast MRI using a new triggering method and comparison with Doppler ultrasound in a sheep model: a pilot study
Related Research units
Abstract
OBJECT: We present the first study demonstrating the feasibility of antenatal blood flow velocimetry performing ECG triggered phase-contrast (PC)-MRI in the fetal aorta by using a newly developed Doppler ultrasound trigger.
MATERIALS AND METHODS: Five pregnant sheep carrying singleton fetuses (gestational age 121 days) were anesthetized to undergo fetal 2D PC-MRI in the fetal descending aorta (1.5 T) using a newly developed MR-compatible Doppler ultrasound trigger for fetal cardiac triggering. Inter-operator variability was assessed for PC-MR measurements and reproducibility was tested by repeated scans in one fetus. Inter-modality comparison was performed by Doppler ultrasound velocimetry.
RESULTS: Fetal cardiac triggering was possible in all examinations. PC-MR velocimetry revealed a mean inter-operator variability of 3 ± 5 %. Average peak systolic flow velocities of 62.5 ± 4.4 cm/s were in good agreement with Doppler ultrasound measurements of 62.0 ± 9.2 cm/s (p (Lord's U test) ≫ 0.05).
CONCLUSION: Fetal PC-MR velocimetry was successfully performed using the newly developed MR-compatible Doppler ultrasound trigger for intrauterine fetal cardiac triggering, demonstrating high inter-operator and inter-modality agreement. This new method has the high potential for alternative assessment of hemodynamic decompensation of the fetal circulation.
Bibliographical data
Original language | English |
---|---|
ISSN | 1352-8661 |
DOIs | |
Publication status | Published - 01.06.2014 |
PubMed | 23934159 |
---|