Evaluation of the influence of acquisition and reconstruction parameters for 16-row multidetector CT on coronary calcium scoring using a stationary and dynamic cardiac phantom.

Standard

Evaluation of the influence of acquisition and reconstruction parameters for 16-row multidetector CT on coronary calcium scoring using a stationary and dynamic cardiac phantom. / Begemann, Philipp; van Stevendaal, Udo; Köster, Ralf; Mahnken, Andreas H; Koops, Andreas; Adam, Gerhard; Grass, Michael; Nolte-Ernsting, Claus.

In: EUR RADIOL, Vol. 17, No. 8, 8, 2007, p. 1985-1994.

Research output: SCORING: Contribution to journalSCORING: Journal articleResearchpeer-review

Harvard

APA

Vancouver

Bibtex

@article{159c162839084732bf47e79dc5d466b2,
title = "Evaluation of the influence of acquisition and reconstruction parameters for 16-row multidetector CT on coronary calcium scoring using a stationary and dynamic cardiac phantom.",
abstract = "A calcium-scoring phantom with hydroxyapatite-filled cylindrical holes (0.5 to 4 mm) was used. High-resolution scans were performed for an accuracy baseline. The phantom was mounted to a moving heart phantom. Non-moving data with the implementation of an ECG-signal were acquired for different pitches (0.2/0.3), heart rates (60/80/95 bpm) and collimations (16 x 0.75/16 x 1.5 mm). Images were reconstructed with a cone-beam multi-cycle algorithm at a standard thickness/increment of 3 mm/1.5 mm and the thinnest possible thickness (0.8/0.4 and 2/1). Subsequently, ECG-gated moving calcium-scoring phantom data were acquired. The calcium volume and Agatston score were measured. The temporal resolution and reconstruction cycles were calculated. High-resolution scans determine the calcium volume with a high accuracy (mean overestimation, 0.8%). In the non-moving measurements, the volume underestimation ranged from about 6% (16 x 0.75 mm; 0.8/0.4 mm) to nearly 25% (16 x 1.5 mm; 3/1.5 mm). Moving scans showed increased measurement errors depending on the reconstructed RR interval, collimation, pitch, heart rate and gantry rotation time. Also, a correlation with the temporal resolution could be found. The reliability of calcium-scoring results can be improved with the use of a narrower collimation, a lower pitch and the reconstruction of thinner images, resulting in higher patient doses. The choice of the correct cardiac phase within the RR interval is essential to minimize measurement errors.",
author = "Philipp Begemann and {van Stevendaal}, Udo and Ralf K{\"o}ster and Mahnken, {Andreas H} and Andreas Koops and Gerhard Adam and Michael Grass and Claus Nolte-Ernsting",
year = "2007",
language = "Deutsch",
volume = "17",
pages = "1985--1994",
journal = "EUR RADIOL",
issn = "0938-7994",
publisher = "Springer",
number = "8",

}

RIS

TY - JOUR

T1 - Evaluation of the influence of acquisition and reconstruction parameters for 16-row multidetector CT on coronary calcium scoring using a stationary and dynamic cardiac phantom.

AU - Begemann, Philipp

AU - van Stevendaal, Udo

AU - Köster, Ralf

AU - Mahnken, Andreas H

AU - Koops, Andreas

AU - Adam, Gerhard

AU - Grass, Michael

AU - Nolte-Ernsting, Claus

PY - 2007

Y1 - 2007

N2 - A calcium-scoring phantom with hydroxyapatite-filled cylindrical holes (0.5 to 4 mm) was used. High-resolution scans were performed for an accuracy baseline. The phantom was mounted to a moving heart phantom. Non-moving data with the implementation of an ECG-signal were acquired for different pitches (0.2/0.3), heart rates (60/80/95 bpm) and collimations (16 x 0.75/16 x 1.5 mm). Images were reconstructed with a cone-beam multi-cycle algorithm at a standard thickness/increment of 3 mm/1.5 mm and the thinnest possible thickness (0.8/0.4 and 2/1). Subsequently, ECG-gated moving calcium-scoring phantom data were acquired. The calcium volume and Agatston score were measured. The temporal resolution and reconstruction cycles were calculated. High-resolution scans determine the calcium volume with a high accuracy (mean overestimation, 0.8%). In the non-moving measurements, the volume underestimation ranged from about 6% (16 x 0.75 mm; 0.8/0.4 mm) to nearly 25% (16 x 1.5 mm; 3/1.5 mm). Moving scans showed increased measurement errors depending on the reconstructed RR interval, collimation, pitch, heart rate and gantry rotation time. Also, a correlation with the temporal resolution could be found. The reliability of calcium-scoring results can be improved with the use of a narrower collimation, a lower pitch and the reconstruction of thinner images, resulting in higher patient doses. The choice of the correct cardiac phase within the RR interval is essential to minimize measurement errors.

AB - A calcium-scoring phantom with hydroxyapatite-filled cylindrical holes (0.5 to 4 mm) was used. High-resolution scans were performed for an accuracy baseline. The phantom was mounted to a moving heart phantom. Non-moving data with the implementation of an ECG-signal were acquired for different pitches (0.2/0.3), heart rates (60/80/95 bpm) and collimations (16 x 0.75/16 x 1.5 mm). Images were reconstructed with a cone-beam multi-cycle algorithm at a standard thickness/increment of 3 mm/1.5 mm and the thinnest possible thickness (0.8/0.4 and 2/1). Subsequently, ECG-gated moving calcium-scoring phantom data were acquired. The calcium volume and Agatston score were measured. The temporal resolution and reconstruction cycles were calculated. High-resolution scans determine the calcium volume with a high accuracy (mean overestimation, 0.8%). In the non-moving measurements, the volume underestimation ranged from about 6% (16 x 0.75 mm; 0.8/0.4 mm) to nearly 25% (16 x 1.5 mm; 3/1.5 mm). Moving scans showed increased measurement errors depending on the reconstructed RR interval, collimation, pitch, heart rate and gantry rotation time. Also, a correlation with the temporal resolution could be found. The reliability of calcium-scoring results can be improved with the use of a narrower collimation, a lower pitch and the reconstruction of thinner images, resulting in higher patient doses. The choice of the correct cardiac phase within the RR interval is essential to minimize measurement errors.

M3 - SCORING: Zeitschriftenaufsatz

VL - 17

SP - 1985

EP - 1994

JO - EUR RADIOL

JF - EUR RADIOL

SN - 0938-7994

IS - 8

M1 - 8

ER -