Evaluation of a New Multiparameter Brain Probe for Simultaneous Measurement of Brain Tissue Oxygenation, Cerebral Blood Flow, Intracranial Pressure, and Brain Temperature in a Porcine Model

Standard

Evaluation of a New Multiparameter Brain Probe for Simultaneous Measurement of Brain Tissue Oxygenation, Cerebral Blood Flow, Intracranial Pressure, and Brain Temperature in a Porcine Model. / Mader, Marius M; Leidorf, Anna; Hecker, Andreas; Heimann, Axel; Mayr, Petra S M; Kempski, Oliver; Alessandri, Beat; Wöbker, Gabriele.

In: NEUROCRIT CARE, Vol. 29, No. 2, 10.2018, p. 291-301.

Research output: SCORING: Contribution to journalSCORING: Journal articleResearchpeer-review

Harvard

APA

Vancouver

Bibtex

@article{5b2d0164157d4b5cb984f1266aeb967f,
title = "Evaluation of a New Multiparameter Brain Probe for Simultaneous Measurement of Brain Tissue Oxygenation, Cerebral Blood Flow, Intracranial Pressure, and Brain Temperature in a Porcine Model",
abstract = "BACKGROUND: A novel multiparameter brain sensor (MPBS) allows the simultaneous measurement of brain tissue oxygenation (ptiO2), cerebral blood flow (CBF), intracranial pressure (ICP), and brain temperature with a single catheter. This laboratory investigation evaluates the MPBS in an animal model in relation to established reference probes.METHODS: The study group consisted of 17 juvenile male pigs. Four MPBS and four reference probes were implanted per pig and compared simultaneously. The measured parameters were challenged by standardized provocations such as hyperoxia, dobutamine, and norepinephrine application, hypercapnia and hypoxia in combination with and without a controlled cortical impact (CCI) injury. Mean values over 2 min were collected for predefined time points and were analyzed using Bland-Altman plots.RESULTS: The protocol was successfully conducted in 15 pigs of which seven received CCI. ICP and ptiO2 were significantly influenced by the provocations. Subtraction of MPBS from reference values revealed a mean difference (limits of agreement) of 3.7 (- 20.5 to 27.9) mm Hg, - 2.9 (- 7.9 to 2.1) mm Hg, and 5.1 (- 134.7 to 145.0) % for ptiO2, ICP, and relative CBF, respectively.CONCLUSIONS: The MPBS is a promising measurement tool for multiparameter neuromonitoring. The conducted study demonstrates the in vivo functionality of the probe. Comparison with standard probes revealed a deviation which is mostly analogous to other multiparameter devices. However, further evaluation of the device is necessary before it can reliably be used for clinical decision making.",
keywords = "Journal Article",
author = "Mader, {Marius M} and Anna Leidorf and Andreas Hecker and Axel Heimann and Mayr, {Petra S M} and Oliver Kempski and Beat Alessandri and Gabriele W{\"o}bker",
year = "2018",
month = oct,
doi = "10.1007/s12028-018-0541-9",
language = "English",
volume = "29",
pages = "291--301",
journal = "NEUROCRIT CARE",
issn = "1541-6933",
publisher = "Humana Press",
number = "2",

}

RIS

TY - JOUR

T1 - Evaluation of a New Multiparameter Brain Probe for Simultaneous Measurement of Brain Tissue Oxygenation, Cerebral Blood Flow, Intracranial Pressure, and Brain Temperature in a Porcine Model

AU - Mader, Marius M

AU - Leidorf, Anna

AU - Hecker, Andreas

AU - Heimann, Axel

AU - Mayr, Petra S M

AU - Kempski, Oliver

AU - Alessandri, Beat

AU - Wöbker, Gabriele

PY - 2018/10

Y1 - 2018/10

N2 - BACKGROUND: A novel multiparameter brain sensor (MPBS) allows the simultaneous measurement of brain tissue oxygenation (ptiO2), cerebral blood flow (CBF), intracranial pressure (ICP), and brain temperature with a single catheter. This laboratory investigation evaluates the MPBS in an animal model in relation to established reference probes.METHODS: The study group consisted of 17 juvenile male pigs. Four MPBS and four reference probes were implanted per pig and compared simultaneously. The measured parameters were challenged by standardized provocations such as hyperoxia, dobutamine, and norepinephrine application, hypercapnia and hypoxia in combination with and without a controlled cortical impact (CCI) injury. Mean values over 2 min were collected for predefined time points and were analyzed using Bland-Altman plots.RESULTS: The protocol was successfully conducted in 15 pigs of which seven received CCI. ICP and ptiO2 were significantly influenced by the provocations. Subtraction of MPBS from reference values revealed a mean difference (limits of agreement) of 3.7 (- 20.5 to 27.9) mm Hg, - 2.9 (- 7.9 to 2.1) mm Hg, and 5.1 (- 134.7 to 145.0) % for ptiO2, ICP, and relative CBF, respectively.CONCLUSIONS: The MPBS is a promising measurement tool for multiparameter neuromonitoring. The conducted study demonstrates the in vivo functionality of the probe. Comparison with standard probes revealed a deviation which is mostly analogous to other multiparameter devices. However, further evaluation of the device is necessary before it can reliably be used for clinical decision making.

AB - BACKGROUND: A novel multiparameter brain sensor (MPBS) allows the simultaneous measurement of brain tissue oxygenation (ptiO2), cerebral blood flow (CBF), intracranial pressure (ICP), and brain temperature with a single catheter. This laboratory investigation evaluates the MPBS in an animal model in relation to established reference probes.METHODS: The study group consisted of 17 juvenile male pigs. Four MPBS and four reference probes were implanted per pig and compared simultaneously. The measured parameters were challenged by standardized provocations such as hyperoxia, dobutamine, and norepinephrine application, hypercapnia and hypoxia in combination with and without a controlled cortical impact (CCI) injury. Mean values over 2 min were collected for predefined time points and were analyzed using Bland-Altman plots.RESULTS: The protocol was successfully conducted in 15 pigs of which seven received CCI. ICP and ptiO2 were significantly influenced by the provocations. Subtraction of MPBS from reference values revealed a mean difference (limits of agreement) of 3.7 (- 20.5 to 27.9) mm Hg, - 2.9 (- 7.9 to 2.1) mm Hg, and 5.1 (- 134.7 to 145.0) % for ptiO2, ICP, and relative CBF, respectively.CONCLUSIONS: The MPBS is a promising measurement tool for multiparameter neuromonitoring. The conducted study demonstrates the in vivo functionality of the probe. Comparison with standard probes revealed a deviation which is mostly analogous to other multiparameter devices. However, further evaluation of the device is necessary before it can reliably be used for clinical decision making.

KW - Journal Article

U2 - 10.1007/s12028-018-0541-9

DO - 10.1007/s12028-018-0541-9

M3 - SCORING: Journal article

C2 - 29949006

VL - 29

SP - 291

EP - 301

JO - NEUROCRIT CARE

JF - NEUROCRIT CARE

SN - 1541-6933

IS - 2

ER -