Contrast-dose relation in first-pass myocardial MR perfusion imaging.

  • Wolfgang Utz
  • Thoralf Niendorf
  • Ralf Wassmuth
  • Daniel Messroghli
  • Rainer Dietz
  • Jeanette Schulz-Menger

Related Research units

Abstract

PURPOSE: To determine the regime of linear contrast enhancement in human first-pass perfusion cardiovascular magnetic resonance (CMR) imaging to improve accuracy in myocardial perfusion quantification. MATERIALS AND METHODS: A total of 10 healthy subjects were studied on a clinical 1.5T MR scanner. Seven doses of Gd-DTPA ranging from 0.00125 to 0.1 mmol/kg of body weight (b.w.) were administered as equal volumes by rapid bolus injection (6 mL/second). Resting periods of 15 minutes were introduced after delivery of Gd doses >0.01 mmol/kg b.w. For each subject, two series of rest perfusion scans were performed using two different multislice saturation-recovery perfusion sequences. Maximum contrast enhancement and maximum upslope were obtained in the blood pool of the left ventricular (LV) cavity and in the myocardium. The range of linear contrast-dose relation was determined by linear regression analysis. RESULTS: MR signal intensity increased linearly for contrast agent concentrations up to 0.01 mmol/kg b.w. in the LV blood pool and up to 0.05 mmol/kg b.w. in the myocardium. For Gd concentrations exceeding these thresholds the signal intensity response was not linear with respect to the contrast agent dose. CONCLUSION: Quantitative evaluation of cardiac MR perfusion data needs to account for signal saturation in both the LV blood pool and the myocardium.

Bibliographical data

Original languageGerman
Article number6
ISSN1053-1807
Publication statusPublished - 2007
pubmed 17520736