Comparison of antero-lateral versus antero-posterior electrode position for biphasic external cardioversion of atrial flutter

  • Tim Risius
  • Kai Mortensen
  • Tjark F Schwemer
  • Muhammet A Aydin
  • Hanno U Klemm
  • Rodolfo Ventura
  • Achim Barmeyer
  • Boris Hoffmann
  • Thomas Rostock
  • Thomas Meinertz
  • Stephan Willems

Abstract

External cardioversion is an established and very important tool to terminate symptomatic atrial flutter. The superiority of the biphasic waveform has been demonstrated for atrial flutter, but whether electrode position affects the efficacy of cardioversion in this population is not known. The aim of this trial was to evaluate whether anterior-lateral (A-L) compared with anterior-posterior (A-P) electrode position improves cardioversion results. Of 130 screened patients, 96 (72 men, mean age 62 +/- 12 years) were included and randomly assigned to a cardioversion protocol with either A-L or A-P electrode position. In each group, 48 patients received sequential biphasic waveform shocks using a step-up protocol consisting of 50, 75, 100, 150, or 200 J. The mean energy (65 +/- 13 J for A-L vs 77 +/- 13 J for A-P, p = 0.001) and mean number of shocks (1.48 +/- 1.01 for A-L vs 1.96 +/- 1.00 for A-P, p = 0.001) required for successful cardioversion were significantly lower in the A-L group. The efficacy of the first shock with 50 J in the A-L electrode position (35 of 48 patients [73%]) was also highly significantly greater than the first shock with 50 J in the A-P electrode position (18 of 48 patients [36%]) (p = 0.001). In conclusion, the A-L electrode position increases efficacy and requires fewer energy and shocks in external electrical cardioversion of common atrial flutter. Therefore, A-L electrode positioning should be recommended for the external cardioversion of common atrial flutter.

Bibliographical data

Original languageEnglish
ISSN0002-9149
DOIs
Publication statusPublished - 01.12.2009
PubMed 19932790