Autocalibrating pulse contour analysis based on radial artery applanation tonometry for continuous non-invasive cardiac output monitoring in intensive care unit patients after major gastrointestinal surgery--a prospective method comparison study

Standard

Harvard

APA

Vancouver

Bibtex

@article{1ea4b29cdba64009b7f0539b2ba12142,
title = "Autocalibrating pulse contour analysis based on radial artery applanation tonometry for continuous non-invasive cardiac output monitoring in intensive care unit patients after major gastrointestinal surgery--a prospective method comparison study",
abstract = "The T-Line({\textregistered}) system (Tensys({\textregistered}) Medical Inc., San Diego, CA, USA) non-invasively estimates cardiac output (CO) using autocalibrating pulse contour analysis of the radial artery applanation tonometry-derived arterial waveform. We compared T-Line CO measurements (TL-CO) with invasively obtained CO measurements using transpulmonary thermodilution (TDCO) and calibrated pulse contour analysis (PC-CO) in patients after major gastrointestinal surgery. We compared 1) TL-CO versus TD-CO and 2) TL-CO versus PC-CO in 27 patients treated in the intensive care unit (ICU) after major gastrointestinal surgery. For the assessment of TD-CO and PC-CO we used the PiCCO({\textregistered}) system (Pulsion Medical Systems SE, Feldkirchen, Germany). Per patient, we compared two sets of TD-CO and 30 minutes of PC-CO measurements with the simultaneously recorded TL-CO values using Bland-Altman analysis. The mean of differences (± standard deviation; 95% limits of agreement) between TL-CO and TD-CO was -0.8 (±1.6; -4.0 to +2.3) l/minute with a percentage error of 45%. For TL-CO versus PC-CO, we observed a mean of differences of -0.4 (±1.5; -3.4 to +2.5) l/minute with a percentage error of 43%. In ICU patients after major gastrointestinal surgery, continuous non-invasive CO measurement based on autocalibrating pulse contour analysis of the radial artery applanation tonometry-derived arterial waveform (TL-CO) is feasible in a clinical study setting. However, the agreement of TL-CO with TD-CO and PC-CO observed in our study indicates that further improvements are needed before the technology can be recommended for clinical use in these patients.",
keywords = "Aged, Calibration, Cardiac Output, Digestive System Surgical Procedures, Female, Heart Rate, Humans, Intensive Care Units, Male, Manometry, Middle Aged, Prospective Studies, Radial Artery, Thermodilution",
author = "Wagner, {J Y} and M Langemann and G Sch{\"o}n and S Kluge and Reuter, {D A} and B Saugel",
year = "2016",
month = may,
language = "English",
volume = "44",
pages = "340--5",
journal = "ANAESTH INTENS CARE",
issn = "0310-057X",
publisher = "Australian Society of Anaesthetists",
number = "3",

}

RIS

TY - JOUR

T1 - Autocalibrating pulse contour analysis based on radial artery applanation tonometry for continuous non-invasive cardiac output monitoring in intensive care unit patients after major gastrointestinal surgery--a prospective method comparison study

AU - Wagner, J Y

AU - Langemann, M

AU - Schön, G

AU - Kluge, S

AU - Reuter, D A

AU - Saugel, B

PY - 2016/5

Y1 - 2016/5

N2 - The T-Line(®) system (Tensys(®) Medical Inc., San Diego, CA, USA) non-invasively estimates cardiac output (CO) using autocalibrating pulse contour analysis of the radial artery applanation tonometry-derived arterial waveform. We compared T-Line CO measurements (TL-CO) with invasively obtained CO measurements using transpulmonary thermodilution (TDCO) and calibrated pulse contour analysis (PC-CO) in patients after major gastrointestinal surgery. We compared 1) TL-CO versus TD-CO and 2) TL-CO versus PC-CO in 27 patients treated in the intensive care unit (ICU) after major gastrointestinal surgery. For the assessment of TD-CO and PC-CO we used the PiCCO(®) system (Pulsion Medical Systems SE, Feldkirchen, Germany). Per patient, we compared two sets of TD-CO and 30 minutes of PC-CO measurements with the simultaneously recorded TL-CO values using Bland-Altman analysis. The mean of differences (± standard deviation; 95% limits of agreement) between TL-CO and TD-CO was -0.8 (±1.6; -4.0 to +2.3) l/minute with a percentage error of 45%. For TL-CO versus PC-CO, we observed a mean of differences of -0.4 (±1.5; -3.4 to +2.5) l/minute with a percentage error of 43%. In ICU patients after major gastrointestinal surgery, continuous non-invasive CO measurement based on autocalibrating pulse contour analysis of the radial artery applanation tonometry-derived arterial waveform (TL-CO) is feasible in a clinical study setting. However, the agreement of TL-CO with TD-CO and PC-CO observed in our study indicates that further improvements are needed before the technology can be recommended for clinical use in these patients.

AB - The T-Line(®) system (Tensys(®) Medical Inc., San Diego, CA, USA) non-invasively estimates cardiac output (CO) using autocalibrating pulse contour analysis of the radial artery applanation tonometry-derived arterial waveform. We compared T-Line CO measurements (TL-CO) with invasively obtained CO measurements using transpulmonary thermodilution (TDCO) and calibrated pulse contour analysis (PC-CO) in patients after major gastrointestinal surgery. We compared 1) TL-CO versus TD-CO and 2) TL-CO versus PC-CO in 27 patients treated in the intensive care unit (ICU) after major gastrointestinal surgery. For the assessment of TD-CO and PC-CO we used the PiCCO(®) system (Pulsion Medical Systems SE, Feldkirchen, Germany). Per patient, we compared two sets of TD-CO and 30 minutes of PC-CO measurements with the simultaneously recorded TL-CO values using Bland-Altman analysis. The mean of differences (± standard deviation; 95% limits of agreement) between TL-CO and TD-CO was -0.8 (±1.6; -4.0 to +2.3) l/minute with a percentage error of 45%. For TL-CO versus PC-CO, we observed a mean of differences of -0.4 (±1.5; -3.4 to +2.5) l/minute with a percentage error of 43%. In ICU patients after major gastrointestinal surgery, continuous non-invasive CO measurement based on autocalibrating pulse contour analysis of the radial artery applanation tonometry-derived arterial waveform (TL-CO) is feasible in a clinical study setting. However, the agreement of TL-CO with TD-CO and PC-CO observed in our study indicates that further improvements are needed before the technology can be recommended for clinical use in these patients.

KW - Aged

KW - Calibration

KW - Cardiac Output

KW - Digestive System Surgical Procedures

KW - Female

KW - Heart Rate

KW - Humans

KW - Intensive Care Units

KW - Male

KW - Manometry

KW - Middle Aged

KW - Prospective Studies

KW - Radial Artery

KW - Thermodilution