Acute lesion extension following pulmonary vein isolation with two novel single shot devices: Pulsed field ablation versus multielectrode radiofrequency balloon

Standard

Acute lesion extension following pulmonary vein isolation with two novel single shot devices: Pulsed field ablation versus multielectrode radiofrequency balloon. / My, Ilaria; Lemoine, Marc D; Butt, Mahi; Mencke, Celine; Loeck, Fabian W; Obergassel, Julius; Rottner, Laura; Wenzel, Jan-Per; Schleberger, Ruben; Moser, Julia; Moser, Fabian; Kirchhof, Paulus; Reissmann, Bruno; Ouyang, Feifan; Rillig, Andreas; Metzner, Andreas.

In: J CARDIOVASC ELECTR, Vol. 34, No. 9, 09.2023, p. 1802-1807.

Research output: SCORING: Contribution to journalSCORING: Journal articleResearchpeer-review

Harvard

APA

Vancouver

Bibtex

@article{13d8e708a0e0444ab224aa2a8ed7a03e,
title = "Acute lesion extension following pulmonary vein isolation with two novel single shot devices: Pulsed field ablation versus multielectrode radiofrequency balloon",
abstract = "INTRODUCTION: Pulsed-field ablation (PFA) and the multielectrode radiofrequency balloon (RFB) are two novel ablation technologies to perform pulmonary vein isolation (PVI). It is currently unknown whether these technologies differ in lesion formation and lesion extent. We compared the acute lesion extent after PVI induced by PFA and RFB by measuring low-voltage area in high-density maps and the release of biomolecules reflecting cardiac injury.METHODS: PVI was performed with a pentaspline catheter (FARAPULSE) applying PFA or with the compliant multielectrode RFB (HELIOSTAR). Before and after PVI high-density mapping with CARTO 3 was performed. In addition, blood samples were taken before transseptal puncture and after post-PVI remapping and serum concentrations of high-sensitive Troponin I were quantified by immunoassay.RESULTS: Sixty patients undergoing PVI by PFA (n = 28, age 69 ± 12 year, 60% males, 39.3% persistent atrial fibrillation [AF]) or RFB (n = 32, age 65 ± 13 year, 53% males, 21.9% persistent AF) were evaluated. Acute PVI was achieved in all patients in both groups. Mean number of PFA pulses was 34.2 ± 4.5 and mean number RFB applications was 8.5 ± 3 per patient. Total posterior ablation area was significantly larger in PFA (20.7 ± 7.7 cm²) than in RFB (7.1 ± 2.09 cm²; p < .001). Accordingly, posterior ablation area for each PV resulted in larger lesions after PFA versus RFB (LSPV 5.2 ± 2.7 vs. 1.9 ± 0.8 cm², LIPV 5.5 ± 2.3 vs. 1.9 ± 0.8 cm², RSPV 4.7 ± 1.9 vs. 1.6 ± 0.5 cm², RIPV 5.3 ± 2.1 vs. 1.6 ± 0.7 cm,² respectively; p < .001). In a subset of 38 patients, increase of hsTropI was higher after PFA (625 ± 138 pg/mL, n = 28) versus RFB (148 ± 36 pg/mL, n = 10; p = .049) supporting the evidence of larger lesion extent by PFA.CONCLUSION: PFA delivers larger acute lesion areas and higher troponin release upon successful PVI than multielectrode RFB-based PVI in this single-center series.",
keywords = "Male, Humans, Middle Aged, Aged, Aged, 80 and over, Female, Pulmonary Veins/surgery, Treatment Outcome, Catheter Ablation/adverse effects, Atrial Fibrillation/diagnosis, Catheters",
author = "Ilaria My and Lemoine, {Marc D} and Mahi Butt and Celine Mencke and Loeck, {Fabian W} and Julius Obergassel and Laura Rottner and Jan-Per Wenzel and Ruben Schleberger and Julia Moser and Fabian Moser and Paulus Kirchhof and Bruno Reissmann and Feifan Ouyang and Andreas Rillig and Andreas Metzner",
note = "{\textcopyright} 2023 The Authors. Journal of Cardiovascular Electrophysiology published by Wiley Periodicals LLC.",
year = "2023",
month = sep,
doi = "10.1111/jce.16001",
language = "English",
volume = "34",
pages = "1802--1807",
journal = "J CARDIOVASC ELECTR",
issn = "1045-3873",
publisher = "Wiley-Blackwell",
number = "9",

}

RIS

TY - JOUR

T1 - Acute lesion extension following pulmonary vein isolation with two novel single shot devices: Pulsed field ablation versus multielectrode radiofrequency balloon

AU - My, Ilaria

AU - Lemoine, Marc D

AU - Butt, Mahi

AU - Mencke, Celine

AU - Loeck, Fabian W

AU - Obergassel, Julius

AU - Rottner, Laura

AU - Wenzel, Jan-Per

AU - Schleberger, Ruben

AU - Moser, Julia

AU - Moser, Fabian

AU - Kirchhof, Paulus

AU - Reissmann, Bruno

AU - Ouyang, Feifan

AU - Rillig, Andreas

AU - Metzner, Andreas

N1 - © 2023 The Authors. Journal of Cardiovascular Electrophysiology published by Wiley Periodicals LLC.

PY - 2023/9

Y1 - 2023/9

N2 - INTRODUCTION: Pulsed-field ablation (PFA) and the multielectrode radiofrequency balloon (RFB) are two novel ablation technologies to perform pulmonary vein isolation (PVI). It is currently unknown whether these technologies differ in lesion formation and lesion extent. We compared the acute lesion extent after PVI induced by PFA and RFB by measuring low-voltage area in high-density maps and the release of biomolecules reflecting cardiac injury.METHODS: PVI was performed with a pentaspline catheter (FARAPULSE) applying PFA or with the compliant multielectrode RFB (HELIOSTAR). Before and after PVI high-density mapping with CARTO 3 was performed. In addition, blood samples were taken before transseptal puncture and after post-PVI remapping and serum concentrations of high-sensitive Troponin I were quantified by immunoassay.RESULTS: Sixty patients undergoing PVI by PFA (n = 28, age 69 ± 12 year, 60% males, 39.3% persistent atrial fibrillation [AF]) or RFB (n = 32, age 65 ± 13 year, 53% males, 21.9% persistent AF) were evaluated. Acute PVI was achieved in all patients in both groups. Mean number of PFA pulses was 34.2 ± 4.5 and mean number RFB applications was 8.5 ± 3 per patient. Total posterior ablation area was significantly larger in PFA (20.7 ± 7.7 cm²) than in RFB (7.1 ± 2.09 cm²; p < .001). Accordingly, posterior ablation area for each PV resulted in larger lesions after PFA versus RFB (LSPV 5.2 ± 2.7 vs. 1.9 ± 0.8 cm², LIPV 5.5 ± 2.3 vs. 1.9 ± 0.8 cm², RSPV 4.7 ± 1.9 vs. 1.6 ± 0.5 cm², RIPV 5.3 ± 2.1 vs. 1.6 ± 0.7 cm,² respectively; p < .001). In a subset of 38 patients, increase of hsTropI was higher after PFA (625 ± 138 pg/mL, n = 28) versus RFB (148 ± 36 pg/mL, n = 10; p = .049) supporting the evidence of larger lesion extent by PFA.CONCLUSION: PFA delivers larger acute lesion areas and higher troponin release upon successful PVI than multielectrode RFB-based PVI in this single-center series.

AB - INTRODUCTION: Pulsed-field ablation (PFA) and the multielectrode radiofrequency balloon (RFB) are two novel ablation technologies to perform pulmonary vein isolation (PVI). It is currently unknown whether these technologies differ in lesion formation and lesion extent. We compared the acute lesion extent after PVI induced by PFA and RFB by measuring low-voltage area in high-density maps and the release of biomolecules reflecting cardiac injury.METHODS: PVI was performed with a pentaspline catheter (FARAPULSE) applying PFA or with the compliant multielectrode RFB (HELIOSTAR). Before and after PVI high-density mapping with CARTO 3 was performed. In addition, blood samples were taken before transseptal puncture and after post-PVI remapping and serum concentrations of high-sensitive Troponin I were quantified by immunoassay.RESULTS: Sixty patients undergoing PVI by PFA (n = 28, age 69 ± 12 year, 60% males, 39.3% persistent atrial fibrillation [AF]) or RFB (n = 32, age 65 ± 13 year, 53% males, 21.9% persistent AF) were evaluated. Acute PVI was achieved in all patients in both groups. Mean number of PFA pulses was 34.2 ± 4.5 and mean number RFB applications was 8.5 ± 3 per patient. Total posterior ablation area was significantly larger in PFA (20.7 ± 7.7 cm²) than in RFB (7.1 ± 2.09 cm²; p < .001). Accordingly, posterior ablation area for each PV resulted in larger lesions after PFA versus RFB (LSPV 5.2 ± 2.7 vs. 1.9 ± 0.8 cm², LIPV 5.5 ± 2.3 vs. 1.9 ± 0.8 cm², RSPV 4.7 ± 1.9 vs. 1.6 ± 0.5 cm², RIPV 5.3 ± 2.1 vs. 1.6 ± 0.7 cm,² respectively; p < .001). In a subset of 38 patients, increase of hsTropI was higher after PFA (625 ± 138 pg/mL, n = 28) versus RFB (148 ± 36 pg/mL, n = 10; p = .049) supporting the evidence of larger lesion extent by PFA.CONCLUSION: PFA delivers larger acute lesion areas and higher troponin release upon successful PVI than multielectrode RFB-based PVI in this single-center series.

KW - Male

KW - Humans

KW - Middle Aged

KW - Aged

KW - Aged, 80 and over

KW - Female

KW - Pulmonary Veins/surgery

KW - Treatment Outcome

KW - Catheter Ablation/adverse effects

KW - Atrial Fibrillation/diagnosis

KW - Catheters

U2 - 10.1111/jce.16001

DO - 10.1111/jce.16001

M3 - SCORING: Journal article

C2 - 37473404

VL - 34

SP - 1802

EP - 1807

JO - J CARDIOVASC ELECTR

JF - J CARDIOVASC ELECTR

SN - 1045-3873

IS - 9

ER -