Accelerating image acquisition in 64-MDCT: the influence of scan parameters on image resolution and quality in a phantom study.

  • Thorsten Klink
  • Marc Regier
  • Udo van Stevendaal
  • Michael Grass
  • Gerhard Adam
  • Philipp Begemann

Abstract

Computed tomographic (CT) image resolution and quality were evaluated utilizing varying scan protocols with accelerated image acquisition. A resolution phantom with hole diameters from 0.2 to 1.0 mm was scanned in axial, coronal, and sagittal plane using a 64-slice multidetector CT with varying scan parameters. No relevant differences in image resolution and quality were detected between the fastest scan protocol, with the shortest rotation time and highest pitch, and the slowest protocol. Accelerated CT protocols resulted in diagnostic images with adequate resolution and quality.

Bibliographical data

Original languageEnglish
Article number4
ISSN0899-7071
Publication statusPublished - 2012
pubmed 22726972