A multi-vendor, multi-center study on reproducibility and comparability of fast strain-encoded cardiovascular magnetic resonance imaging

  • Jennifer Erley
  • Victoria Zieschang
  • Tomas Lapinskas
  • Aylin Demir
  • Stephanie Wiesemann
  • Markus Haass
  • Nael F Osman
  • Orlando P Simonetti
  • Yingmin Liu
  • Amit R Patel
  • Victor Mor-Avi
  • Orhan Unal
  • Kevin M Johnson
  • Burkert Pieske
  • Jochen Hansmann
  • Jeanette Schulz-Menger
  • Sebastian Kelle

Abstract

Myocardial strain is a convenient parameter to quantify left ventricular (LV) function. Fast strain-encoding (fSENC) enables the acquisition of cardiovascular magnetic resonance images for strain-measurement within a few heartbeats during free-breathing. It is necessary to analyze inter-vendor agreement of techniques to determine strain, such as fSENC, in order to compare existing studies and plan multi-center studies. Therefore, the aim of this study was to investigate inter-vendor agreement and test-retest reproducibility of fSENC for three major MRI-vendors. fSENC-images were acquired three times in the same group of 15 healthy volunteers using 3 Tesla scanners from three different vendors: at the German Heart Institute Berlin, the Charité University Medicine Berlin-Campus Buch and the Theresien-Hospital Mannheim. Volunteers were scanned using the same imaging protocol composed of two fSENC-acquisitions, a 15-min break and another two fSENC-acquisitions. LV global longitudinal and circumferential strain (GLS, GCS) were analyzed by a trained observer (Myostrain 5.0, Myocardial Solutions) and for nine volunteers repeatedly by another observer. Inter-vendor agreement was determined using Bland-Altman analysis. Test-retest reproducibility and intra- and inter-observer reproducibility were analyzed using intraclass correlation coefficient (ICC) and coefficients of variation (CoV). Inter-vendor agreement between all three sites was good for GLS and GCS, with biases of 0.01-1.88%. Test-retest reproducibility of scans before and after the break was high, shown by ICC- and CoV values of 0.63-0.97 and 3-9% for GLS and 0.69-0.82 and 4-7% for GCS, respectively. Intra- and inter-observer reproducibility were excellent for both parameters (ICC of 0.77-0.99, CoV of 2-5%). This trial demonstrates good inter-vendor agreement and test-retest reproducibility of GLS and GCS measurements, acquired at three different scanners from three different vendors using fSENC. The results indicate that it is necessary to account for a possible bias (< 2%) when comparing strain measurements of different scanners. Technical differences between scanners, which impact inter-vendor agreement, should be further analyzed and minimized.DRKS Registration Number: 00013253.Universal Trial Number (UTN): U1111-1207-5874.

Bibliographical data

Original languageEnglish
ISSN1569-5794
DOIs
Publication statusPublished - 05.2020
Externally publishedYes
PubMed 32056087