The role of cone-beam breast-CT for breast cancer detection relative to breast density

  • Susanne Wienbeck
  • Johannes Uhlig
  • Susanne Luftner-Nagel
  • Antonia Zapf
  • Alexey Surov
  • Eva von Fintel
  • Vera Stahnke
  • Joachim Lotz
  • Uwe Fischer

Abstract

OBJECTIVES: To evaluate the impact of breast density on the diagnostic accuracy of non-contrast cone-beam breast computed tomography (CBBCT) in comparison to mammography for the detection of breast masses. METHODS: A retrospective study was conducted from August 2015 to July 2016. Fifty-nine patients (65 breasts, 112 lesions) with BI-RADS, 5th edition 4 or 5 assessment in mammography and/or ultrasound of the breast received an additional non-contrast CBBCT. Independent double blind reading by two radiologists was performed for mammography and CBBCT imaging. Sensitivity, specificity and AUC were compared between the modalities. RESULTS: Breast lesions were histologically examined in 85 of 112 lesions (76%). The overall sensitivity for CBBCT (reader 1: 91%, reader 2: 88%) was higher than in mammography (both: 68%, p<0.001), and also for the high-density group (p<0.05). The specificity and AUC was higher for mammography in comparison to CBBCT (p<0.05 and p<0.001). The interobserver agreement (ICC) between the readers was 90% (95% CI: 86-93%) for mammography and 87% (95% CI: 82-91%) for CBBCT. CONCLUSIONS: Compared with two-view mammography, non-contrast CBBCT has higher sensitivity, lower specificity, and lower AUC for breast mass detection in both high and low density breasts. KEY POINTS: • Overall sensitivity for non-contrast CBBCT ranged between 88%-91%. • Sensitivity was higher for CBBCT than mammography in both density types (p<0.001). • Specificity was higher for mammography than CBBCT in both density types (p<0.05). • AUC was larger for mammography than CBBCT in both density types (p<0.001).

Bibliografische Daten

OriginalspracheEnglisch
ISSN0938-7994
DOIs
StatusVeröffentlicht - 12.2017
PubMed 28677053