The effect of load on heat production, thermal effects and expenditure of time during implant site preparation - an experimental ex vivo comparison between piezosurgery and conventional drilling
Abstract
OBJECTIVES: Piezoelectric surgery (PS) is meant to be a gentle osteotomy method. The aim of this study was to compare piezosurgical vs. conventional drilling methods for implant site preparation (ISP) - focusing on load-dependent thermal effect on hard tissue and the expenditure of ISP time.
MATERIALS AND METHODS: Three hundred and sixty ISP were performed on ex vivo pig heads using piezosurgery, spiral burs (SB) and trephine burs (TB). The load applied was increased from 0 to 1000 g in 100-g intervals. Temperature within the bone was measured with a thermocouple, and duration was recorded with a stop watch. Thermal effects were histomorphometrically analysed. Twelve ISPs per technique were performed at the lateral wall of the maxillary sinus.
RESULTS: PS yields the highest mean temperatures (48.6 ± 3.4°C) and thermal effects (200.7 ± 44.4 μm), both at 900-1000 g. Duration is reduced with a plus of load and significantly longer in either case for PS (P < 0.05). There is a correlation of the applied load with all other examined factors for PS and TB. Temperature and histological effects decrease for SB beyond 500 g.
CONCLUSIONS: PS yields significantly higher temperatures and thermal tissue alterations on load levels higher than 500 g and is significantly slower for ISP compared to SB and TB. For ISP with PS, a maximum load of 400 g should be maintained.