Sensitivity of High-Pitch Dual-Source Computed Tomography for the Detection of Anomalous Pulmonary Venous Connection in Infants

Abstract

PURPOSE:  To evaluate the sensitivity, specificity, and interobserver reliability of high-pitch dual-source computed tomography angiography (CTA) in the detection of anomalous pulmonary venous connection (APVC) in infants with congenital heart defects and to assess the associated radiation exposure.

MATERIALS AND METHODS:  78 pulmonary veins in 17 consecutively enrolled patients with congenital heart defects (6 females; 11 males; median age: 6 days; range: 1-299 days) were retrospectively included in this study. All patients underwent high-pitch dual-source CTA of the chest at low tube voltages (70 kV). APVC was evaluated independently by two radiologists. Sensitivity, specificity, positive (PPV) and negative predictive values (NPV), and interobserver agreement were determined. For standard of reference, one additional observer reviewed CT scans, echocardiography reports, clinical reports as well as surgical reports. In cases of disagreement the additional observer made the final decision based on all available information.

RESULTS:  Detection of APVC with high-pitch dual-source CTA revealed a good sensitivity (91 %) and specificity (99 %), with PPV and NPV of 98 % and 97 %. Interobserver agreement was almost perfect (Kappa = 0.84). The median DLP was 3.8 mGy*cm (IQR 3.3-4.7 mGy*cm) and the median radiation dose was 0.33 mSv (IQR 0.26-0.39 mSv).

CONCLUSION:  High-pitch dual-source CTA in infants with congenital heart defects allows for accurate and reliable assessment of APVC at a low radiation dose.

KEY POINTS:   · High-pitch dual-source CTA enables detection of anomalous pulmonary vein connection with high sensitivity in infants.. · Interrater reliability in the detection of anomalous pulmonary vein connection with high-pitch dual-source CTA is almost perfect.. · Radiation dose of high-pitch dual-source CTA in the cardiac examination of infants is low..

CITATION FORMAT: · Well L, Weinrich JM, Meyer M et al. Sensitivity of High-Pitch Dual-Source Computed Tomography for the Detection of Anomalous Pulmonary Venous Connection in Infants. Fortschr Röntgenstr 2021; 193: 551 - 558.