Proposing Metrics for Benchmarking Novel EEG Technologies Towards Real-World Measurements

Standard

Proposing Metrics for Benchmarking Novel EEG Technologies Towards Real-World Measurements. / Oliveira, Anderson S; Schlink, Bryan R; Hairston, W David; König, Peter; Ferris, Daniel P.

in: FRONT HUM NEUROSCI, Jahrgang 10, Nr. 10, 2016, S. 188.

Publikationen: SCORING: Beitrag in Fachzeitschrift/ZeitungSCORING: ZeitschriftenaufsatzForschungBegutachtung

Harvard

Oliveira, AS, Schlink, BR, Hairston, WD, König, P & Ferris, DP 2016, 'Proposing Metrics for Benchmarking Novel EEG Technologies Towards Real-World Measurements', FRONT HUM NEUROSCI, Jg. 10, Nr. 10, S. 188. https://doi.org/10.3389/fnhum.2016.00188

APA

Oliveira, A. S., Schlink, B. R., Hairston, W. D., König, P., & Ferris, D. P. (2016). Proposing Metrics for Benchmarking Novel EEG Technologies Towards Real-World Measurements. FRONT HUM NEUROSCI, 10(10), 188. https://doi.org/10.3389/fnhum.2016.00188

Vancouver

Bibtex

@article{abf33b70428843f4a8e8b4d2a39a9b0b,
title = "Proposing Metrics for Benchmarking Novel EEG Technologies Towards Real-World Measurements",
abstract = "Recent advances in electroencephalographic (EEG) acquisition allow for recordings using wet and dry sensors during whole-body motion. The large variety of commercially available EEG systems contrasts with the lack of established methods for objectively describing their performance during whole-body motion. Therefore, the aim of this study was to introduce methods for benchmarking the suitability of new EEG technologies for that context. Subjects performed an auditory oddball task using three different EEG systems (Biosemi wet-BSM, Cognionics Wet-Cwet, Conionics Dry-Cdry). Nine subjects performed the oddball task while seated and walking on a treadmill. We calculated EEG epoch rejection rate, pre-stimulus noise (PSN), signal-to-noise ratio (SNR) and EEG amplitude variance across the P300 event window (CVERP) from a subset of 12 channels common to all systems. We also calculated test-retest reliability and the subject's level of comfort while using each system. Our results showed that using the traditional 75 μV rejection threshold BSM and Cwet epoch rejection rates are ~25% and ~47% in the seated and walking conditions respectively. However, this threshold rejects ~63% of epochs for Cdry in the seated condition and excludes 100% of epochs for the majority of subjects during walking. BSM showed predominantly no statistical differences between seated and walking condition for all metrics, whereas Cwet showed increases in PSN and CVERP, as well as reduced SNR in the walking condition. Data quality from Cdry in seated conditions were predominantly inferior in comparison to the wet systems. Test-retest reliability was mostly moderate/good for these variables, especially in seated conditions. In addition, subjects felt less discomfort and were motivated for longer recording periods while using wet EEG systems in comparison to the dry system. The proposed method was successful in identifying differences across systems that are mostly caused by motion-related artifacts and usability issues. We conclude that the extraction of the selected metrics from an auditory oddball paradigm may be used as a benchmark method for testing the performance of different EEG systems in mobile conditions. Moreover dry EEG systems may need substantial improvements to meet the quality standards of wet electrodes.",
keywords = "Journal Article",
author = "Oliveira, {Anderson S} and Schlink, {Bryan R} and Hairston, {W David} and Peter K{\"o}nig and Ferris, {Daniel P}",
year = "2016",
doi = "10.3389/fnhum.2016.00188",
language = "English",
volume = "10",
pages = "188",
journal = "FRONT HUM NEUROSCI",
issn = "1662-5161",
publisher = "Frontiers Research Foundation",
number = "10",

}

RIS

TY - JOUR

T1 - Proposing Metrics for Benchmarking Novel EEG Technologies Towards Real-World Measurements

AU - Oliveira, Anderson S

AU - Schlink, Bryan R

AU - Hairston, W David

AU - König, Peter

AU - Ferris, Daniel P

PY - 2016

Y1 - 2016

N2 - Recent advances in electroencephalographic (EEG) acquisition allow for recordings using wet and dry sensors during whole-body motion. The large variety of commercially available EEG systems contrasts with the lack of established methods for objectively describing their performance during whole-body motion. Therefore, the aim of this study was to introduce methods for benchmarking the suitability of new EEG technologies for that context. Subjects performed an auditory oddball task using three different EEG systems (Biosemi wet-BSM, Cognionics Wet-Cwet, Conionics Dry-Cdry). Nine subjects performed the oddball task while seated and walking on a treadmill. We calculated EEG epoch rejection rate, pre-stimulus noise (PSN), signal-to-noise ratio (SNR) and EEG amplitude variance across the P300 event window (CVERP) from a subset of 12 channels common to all systems. We also calculated test-retest reliability and the subject's level of comfort while using each system. Our results showed that using the traditional 75 μV rejection threshold BSM and Cwet epoch rejection rates are ~25% and ~47% in the seated and walking conditions respectively. However, this threshold rejects ~63% of epochs for Cdry in the seated condition and excludes 100% of epochs for the majority of subjects during walking. BSM showed predominantly no statistical differences between seated and walking condition for all metrics, whereas Cwet showed increases in PSN and CVERP, as well as reduced SNR in the walking condition. Data quality from Cdry in seated conditions were predominantly inferior in comparison to the wet systems. Test-retest reliability was mostly moderate/good for these variables, especially in seated conditions. In addition, subjects felt less discomfort and were motivated for longer recording periods while using wet EEG systems in comparison to the dry system. The proposed method was successful in identifying differences across systems that are mostly caused by motion-related artifacts and usability issues. We conclude that the extraction of the selected metrics from an auditory oddball paradigm may be used as a benchmark method for testing the performance of different EEG systems in mobile conditions. Moreover dry EEG systems may need substantial improvements to meet the quality standards of wet electrodes.

AB - Recent advances in electroencephalographic (EEG) acquisition allow for recordings using wet and dry sensors during whole-body motion. The large variety of commercially available EEG systems contrasts with the lack of established methods for objectively describing their performance during whole-body motion. Therefore, the aim of this study was to introduce methods for benchmarking the suitability of new EEG technologies for that context. Subjects performed an auditory oddball task using three different EEG systems (Biosemi wet-BSM, Cognionics Wet-Cwet, Conionics Dry-Cdry). Nine subjects performed the oddball task while seated and walking on a treadmill. We calculated EEG epoch rejection rate, pre-stimulus noise (PSN), signal-to-noise ratio (SNR) and EEG amplitude variance across the P300 event window (CVERP) from a subset of 12 channels common to all systems. We also calculated test-retest reliability and the subject's level of comfort while using each system. Our results showed that using the traditional 75 μV rejection threshold BSM and Cwet epoch rejection rates are ~25% and ~47% in the seated and walking conditions respectively. However, this threshold rejects ~63% of epochs for Cdry in the seated condition and excludes 100% of epochs for the majority of subjects during walking. BSM showed predominantly no statistical differences between seated and walking condition for all metrics, whereas Cwet showed increases in PSN and CVERP, as well as reduced SNR in the walking condition. Data quality from Cdry in seated conditions were predominantly inferior in comparison to the wet systems. Test-retest reliability was mostly moderate/good for these variables, especially in seated conditions. In addition, subjects felt less discomfort and were motivated for longer recording periods while using wet EEG systems in comparison to the dry system. The proposed method was successful in identifying differences across systems that are mostly caused by motion-related artifacts and usability issues. We conclude that the extraction of the selected metrics from an auditory oddball paradigm may be used as a benchmark method for testing the performance of different EEG systems in mobile conditions. Moreover dry EEG systems may need substantial improvements to meet the quality standards of wet electrodes.

KW - Journal Article

U2 - 10.3389/fnhum.2016.00188

DO - 10.3389/fnhum.2016.00188

M3 - SCORING: Journal article

C2 - 27242467

VL - 10

SP - 188

JO - FRONT HUM NEUROSCI

JF - FRONT HUM NEUROSCI

SN - 1662-5161

IS - 10

ER -