Multicenter Evaluation of a 0-Hour/1-Hour Algorithm in the Diagnosis of Myocardial Infarction With High-Sensitivity Cardiac Troponin T

  • Christian Mueller
  • Evangelos Giannitsis
  • Michael Christ
  • Jorge Ordóñez-Llanos
  • Christopher deFilippi
  • James McCord
  • Richard Body
  • Mauro Panteghini
  • Tomas Jernberg
  • Mario Plebani
  • Franck Verschuren
  • John French
  • Robert Christenson
  • Silvia Weiser
  • Garnet Bendig
  • Peter Dilba
  • Bertil Lindahl
  • TRAPID-AMI Investigators
  • Raphael Twerenbold

Abstract

STUDY OBJECTIVE: We aim to prospectively validate the diagnostic accuracy of the recently developed 0-h/1-h algorithm, using high-sensitivity cardiac troponin T (hs-cTnT) for the early rule-out and rule-in of acute myocardial infarction.

METHODS: We enrolled patients presenting with suspected acute myocardial infarction and recent (<6 hours) onset of symptoms to the emergency department in a global multicenter diagnostic study. Hs-cTnT (Roche Diagnostics) and sensitive cardiac troponin I (Siemens Healthcare) were measured at presentation and after 1 hour, 2 hours, and 4 to 14 hours in a central laboratory. Patient triage according to the predefined hs-cTnT 0-hour/1-hour algorithm (hs-cTnT below 12 ng/L and Δ1 hour below 3 ng/L to rule out; hs-cTnT at least 52 ng/L or Δ1 hour at least 5 ng/L to rule in; remaining patients to the "observational zone") was compared against a centrally adjudicated final diagnosis by 2 independent cardiologists (reference standard). The final diagnosis was based on all available information, including coronary angiography and echocardiography results, follow-up data, and serial measurements of sensitive cardiac troponin I, whereas adjudicators remained blinded to hs-cTnT.

RESULTS: Among 1,282 patients enrolled, acute myocardial infarction was the final diagnosis for 213 (16.6%) patients. Applying the hs-cTnT 0-hour/1-hour algorithm, 813 (63.4%) patients were classified as rule out, 184 (14.4%) were classified as rule in, and 285 (22.2%) were triaged to the observational zone. This resulted in a negative predictive value and sensitivity for acute myocardial infarction of 99.1% (95% confidence interval [CI] 98.2% to 99.7%) and 96.7% (95% CI 93.4% to 98.7%) in the rule-out zone (7 patients with false-negative results), a positive predictive value and specificity for acute myocardial infarction of 77.2% (95% CI 70.4% to 83.0%) and 96.1% (95% CI 94.7% to 97.2%) in the rule-in zone, and a prevalence of acute myocardial infarction of 22.5% in the observational zone.

CONCLUSION: The hs-cTnT 0-hour/1-hour algorithm performs well for early rule-out and rule-in of acute myocardial infarction.

Bibliografische Daten

OriginalspracheEnglisch
ISSN0196-0644
DOIs
StatusVeröffentlicht - 07.2016
Extern publiziertJa

Anmerkungen des Dekanats

Copyright © 2016 American College of Emergency Physicians. Published by Elsevier Inc. All rights reserved.

PubMed 26794254