Long-term experience with AutoCapture-controlled epicardial pacing in children

  • Maren Tomaske
  • Paul Harpes
  • Rene Pretre
  • Ali Dodge-Khatami
  • Urs Bauersfeld

Abstract

AIMS: To examine the feasibility and safety of AutoCapture (AC)-controlled pacing with epicardial leads in children, and study the effects on device longevity.

METHODS: A total of 62 children were prospectively enrolled. Pre-discharge testing precluded AC function in six children. In 56 (90%) children, devices with AC-controlled pacing were followed up to 9years. Calculated battery life in AC-controlled pacing was compared with theoretical calculations, using a two-fold stimulation output of measured thresholds.

RESULTS: In 53 of 56 children, no differences were observed for evoked response signals (13.3 vs. 11.5mV, P = 0.20) or lead polarization safety margins (5.5 vs. 4.1, P = 0.25) at 6-month and 4-year follow-up. A crossover to conventional pacing was required in 3 of 56 children. AC-controlled pacing prolonged the calculated battery life up to 15% for the identity and integrity devices with 0.95A h capacity, compared with theoretical conventional settings (P = 0.008). In patients with ventricular pacing thresholds >1.5V at 0.5ms, battery life was increased by 30% compared with theoretical conventional settings (P < 0.001).

CONCLUSION: AC-controlled pacing with epicardial leads is feasible and safe in children during long-term follow-up. An adequate lead polarization safety margin persists in most patients. Calculated battery life was prolonged up to 15% with AC-controlled pacing. Patients with high or fluctuating pacing thresholds benefit the most.

Bibliografische Daten

OriginalspracheEnglisch
ISSN1099-5129
DOIs
StatusVeröffentlicht - 08.2007
PubMed 17630391