Laser Doppler flowmetry evaluation of gingival recovery response after laser treatment

  • Carmen Todea
  • Silvana Canjau
  • Dorin Dodenciu
  • Mariana I. Miron
  • Anca Tudor
  • Cosmin Balabuc

Abstract

This study was performed in order to evaluate in vivo the applicability of Laser Doppler Flowmetry (LDF) in recording the gingival blood flow and to assess the changes of gingival blood flow following gingival reshaping performed with Er:YAG and 980 nm diode lasers. The LDF evaluation was performed on 20 anterior teeth, which underwent reshaping of gingiva, corresponding to 5 female patients (4 anterior teeth/patient), aged between 20 and 35. One part of the mouth was treated with Er:YAG laser (LP, VLP modes, 140 - 250 mJ, 10 - 20 Hz, using cylindrical sapphire tips) and other part with 980 nm diode laser (CW, 4 W, contact mode and saline solution cooling). The gingival blood flow was monitored using a MoorLab laser Doppler equipment (Moor Instruments Ltd., Axminster, UK) with a straight optical probe, MP3b, 10 mm. The data were processed using statistical analysis software SPSS v16.0.1. The investigation showed an evident decrease in perfusion for both areas in comparison with the baseline values 24 hours after treatment. The microvascular blood flow increased significantly after 7 days in both areas but mostly in diode area (p

Bibliografische Daten

OriginalspracheEnglisch
TitelHEAD AND NECK OPTICAL DIAGNOSTICS
Redakteure/-innenC Betz, BJF Wong
Band8805
Erscheinungsort1000 20TH ST, PO BOX 10, BELLINGHAM, WA 98227-0010 USA
Herausgeber (Verlag)SPIE-INT SOC OPTICAL ENGINEERING
Erscheinungsdatum2013
ISBN (Print)978-0-8194-9813-7
DOIs
StatusVeröffentlicht - 2013
Extern publiziertJa

Anmerkungen des Dekanats

5th Scientific Meeting of the Head-and-Neck-Optical-Diagnostic-Society (HNODS), Munich, GERMANY, MAY 13-14, 2013