Intraoperative Blood Pressure Monitoring in Obese Patients: Arterial Catheter, Finger Cuff, and Oscillometry

Standard

Intraoperative Blood Pressure Monitoring in Obese Patients: Arterial Catheter, Finger Cuff, and Oscillometry. / Schumann, Roman; Meidert, Agnes S; Bonney, Iwona; Koutentis, Christos; Wesselink, Wilbert; Kouz, Karim; Saugel, Bernd.

in: ANESTHESIOLOGY, Jahrgang 134, Nr. 2, 01.02.2021, S. 179-188.

Publikationen: SCORING: Beitrag in Fachzeitschrift/ZeitungSCORING: ZeitschriftenaufsatzForschungBegutachtung

Harvard

APA

Vancouver

Bibtex

@article{4895220a0b4048f2b9641a46b3204eba,
title = "Intraoperative Blood Pressure Monitoring in Obese Patients: Arterial Catheter, Finger Cuff, and Oscillometry",
abstract = "BACKGROUND: The optimal method for blood pressure monitoring in obese surgical patients remains unknown. Arterial catheters can cause potential complications, and noninvasive oscillometry provides only intermittent values. Finger cuff methods allow continuous noninvasive monitoring. The authors tested the hypothesis that the agreement between finger cuff and intraarterial measurements is better than the agreement between oscillometric and intraarterial measurements.METHODS: This prospective study compared intraarterial (reference method), finger cuff, and oscillometric (upper arm, forearm, and lower leg) blood pressure measurements in 90 obese patients having bariatric surgery using Bland-Altman analysis, four-quadrant plot and concordance analysis (to assess the ability of monitoring methods to follow blood pressure changes), and error grid analysis (to describe the clinical relevance of measurement differences).RESULTS: The difference (mean ± SD) between finger cuff and intraarterial measurements was -1 mmHg (± 11 mmHg) for mean arterial pressure, -7 mmHg (± 14 mmHg) for systolic blood pressure, and 0 mmHg (± 11 mmHg) for diastolic blood pressure. Concordance between changes in finger cuff and intraarterial measurements was 88% (mean arterial pressure), 85% (systolic blood pressure), and 81% (diastolic blood pressure). In error grid analysis comparing finger cuff and intraarterial measurements, the proportions of measurements in risk zones A to E were 77.1%, 21.6%, 0.9%, 0.4%, and 0.0% for mean arterial pressure, respectively, and 89.5%, 9.8%, 0.2%, 0.4%, and 0.2%, respectively, for systolic blood pressure. For mean arterial pressure and diastolic blood pressure, absolute agreement and trending agreement between finger cuff and intraarterial measurements were better than between oscillometric (at each of the three measurement sites) and intraarterial measurements. Forearm performed better than upper arm and lower leg monitoring with regard to absolute agreement and trending agreement with intraarterial monitoring.CONCLUSIONS: The agreement between finger cuff and intraarterial measurements was better than the agreement between oscillometric and intraarterial measurements for mean arterial pressure and diastolic blood pressure in obese patients during surgery. Forearm oscillometry exhibits better measurement performance than upper arm or lower leg oscillometry.",
keywords = "Bariatric Surgery, Blood Pressure, Blood Pressure Determination/methods, Catheterization, Peripheral/methods, Female, Fingers, Humans, Male, Middle Aged, Monitoring, Intraoperative/methods, Obesity/surgery, Oscillometry/methods, Prospective Studies, Reproducibility of Results",
author = "Roman Schumann and Meidert, {Agnes S} and Iwona Bonney and Christos Koutentis and Wilbert Wesselink and Karim Kouz and Bernd Saugel",
note = "Copyright {\textcopyright} 2020, the American Society of Anesthesiologists, Inc. All Rights Reserved.",
year = "2021",
month = feb,
day = "1",
doi = "10.1097/ALN.0000000000003636",
language = "English",
volume = "134",
pages = "179--188",
journal = "ANESTHESIOLOGY",
issn = "0003-3022",
publisher = "Lippincott Williams and Wilkins",
number = "2",

}

RIS

TY - JOUR

T1 - Intraoperative Blood Pressure Monitoring in Obese Patients: Arterial Catheter, Finger Cuff, and Oscillometry

AU - Schumann, Roman

AU - Meidert, Agnes S

AU - Bonney, Iwona

AU - Koutentis, Christos

AU - Wesselink, Wilbert

AU - Kouz, Karim

AU - Saugel, Bernd

N1 - Copyright © 2020, the American Society of Anesthesiologists, Inc. All Rights Reserved.

PY - 2021/2/1

Y1 - 2021/2/1

N2 - BACKGROUND: The optimal method for blood pressure monitoring in obese surgical patients remains unknown. Arterial catheters can cause potential complications, and noninvasive oscillometry provides only intermittent values. Finger cuff methods allow continuous noninvasive monitoring. The authors tested the hypothesis that the agreement between finger cuff and intraarterial measurements is better than the agreement between oscillometric and intraarterial measurements.METHODS: This prospective study compared intraarterial (reference method), finger cuff, and oscillometric (upper arm, forearm, and lower leg) blood pressure measurements in 90 obese patients having bariatric surgery using Bland-Altman analysis, four-quadrant plot and concordance analysis (to assess the ability of monitoring methods to follow blood pressure changes), and error grid analysis (to describe the clinical relevance of measurement differences).RESULTS: The difference (mean ± SD) between finger cuff and intraarterial measurements was -1 mmHg (± 11 mmHg) for mean arterial pressure, -7 mmHg (± 14 mmHg) for systolic blood pressure, and 0 mmHg (± 11 mmHg) for diastolic blood pressure. Concordance between changes in finger cuff and intraarterial measurements was 88% (mean arterial pressure), 85% (systolic blood pressure), and 81% (diastolic blood pressure). In error grid analysis comparing finger cuff and intraarterial measurements, the proportions of measurements in risk zones A to E were 77.1%, 21.6%, 0.9%, 0.4%, and 0.0% for mean arterial pressure, respectively, and 89.5%, 9.8%, 0.2%, 0.4%, and 0.2%, respectively, for systolic blood pressure. For mean arterial pressure and diastolic blood pressure, absolute agreement and trending agreement between finger cuff and intraarterial measurements were better than between oscillometric (at each of the three measurement sites) and intraarterial measurements. Forearm performed better than upper arm and lower leg monitoring with regard to absolute agreement and trending agreement with intraarterial monitoring.CONCLUSIONS: The agreement between finger cuff and intraarterial measurements was better than the agreement between oscillometric and intraarterial measurements for mean arterial pressure and diastolic blood pressure in obese patients during surgery. Forearm oscillometry exhibits better measurement performance than upper arm or lower leg oscillometry.

AB - BACKGROUND: The optimal method for blood pressure monitoring in obese surgical patients remains unknown. Arterial catheters can cause potential complications, and noninvasive oscillometry provides only intermittent values. Finger cuff methods allow continuous noninvasive monitoring. The authors tested the hypothesis that the agreement between finger cuff and intraarterial measurements is better than the agreement between oscillometric and intraarterial measurements.METHODS: This prospective study compared intraarterial (reference method), finger cuff, and oscillometric (upper arm, forearm, and lower leg) blood pressure measurements in 90 obese patients having bariatric surgery using Bland-Altman analysis, four-quadrant plot and concordance analysis (to assess the ability of monitoring methods to follow blood pressure changes), and error grid analysis (to describe the clinical relevance of measurement differences).RESULTS: The difference (mean ± SD) between finger cuff and intraarterial measurements was -1 mmHg (± 11 mmHg) for mean arterial pressure, -7 mmHg (± 14 mmHg) for systolic blood pressure, and 0 mmHg (± 11 mmHg) for diastolic blood pressure. Concordance between changes in finger cuff and intraarterial measurements was 88% (mean arterial pressure), 85% (systolic blood pressure), and 81% (diastolic blood pressure). In error grid analysis comparing finger cuff and intraarterial measurements, the proportions of measurements in risk zones A to E were 77.1%, 21.6%, 0.9%, 0.4%, and 0.0% for mean arterial pressure, respectively, and 89.5%, 9.8%, 0.2%, 0.4%, and 0.2%, respectively, for systolic blood pressure. For mean arterial pressure and diastolic blood pressure, absolute agreement and trending agreement between finger cuff and intraarterial measurements were better than between oscillometric (at each of the three measurement sites) and intraarterial measurements. Forearm performed better than upper arm and lower leg monitoring with regard to absolute agreement and trending agreement with intraarterial monitoring.CONCLUSIONS: The agreement between finger cuff and intraarterial measurements was better than the agreement between oscillometric and intraarterial measurements for mean arterial pressure and diastolic blood pressure in obese patients during surgery. Forearm oscillometry exhibits better measurement performance than upper arm or lower leg oscillometry.

KW - Bariatric Surgery

KW - Blood Pressure

KW - Blood Pressure Determination/methods

KW - Catheterization, Peripheral/methods

KW - Female

KW - Fingers

KW - Humans

KW - Male

KW - Middle Aged

KW - Monitoring, Intraoperative/methods

KW - Obesity/surgery

KW - Oscillometry/methods

KW - Prospective Studies

KW - Reproducibility of Results

U2 - 10.1097/ALN.0000000000003636

DO - 10.1097/ALN.0000000000003636

M3 - SCORING: Journal article

C2 - 33326001

VL - 134

SP - 179

EP - 188

JO - ANESTHESIOLOGY

JF - ANESTHESIOLOGY

SN - 0003-3022

IS - 2

ER -