Intra-aneurysmal flow disruption after implantation of the Medina® Embolization Device depends on aneurysm neck coverage

Standard

Intra-aneurysmal flow disruption after implantation of the Medina® Embolization Device depends on aneurysm neck coverage. / Frölich, Andreas Maximilian; Nawka, Marie Teresa; Ernst, Marielle; Frischmuth, Isabell; Fiehler, Jens; Buhk, Jan-Hendrik.

in: PLOS ONE, Jahrgang 13, Nr. 2, 2018, S. e0191975.

Publikationen: SCORING: Beitrag in Fachzeitschrift/ZeitungSCORING: ZeitschriftenaufsatzForschungBegutachtung

Harvard

APA

Vancouver

Bibtex

@article{7c35a4dca3494a1491ced734864f2a23,
title = "Intra-aneurysmal flow disruption after implantation of the Medina{\textregistered} Embolization Device depends on aneurysm neck coverage",
abstract = "BACKGROUND AND PURPOSE: Flow disruption achieved by braided intrasaccular implants is a novel treatment strategy for cerebrovascular aneurysms. We hypothesized that the degree of intra-aneurysmal flow disruption can be quantified in vitro and is influenced by device position across the aneurysm neck. We tested this hypothesis using the Medina{\textregistered} Embolization Device (MED).METHODS: Ten different patient-specific elastic vascular models were manufactured. Models were connected to a pulsatile flow circuit, filled with a blood-mimicking fluid and treated by two operators using a single MED. Intra-aneurysmal flow velocity was measured using conventional and high-frequency digital subtraction angiography (HF-DSA) before and after each deployment. Aneurysm neck coverage by the implanted devices was assessed with flat detector computed tomography on a three-point Likert scale.RESULTS: A total of 80 individual MED deployments were performed by the two operators. The mean intra-aneurysmal flow velocity reduction after MED implantation was 33.6% (27.5-39.7%). No significant differences in neck coverage (p = 0.99) or flow disruption (p = 0.84) were observed between operators. The degree of flow disruption significantly correlated with neck coverage (ρ = 0.42, 95% CI: 0.21-0.59, p = 0.002) as well as with neck area (ρ = -0,35, 95% CI: -0.54 --0.13, p = 0.024). On multiple regression analysis, both neck coverage and total neck area were independent predictors of flow disruption.CONCLUSIONS: The degree of intra-aneurysmal flow disruption after MED implantation can be quantified in vitro and varies considerably between different aneurysms and different device configurations. Optimal device coverage across the aneurysm neck improves flow disruption and may thus contribute to aneurysm occlusion.",
keywords = "Journal Article",
author = "Fr{\"o}lich, {Andreas Maximilian} and Nawka, {Marie Teresa} and Marielle Ernst and Isabell Frischmuth and Jens Fiehler and Jan-Hendrik Buhk",
year = "2018",
doi = "10.1371/journal.pone.0191975",
language = "English",
volume = "13",
pages = "e0191975",
journal = "PLOS ONE",
issn = "1932-6203",
publisher = "Public Library of Science",
number = "2",

}

RIS

TY - JOUR

T1 - Intra-aneurysmal flow disruption after implantation of the Medina® Embolization Device depends on aneurysm neck coverage

AU - Frölich, Andreas Maximilian

AU - Nawka, Marie Teresa

AU - Ernst, Marielle

AU - Frischmuth, Isabell

AU - Fiehler, Jens

AU - Buhk, Jan-Hendrik

PY - 2018

Y1 - 2018

N2 - BACKGROUND AND PURPOSE: Flow disruption achieved by braided intrasaccular implants is a novel treatment strategy for cerebrovascular aneurysms. We hypothesized that the degree of intra-aneurysmal flow disruption can be quantified in vitro and is influenced by device position across the aneurysm neck. We tested this hypothesis using the Medina® Embolization Device (MED).METHODS: Ten different patient-specific elastic vascular models were manufactured. Models were connected to a pulsatile flow circuit, filled with a blood-mimicking fluid and treated by two operators using a single MED. Intra-aneurysmal flow velocity was measured using conventional and high-frequency digital subtraction angiography (HF-DSA) before and after each deployment. Aneurysm neck coverage by the implanted devices was assessed with flat detector computed tomography on a three-point Likert scale.RESULTS: A total of 80 individual MED deployments were performed by the two operators. The mean intra-aneurysmal flow velocity reduction after MED implantation was 33.6% (27.5-39.7%). No significant differences in neck coverage (p = 0.99) or flow disruption (p = 0.84) were observed between operators. The degree of flow disruption significantly correlated with neck coverage (ρ = 0.42, 95% CI: 0.21-0.59, p = 0.002) as well as with neck area (ρ = -0,35, 95% CI: -0.54 --0.13, p = 0.024). On multiple regression analysis, both neck coverage and total neck area were independent predictors of flow disruption.CONCLUSIONS: The degree of intra-aneurysmal flow disruption after MED implantation can be quantified in vitro and varies considerably between different aneurysms and different device configurations. Optimal device coverage across the aneurysm neck improves flow disruption and may thus contribute to aneurysm occlusion.

AB - BACKGROUND AND PURPOSE: Flow disruption achieved by braided intrasaccular implants is a novel treatment strategy for cerebrovascular aneurysms. We hypothesized that the degree of intra-aneurysmal flow disruption can be quantified in vitro and is influenced by device position across the aneurysm neck. We tested this hypothesis using the Medina® Embolization Device (MED).METHODS: Ten different patient-specific elastic vascular models were manufactured. Models were connected to a pulsatile flow circuit, filled with a blood-mimicking fluid and treated by two operators using a single MED. Intra-aneurysmal flow velocity was measured using conventional and high-frequency digital subtraction angiography (HF-DSA) before and after each deployment. Aneurysm neck coverage by the implanted devices was assessed with flat detector computed tomography on a three-point Likert scale.RESULTS: A total of 80 individual MED deployments were performed by the two operators. The mean intra-aneurysmal flow velocity reduction after MED implantation was 33.6% (27.5-39.7%). No significant differences in neck coverage (p = 0.99) or flow disruption (p = 0.84) were observed between operators. The degree of flow disruption significantly correlated with neck coverage (ρ = 0.42, 95% CI: 0.21-0.59, p = 0.002) as well as with neck area (ρ = -0,35, 95% CI: -0.54 --0.13, p = 0.024). On multiple regression analysis, both neck coverage and total neck area were independent predictors of flow disruption.CONCLUSIONS: The degree of intra-aneurysmal flow disruption after MED implantation can be quantified in vitro and varies considerably between different aneurysms and different device configurations. Optimal device coverage across the aneurysm neck improves flow disruption and may thus contribute to aneurysm occlusion.

KW - Journal Article

U2 - 10.1371/journal.pone.0191975

DO - 10.1371/journal.pone.0191975

M3 - SCORING: Journal article

C2 - 29408857

VL - 13

SP - e0191975

JO - PLOS ONE

JF - PLOS ONE

SN - 1932-6203

IS - 2

ER -