Inter-rater reliability and double reading analysis of an automated three-dimensional breast ultrasound system: comparison of two independent examiners

  • Anna Maier
  • Joerg Heil
  • Anna Lauer
  • Aba Harcos
  • Benedikt Schaefgen
  • Alexandra von Au
  • Julia Spratte
  • Fabian Riedel
  • Geraldine Rauch
  • André Hennigs
  • Christoph Domschke
  • Sarah Schott
  • Joachim Rom
  • Florian Schuetz
  • Christof Sohn
  • Michael Golatta

Abstract

PURPOSE: Breast ultrasound could be a valuable tool complementary to mammography in breast cancer screening. Automated 3D breast ultrasound (ABUS) addresses challenges of hand-held ultrasound and could allow double reading analysis of ultrasound images. This trial assesses the inter-rater reliability and double reading analysis of an ABUS system. METHODS: To assess the reproducibility and diagnostic validity of the ABUS system, SomoV™, a blinded double reading analysis, was performed in 1019 patients (2038 breasts) by two examiners (examiner A/B) and compared to single reading results, as well as to the reference standard regarding its diagnostic validity. Cohen's kappa coefficients were calculated to measure the inter-rater reliability and agreement of the different diagnostic modalities. Patient comfort and time consumption for image acquisition and reading were analyzed descriptively as secondary objectives. RESULTS: Analysis of inter-rater reliability yielded agreement in 81.6% (κ = 0.37; p < 0.0001) showing fair agreement. Single reading analysis of SomoV™ exams (examiner A/examiner B) compared to reference standard showed good specificity (examiner A: 88.3%/examiner B: 84.5%), fair inter-rater agreement (examiner A: κ = 0.31/examiner B: κ = 0.31), and adequate sensitivity (examiner A: 53.1%/examiner B: 64.2%). Double reading analysis yielded good sensitivity and specificity (73.7 and 77.7%). Mammography (n = 1911) alone detected 160 of 176 carcinomas (sensitivity 90.1%). Adding SomoV™ to mammography would have detected 12 additional carcinomas, resulting in a higher sensitivity of 97.7%. CONCLUSION: SomoV™ is a promising technique with good sensitivity, high patient comfort, and fair inter-examiner reliability. It allows double reading analysis that, in combination with mammography, could increase detection rates in breast cancer screening.

Bibliografische Daten

OriginalspracheEnglisch
ISSN0932-0067
DOIs
StatusVeröffentlicht - 09.2017
PubMed 28748340