Correct placement of iliosacral screws remains a surgical challenge. The aim of this retrospective study was to identify parameters which impact the accuracy of this technically demanding procedure. Eighty-two patients with vertically unstable pelvic injuries treated with a total of 147 iliosacral screws were included. Assessment of postoperative CT scans revealed screw misplacement in 13 cases (8%), of which six occurred following insertion of two unilateral screws into S1. Six screw misplacements occurred in patients with dislocation injuries of the posterior pelvis. Comparison of a navigated and the standard technique revealed a decreased screw misplacement rate in the navigated group (15% standard vs. 3% navigation, p?