Evaluation of a new continuous thermodilution cardiac output monitor in cardiac surgical patients: a prospective criterion standard study.
Standard
Evaluation of a new continuous thermodilution cardiac output monitor in cardiac surgical patients: a prospective criterion standard study. / Zöllner, Christian; Polasek, J; Kilger, E; Pichler, B; Jaenicke, U; Briegel, J; Vetter, H O; Haller, M.
in: CRIT CARE MED, Jahrgang 27, Nr. 2, 2, 1999, S. 293-298.Publikationen: SCORING: Beitrag in Fachzeitschrift/Zeitung › SCORING: Zeitschriftenaufsatz › Forschung › Begutachtung
Harvard
APA
Vancouver
Bibtex
}
RIS
TY - JOUR
T1 - Evaluation of a new continuous thermodilution cardiac output monitor in cardiac surgical patients: a prospective criterion standard study.
AU - Zöllner, Christian
AU - Polasek, J
AU - Kilger, E
AU - Pichler, B
AU - Jaenicke, U
AU - Briegel, J
AU - Vetter, H O
AU - Haller, M
PY - 1999
Y1 - 1999
N2 - OBJECTIVE: To evaluate the accuracy of a new continuous cardiac output monitor in critically ill patients. DESIGN: Criterion standard study. SETTING: Cardiac surgery intensive care unit in a university hospital. PATIENTS: Twenty cardiac surgical patients requiring intensive care treatment with pulmonary artery catheters after surgery. INTERVENTIONS: None. MEASUREMENTS AND MAIN RESULTS: Cardiac output was monitored continuously with a modified pulmonary artery catheter that has a heating filament on the outside of the catheter. Four modes of data processing with different response times ("Normal," "Fast," "FastFilter," and "Urgent" modes) used by the monitoring system. A total of 240 determinations of cardiac output were performed using conventional bolus thermodilution technique; these results were compared with those obtained using three of the four continuous measuring modes available ("Normal," "FastFilter," and "Urgent"). Cardiac output ranged from 3.47 to 15.77 L/min (bolus thermodilution). The mean (bias) +/- SD of differences (precision) for all measurements was 0.40+/-1.26 L/min in the Normal mode (cardiac output 10 L/min did not agree with the results of the bolus thermodilution method.
AB - OBJECTIVE: To evaluate the accuracy of a new continuous cardiac output monitor in critically ill patients. DESIGN: Criterion standard study. SETTING: Cardiac surgery intensive care unit in a university hospital. PATIENTS: Twenty cardiac surgical patients requiring intensive care treatment with pulmonary artery catheters after surgery. INTERVENTIONS: None. MEASUREMENTS AND MAIN RESULTS: Cardiac output was monitored continuously with a modified pulmonary artery catheter that has a heating filament on the outside of the catheter. Four modes of data processing with different response times ("Normal," "Fast," "FastFilter," and "Urgent" modes) used by the monitoring system. A total of 240 determinations of cardiac output were performed using conventional bolus thermodilution technique; these results were compared with those obtained using three of the four continuous measuring modes available ("Normal," "FastFilter," and "Urgent"). Cardiac output ranged from 3.47 to 15.77 L/min (bolus thermodilution). The mean (bias) +/- SD of differences (precision) for all measurements was 0.40+/-1.26 L/min in the Normal mode (cardiac output 10 L/min did not agree with the results of the bolus thermodilution method.