Electrical cardioversion of patients with implanted pacemaker or cardioverter-defibrillator: results of a survey of german centers and systematic review of the literature

Standard

Electrical cardioversion of patients with implanted pacemaker or cardioverter-defibrillator: results of a survey of german centers and systematic review of the literature. / Lüker, J; Sultan, A; Plenge, T; van den Bruck, J; Heeger, C-H; Meyer, S; Mischke, K; Tilz, R R; Vollmann, D; Nölker, G; Schäffer, B; Willems, S; Steven, D.

in: CLIN RES CARDIOL, Jahrgang 107, Nr. 3, 03.2018, S. 249-258.

Publikationen: SCORING: Beitrag in Fachzeitschrift/ZeitungSCORING: ReviewForschung

Harvard

Lüker, J, Sultan, A, Plenge, T, van den Bruck, J, Heeger, C-H, Meyer, S, Mischke, K, Tilz, RR, Vollmann, D, Nölker, G, Schäffer, B, Willems, S & Steven, D 2018, 'Electrical cardioversion of patients with implanted pacemaker or cardioverter-defibrillator: results of a survey of german centers and systematic review of the literature', CLIN RES CARDIOL, Jg. 107, Nr. 3, S. 249-258. https://doi.org/10.1007/s00392-017-1178-y

APA

Lüker, J., Sultan, A., Plenge, T., van den Bruck, J., Heeger, C-H., Meyer, S., Mischke, K., Tilz, R. R., Vollmann, D., Nölker, G., Schäffer, B., Willems, S., & Steven, D. (2018). Electrical cardioversion of patients with implanted pacemaker or cardioverter-defibrillator: results of a survey of german centers and systematic review of the literature. CLIN RES CARDIOL, 107(3), 249-258. https://doi.org/10.1007/s00392-017-1178-y

Vancouver

Bibtex

@article{ef235e214d794d4cbc2a8d07fe1c3b51,
title = "Electrical cardioversion of patients with implanted pacemaker or cardioverter-defibrillator: results of a survey of german centers and systematic review of the literature",
abstract = "AIMS: A relevant number of patients presenting for electrical cardioversion carry a pacemaker (PM) or ICD. Case reports suggest a potential hazard of external cardioversion/defibrillation. The incidence of shock related device complications is unknown. No guidelines or recommendations by international medical societies for a cardioversion protocol of cardiovascular implantable electronic device (CIED) patients exist. We conducted a nationwide survey to gather real-world clinical data on the current clinical approach towards these patients during electrical cardioversion and to estimate the incidence of shock-related complications.METHODS AND RESULTS: Ninety hospitals with > 380 ECV in 2014 were identified from mandatory hospital quality reports and 60 were randomly selected. All centers were provided with a standardized questionnaire on the general proceedings and complications during electrical cardioversion of pacemaker, ICD and CRT patients (CIED patients). Thirty-two centers (53%) participated in the survey. In total, 16,554 ECV were reported (534 ± 314 per center). Biphasic cardioversion with a first shock energy of ≥ 150 J via adhesive patches in antero-posterior orientation was preferred by most centers (78%). Eleven percent (n = 1809) of pts were reported to carry a PM/ICD. The ECV protocol was heterogeneous among centers. Complications associated with electrical cardioversion were reported in 11/1809 patients (0.6%), all were transitory elevations of pacing thresholds.CONCLUSIONS: In this nationwide snapshot survey of cardioversion procedures in Germany, approximately 11% of patients presenting for elective electrical cardioversion were pacemaker or ICD carriers. Cardioversion protocols in these patients are heterogeneous throughout centers and mostly not in accordance with recommendation of the German Cardiac Society. Complications associated with external electrical cardioversion are rare. Controlled trials and large registries are necessary to provide evidence for future recommendations.",
keywords = "Arrhythmias, Cardiac/epidemiology, Defibrillators, Implantable, Electric Countershock/methods, Electrocardiography, Germany, Humans, Morbidity/trends, Pacemaker, Artificial, Registries, Societies, Medical, Survival Rate/trends",
author = "J L{\"u}ker and A Sultan and T Plenge and {van den Bruck}, J and C-H Heeger and S Meyer and K Mischke and Tilz, {R R} and D Vollmann and G N{\"o}lker and B Sch{\"a}ffer and S Willems and D Steven",
year = "2018",
month = mar,
doi = "10.1007/s00392-017-1178-y",
language = "English",
volume = "107",
pages = "249--258",
journal = "CLIN RES CARDIOL",
issn = "1861-0684",
publisher = "D. Steinkopff-Verlag",
number = "3",

}

RIS

TY - JOUR

T1 - Electrical cardioversion of patients with implanted pacemaker or cardioverter-defibrillator: results of a survey of german centers and systematic review of the literature

AU - Lüker, J

AU - Sultan, A

AU - Plenge, T

AU - van den Bruck, J

AU - Heeger, C-H

AU - Meyer, S

AU - Mischke, K

AU - Tilz, R R

AU - Vollmann, D

AU - Nölker, G

AU - Schäffer, B

AU - Willems, S

AU - Steven, D

PY - 2018/3

Y1 - 2018/3

N2 - AIMS: A relevant number of patients presenting for electrical cardioversion carry a pacemaker (PM) or ICD. Case reports suggest a potential hazard of external cardioversion/defibrillation. The incidence of shock related device complications is unknown. No guidelines or recommendations by international medical societies for a cardioversion protocol of cardiovascular implantable electronic device (CIED) patients exist. We conducted a nationwide survey to gather real-world clinical data on the current clinical approach towards these patients during electrical cardioversion and to estimate the incidence of shock-related complications.METHODS AND RESULTS: Ninety hospitals with > 380 ECV in 2014 were identified from mandatory hospital quality reports and 60 were randomly selected. All centers were provided with a standardized questionnaire on the general proceedings and complications during electrical cardioversion of pacemaker, ICD and CRT patients (CIED patients). Thirty-two centers (53%) participated in the survey. In total, 16,554 ECV were reported (534 ± 314 per center). Biphasic cardioversion with a first shock energy of ≥ 150 J via adhesive patches in antero-posterior orientation was preferred by most centers (78%). Eleven percent (n = 1809) of pts were reported to carry a PM/ICD. The ECV protocol was heterogeneous among centers. Complications associated with electrical cardioversion were reported in 11/1809 patients (0.6%), all were transitory elevations of pacing thresholds.CONCLUSIONS: In this nationwide snapshot survey of cardioversion procedures in Germany, approximately 11% of patients presenting for elective electrical cardioversion were pacemaker or ICD carriers. Cardioversion protocols in these patients are heterogeneous throughout centers and mostly not in accordance with recommendation of the German Cardiac Society. Complications associated with external electrical cardioversion are rare. Controlled trials and large registries are necessary to provide evidence for future recommendations.

AB - AIMS: A relevant number of patients presenting for electrical cardioversion carry a pacemaker (PM) or ICD. Case reports suggest a potential hazard of external cardioversion/defibrillation. The incidence of shock related device complications is unknown. No guidelines or recommendations by international medical societies for a cardioversion protocol of cardiovascular implantable electronic device (CIED) patients exist. We conducted a nationwide survey to gather real-world clinical data on the current clinical approach towards these patients during electrical cardioversion and to estimate the incidence of shock-related complications.METHODS AND RESULTS: Ninety hospitals with > 380 ECV in 2014 were identified from mandatory hospital quality reports and 60 were randomly selected. All centers were provided with a standardized questionnaire on the general proceedings and complications during electrical cardioversion of pacemaker, ICD and CRT patients (CIED patients). Thirty-two centers (53%) participated in the survey. In total, 16,554 ECV were reported (534 ± 314 per center). Biphasic cardioversion with a first shock energy of ≥ 150 J via adhesive patches in antero-posterior orientation was preferred by most centers (78%). Eleven percent (n = 1809) of pts were reported to carry a PM/ICD. The ECV protocol was heterogeneous among centers. Complications associated with electrical cardioversion were reported in 11/1809 patients (0.6%), all were transitory elevations of pacing thresholds.CONCLUSIONS: In this nationwide snapshot survey of cardioversion procedures in Germany, approximately 11% of patients presenting for elective electrical cardioversion were pacemaker or ICD carriers. Cardioversion protocols in these patients are heterogeneous throughout centers and mostly not in accordance with recommendation of the German Cardiac Society. Complications associated with external electrical cardioversion are rare. Controlled trials and large registries are necessary to provide evidence for future recommendations.

KW - Arrhythmias, Cardiac/epidemiology

KW - Defibrillators, Implantable

KW - Electric Countershock/methods

KW - Electrocardiography

KW - Germany

KW - Humans

KW - Morbidity/trends

KW - Pacemaker, Artificial

KW - Registries

KW - Societies, Medical

KW - Survival Rate/trends

U2 - 10.1007/s00392-017-1178-y

DO - 10.1007/s00392-017-1178-y

M3 - SCORING: Review article

C2 - 29151182

VL - 107

SP - 249

EP - 258

JO - CLIN RES CARDIOL

JF - CLIN RES CARDIOL

SN - 1861-0684

IS - 3

ER -