Dislocation of the polyethylene liner in hip arthroplasty: 6 patients with dislocation within 3 years.
Standard
Dislocation of the polyethylene liner in hip arthroplasty: 6 patients with dislocation within 3 years. / Niggemeyer, Oliver; Fink, Bernd; Rüther, Wolfgang.
in: ACTA ORTHOP SCAND, Jahrgang 73, Nr. 4, 4, 2002, S. 403-406.Publikationen: SCORING: Beitrag in Fachzeitschrift/Zeitung › SCORING: Zeitschriftenaufsatz › Forschung › Begutachtung
Harvard
APA
Vancouver
Bibtex
}
RIS
TY - JOUR
T1 - Dislocation of the polyethylene liner in hip arthroplasty: 6 patients with dislocation within 3 years.
AU - Niggemeyer, Oliver
AU - Fink, Bernd
AU - Rüther, Wolfgang
PY - 2002
Y1 - 2002
N2 - We revised 6 patients with early destruction of the liner in a cementless press-fit acetabular cup. They had no signs of infection or evidence of trauma. The failures were probably due to rotation of the liner inside the metal shell and destruction of the polyethylene by the cutting edge of the metal. This early complication may be difficult to diagnose because of normal radiographs. We suggest that the manufacturer should provide the insert with a metal marker to make this complication easier to detect.
AB - We revised 6 patients with early destruction of the liner in a cementless press-fit acetabular cup. They had no signs of infection or evidence of trauma. The failures were probably due to rotation of the liner inside the metal shell and destruction of the polyethylene by the cutting edge of the metal. This early complication may be difficult to diagnose because of normal radiographs. We suggest that the manufacturer should provide the insert with a metal marker to make this complication easier to detect.