Diagnostic value of ST-segment deviations during cardiac exercise stress testing: Systematic comparison of different ECG leads and time-points
Beteiligte Einrichtungen
Abstract
BACKGROUND: Exercise ECG stress testing is the most widely available method for evaluation of patients with suspected myocardial ischemia. Its major limitation is the relatively poor accuracy of ST-segment changes regarding ischemia detection. Little is known about the optimal method to assess ST-deviations.
METHODS: A total of 1558 consecutive patients undergoing bicycle exercise stress myocardial perfusion imaging (MPI) were enrolled. Presence of inducible myocardial ischemia was adjudicated using MPI results. The diagnostic value of ST-deviations for detection of exercise-induced myocardial ischemia was systematically analyzed 1) for each individual lead, 2) at three different intervals after the J-point (J+40ms, J+60ms, J+80ms), and 3) at different time points during the test (baseline, maximal workload, 2min into recovery).
RESULTS: Exercise-induced ischemia was detected in 481 (31%) patients. The diagnostic accuracy of ST-deviations was highest at +80ms after the J-point, and at 2min into recovery. At this point, ST-amplitude showed an AUC of 0.63 (95% CI 0.59-0.66) for the best-performing lead I. The combination of ST-amplitude and ST-slope in lead I did not increase the AUC. Lead I reached a sensitivity of 37% and a specificity of 83%, with similar sensitivity to manual ECG analysis (34%, p=0.31) but lower specificity (90%, p<0.001).
CONCLUSION: When using ECG stress testing for evaluation of patients with suspected myocardial ischemia, the diagnostic accuracy of ST-deviations is highest when evaluated at +80ms after the J-point, and at 2min into recovery.
Bibliografische Daten
Originalsprache | Englisch |
---|---|
ISSN | 0167-5273 |
DOIs | |
Status | Veröffentlicht - 01.07.2017 |
Anmerkungen des Dekanats
Copyright © 2017 Elsevier B.V. All rights reserved.
PubMed | 28320607 |
---|