Development of radiation exposure in patients undergoing pulmonary vein isolation in Germany between 2007 and 2014: great potential to minimize radiation dosage

  • Thomas Kleemann
  • Johannes Brachmann
  • Thorsten Lewalter
  • Dietrich Andresen
  • Stephan Willems
  • Stefan G Spitzer
  • Ellen Hoffmann
  • Lars Eckardt
  • Matthias Hochadel
  • Jochen Senges
  • Karl-Heinz Kuck
  • Karlheinz Seidl
  • Ralf Zahn

Abstract

AIM: The aim of this study was to analyze the radiation usage in patients undergoing pulmonary vein isolation (PVI) in Germany and to evaluate the possibility to reduce radiation dose.

METHODS AND RESULTS: A total of 6617 patients with atrial fibrillation (AF) from the German ablation registry and the FREEZEplus registry (control group), who underwent first PVI between 2007 and 2014, were analyzed. In the second step, the effect of optimized conventional fluoroscopy and optimized 3D mapping use was evaluated in 526 consecutive patients with AF who underwent first PVI at the Klinikum Ludwigshafen (optimized group) between 2007 and 2014. In the control group, the median dose area product (DAP) for PVI was 34 Gy cm(2), and the median DAP rate was 1.3 Gy cm(2)/min. The DAP decreased from 37 to 28 Gy cm(2), whereas the DAP rate increased from 1.3 to 1.6 Gy cm(2)/min between 2007 and 2014. In the optimized group, optimized radiation application and use of 3D mapping resulted in a continuous decrease in the DAP from 67 to 2 Gy cm(2) and DAP rate from 1.0 to 0.2 Gy cm(2)/min.

CONCLUSION: Currently, the median radiation exposure during PVI in Germany is 28 Gy cm(2). Optimized fluoroscopy by simple means can safely reduce the radiation dose to about 2 Gy cm(2) or even lower when using 3D mapping. Before introducing novel expensive technologies for radiation reduction optimizing of the conventional fluoroscopy is mandatory.

Bibliografische Daten

OriginalspracheEnglisch
ISSN1861-0684
DOIs
StatusVeröffentlicht - 10.2016
PubMed 27142281