CT Pulmonary Angiography at Reduced Radiation Exposure and Contrast Material Volume Using Iterative Model Reconstruction and iDose4 Technique in Comparison to FBP

Abstract

PURPOSE: To assess image quality of CT pulmonary angiography (CTPA) at reduced radiation exposure (RD-CTPA) and contrast medium (CM) volume using two different iterative reconstruction (IR) algorithms (iDose4 and iterative model reconstruction (IMR)) in comparison to filtered back projection (FBP). MATERIALS AND METHODS: 52 patients (body weight < 100 kg, mean BMI: 23.9) with suspected pulmonary embolism (PE) underwent RD-CTPA (tube voltage: 80 kV; mean CTDIvol: 1.9 mGy) using 40 ml CM. Data were reconstructed using FBP and two different IR algorithms (iDose4 and IMR). Subjective and objective image quality and conspicuity of PE were assessed in central, segmental, and subsegmental arteries. RESULTS: Noise reduction of 55% was achieved with iDose4 and of 85% with IMR compared to FBP. Contrast-to-noise ratio significantly increased with iDose4 and IMR compared to FBP (p<0.05). Subjective image quality was rated significantly higher at IMR reconstructions in comparison to iDose4 and FBP. Conspicuity of central and segmental PE significantly improved with the use of IMR. In subsegmental arteries, iDose4 was superior to IMR. CONCLUSIONS: CTPA at reduced radiation exposure and contrast medium volume is feasible with the use of IMR, which provides improved image quality and conspicuity of pulmonary embolism in central and segmental arteries.

Bibliografische Daten

OriginalspracheEnglisch
ISSN1932-6203
DOIs
StatusVeröffentlicht - 2016
PubMed 27611448